Skip to content

TRUMATCH 420579Recalls & reported problems

Depuy (Ireland) · TRUMATCH PERSONALIZED SOLUTIONS PIN GUIDE SET - CT KNEE FOR USE WITH: RIGHT · Class 3

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295430803), received between Jul 10, 2019 and May 1, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (4 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 10, 2021Terminated

    Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.

    Z-2395-2021 · Root cause: Software Design Change · FDA record ↗

Most reported problems

Device problems

  • Patient-Device Incompatibility4
  • Computer Software Problem2
  • Incorrect Measurement2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • Not Applicable1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 1, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THE LACK OF ACCURACY OF THE PIN GUIDE DUE TO THE POSTERIOR SLOPE OVER 10. EVENT OCCURRED DURING INTRA-OP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21941707 ↗
  • MalfunctionMay 10, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT THE PROXIMAL TIBIA RESECTION WAS CUT WITH APPROXIMATELY 2.5 DEGREES OF ANTERIOR SLOPE. THE PROPOSED PLAN USING THE TIBIA PIN GUIDE WAS SET TO 3 DEGREES OF POSTERIOR SLOPE.

    Patient problems: Insufficient Information

    MAUDE report 19288208 ↗
  • MalfunctionJul 28, 2021Computer Software Problem; Patient-Device Incompatibility; Computer Software Problem; Patient-Device Incompatibility

    PATIENT PLAN DID NOT ALIGN WITH PATIENT ANATOMY, PER SALES REP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12240988 ↗
  • MalfunctionJul 28, 2021Computer Software Problem; Patient-Device Incompatibility; Computer Software Problem; Patient-Device Incompatibility

    PATIENT PLAN DID NOT ALIGN WITH PATIENT ANATOMY, PER SALES REP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12240933 ↗
  • MalfunctionJul 9, 2020Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE TRUMATCH ATTUNE PS PIN GUIDE ON FEMORAL SIDE DID NOT FIT. SURGEON SPENT 5 MINUTES TRYING TO FIT GUIDE ON TO PATIENT FEMUR. GUIDE WOULD NOT FIT WHICH RESULTED IN USING MANUAL INSTRUMENTS. A SURGICAL DELAY OF 10 MINUTES WAS NOTED.

    Patient problems: Not Applicable

    MAUDE report 10252387 ↗
  • MalfunctionJul 10, 2019Incorrect Measurement; Incorrect Measurement

    THE FEMUR GUIDE FIT GREAT BUT WHEN HE MEASURED THE POSTERIOR MEDIAL BONE CHIP HE MEASURED 15MM BUT THE PLAN CALLED FOR 12.1MM RESECTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8775733 ↗

Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices

All Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.