Skip to content

ATTUNE 2545-00-127Recalls & reported problems

Depuy (Ireland) · ATTUNE EVALUATION BULLET - NON-SPIKED SIZE 1-10 · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10603295427551), received between Jul 10, 2019 and Jun 17, 2026 — classified as 17 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Device-Device Incompatibility7
  • Material Deformation3
  • Naturally Worn2
  • Scratched Material2

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Insufficient Information4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 17, 2026Material Deformation

    IT WAS REPORTED THAT PRIOR TO SURGERY, IT WAS OBSERVED THAT THE DEVICE HAD BECOME MISSHAPEN OVER TIME. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25534238 ↗
  • MalfunctionMay 11, 2026Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT THE DEVICE WON'T SIT IN TRAY. DAMAGED AT THE BOTTOM. IT WAS BROKEN AT THE BOTTOM BENT AND WON'T SIT IN TRAY. THERE WAS NO TIME ADDED ON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25142613 ↗
  • MalfunctionApr 15, 2026Break

    IT WAS REPORTED THAT THE ATTUNE SYSTEM IMPACTOR, ATTUNE FB TIBIAL IMPACTOR, ATTUNE FEMORAL IMPACTOR, AND ATTUNE EVAL BLT NO-SPK SZ 1-10 WERE BROKEN INTRAOPERATIVELY. THE EVENT INVOLVED A BROKEN MODULAR FB AND FEMORAL IMPACTOR, A BROKEN TIBIAL EVALUATION BULLET, AND A UNIVERSAL IMPACTOR. THERE WAS NO PATIENT INVOLVEMENT OR OBSERVABLE CLINICAL SYMPTOMS, AND NO CONSEQUENCES OR IMPACT TO THE PATIENT WERE IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24887740 ↗
  • MalfunctionJun 20, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT IS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22302612 ↗
  • MalfunctionJan 28, 2025Break; Break

    IT WAS REPORTED THAT THE 5 AND 6 RP INSERT TRIALS WERE BROKEN AND EVAL BULLETS WERE CHIPPED.

    Patient problems: Insufficient Information

    MAUDE report 21253176 ↗
  • MalfunctionJan 28, 2025Break; Break

    IT WAS REPORTED THAT THE 5 AND 6 RP INSERT TRIALS WERE BROKEN AND EVAL BULLETS WERE CHIPPED.

    Patient problems: Insufficient Information

    MAUDE report 21251170 ↗
  • MalfunctionJul 18, 2024Break; Break

    IT WAS REPORTED THAT THERE WERE 2 BROKEN EVAL BULLETS THAT NEED REPLACED, AND A BROKEN SURFACE PIECE. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19770620 ↗
  • MalfunctionJul 18, 2024Break; Break

    IT WAS REPORTED THAT THERE WERE 2 BROKEN EVAL BULLETS THAT NEED REPLACED, AND A BROKEN SURFACE PIECE. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19770617 ↗
  • MalfunctionJun 24, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE FEMORAL AND RP IMPACTORS BROKE, ANGLE WING BENT, EVAL BULLET OLD AND WORN, RP TRIAL POST IS CRACKED. UNKNOWN SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 19597461 ↗
  • MalfunctionJun 24, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE FEMORAL AND RP IMPACTORS BROKE, ANGLE WING BENT, EVAL BULLET OLD AND WORN, RP TRIAL POST IS CRACKED. UNKNOWN SURGICAL DELAY.

    Patient problems: Insufficient Information

    MAUDE report 19597411 ↗
  • MalfunctionFeb 29, 2024Break; Break

    IT WAS REPORTED THAT THE EVAL BULLET WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18810083 ↗
  • MalfunctionNov 15, 2023Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED THAT THE EVALUATION BULLET HAD SCORING MARKS ON ITS SURFACE, RESTRICTING THE MOVEMENT OF RP ARTICULATING SURFACES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18141639 ↗
  • MalfunctionNov 15, 2023Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    IT WAS REPORTED THAT THE EVALUATION BULLET HAD SCORING MARKS ON ITS SURFACE, RESTRICTING THE MOVEMENT OF RP ARTICULATING SURFACES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18141625 ↗
  • MalfunctionSep 24, 2022Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT UPON INSPECTION OF THE BULLETS, I NOTICED ALL WERE NOT SITTING FLUSH WITHIN THE TIBIAL BASE TRIAL, THEY DO NOT APPEAR TO BE DAMAGED BUT NONE OF THEM SEAT CORRECTLY. I COMPARED THEM TO ANOTHER HOSPITAL AND IT APPEARS THEY ARE SLIGHTLY OUT OF SHAPE CAUSING THIS OBSTRUCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15476172 ↗
  • MalfunctionAug 11, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TIBIAL TRAY TRIAL WOULDN¿T CONNECT TO THE TRAY HANDLE AND THE INSERT WOULDN¿T SIT DOWN INSIDE THE TRAY. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12308365 ↗
  • MalfunctionNov 23, 2020Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE EVAL BULLET NO LONGER SITS FLUSH IN TIBIA TRAY. NO SURGICAL DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10888248 ↗
  • MalfunctionJul 10, 2019Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ONCE THE REVISION SURGERY WAS COMPLETED (IT¿S REPORTED AS (B)(4)). IT WAS REPORTED THAT THE SURGEON ATTEMPTED TO CHECK EXTENSION AND FLEXION OF THE PATIENT'S KNEE JOINT DURING THE REVISION SURGERY ON (B)(6) 2019 BY USING THE TIBIAL INSERT TRIAL (P/N: 251700406) WITH THE INSTRUMENTS WHICH WERE COMBINED WITH THE EVALUATION BULLET NO SPIKE (P/N: 254500127) AND THE KEEL PUNCH (P/N: 254500173), HOWEVER THE SHAPE OF THE TIBIAL INSERT TRIAL DID NOT FIT WITH THE INSTRUMENTS AND DID NOT ENGAGED. THE SURGERY WAS COMPLETED, AND IT WAS UNKNOWN WHETHER THERE WAS A SURGICAL DELAY OR NOT. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8775886 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates