Skip to content

TRUMATCH 420581Recalls & reported problems

Depuy (Ireland) · TRUMATCH PERSONALIZED SOLUTIONS PIN GUIDE FEMUR - CT KNEE FOR USE WITH: RIGHT · Class 3

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295430773), received between Jun 26, 2019 and Jun 26, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect Measurement” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Aug 10, 2021Terminated

    Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.

    Z-2396-2021 · Root cause: Software Design Change · FDA record ↗

Most reported problems

Device problems

  • Incorrect Measurement1

Patient problems

  • Not Applicable1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2019Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE TRUMATCH GUIDE DID NOT FIT AS DESIRED AND SET THE FEMUR IN VARUS. IT WAS NOTICED INTER-OP AND CORRECTED USING TRADITIONAL CUTTING GUIDE.

    Patient problems: Not Applicable

    MAUDE report 8734825 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices →