Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295430803), received between Jul 10, 2019 and May 1, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Patient-Device Incompatibility” (4 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Aug 10, 2021Terminated
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
Z-2395-2021 · Root cause: Software Design Change · FDA record ↗
Most reported problems
Device problems
- Patient-Device Incompatibility4
- Computer Software Problem2
- Incorrect Measurement2
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
- No Consequences Or Impact To Patient1
- Not Applicable1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2025Patient-Device Incompatibility; Patient-Device Incompatibility
ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THE LACK OF ACCURACY OF THE PIN GUIDE DUE TO THE POSTERIOR SLOPE OVER 10. EVENT OCCURRED DURING INTRA-OP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21941707 ↗MalfunctionMay 10, 2024Patient-Device Incompatibility; Patient-Device Incompatibility
IT WAS REPORTED THAT THE PROXIMAL TIBIA RESECTION WAS CUT WITH APPROXIMATELY 2.5 DEGREES OF ANTERIOR SLOPE. THE PROPOSED PLAN USING THE TIBIA PIN GUIDE WAS SET TO 3 DEGREES OF POSTERIOR SLOPE.
Patient problems: Insufficient Information
MAUDE report 19288208 ↗MalfunctionJul 28, 2021Computer Software Problem; Patient-Device Incompatibility; Computer Software Problem; Patient-Device Incompatibility
PATIENT PLAN DID NOT ALIGN WITH PATIENT ANATOMY, PER SALES REP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12240988 ↗MalfunctionJul 28, 2021Computer Software Problem; Patient-Device Incompatibility; Computer Software Problem; Patient-Device Incompatibility
PATIENT PLAN DID NOT ALIGN WITH PATIENT ANATOMY, PER SALES REP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12240933 ↗MalfunctionJul 9, 2020Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE TRUMATCH ATTUNE PS PIN GUIDE ON FEMORAL SIDE DID NOT FIT. SURGEON SPENT 5 MINUTES TRYING TO FIT GUIDE ON TO PATIENT FEMUR. GUIDE WOULD NOT FIT WHICH RESULTED IN USING MANUAL INSTRUMENTS. A SURGICAL DELAY OF 10 MINUTES WAS NOTED.
Patient problems: Not Applicable
MAUDE report 10252387 ↗MalfunctionJul 10, 2019Incorrect Measurement; Incorrect Measurement
THE FEMUR GUIDE FIT GREAT BUT WHEN HE MEASURED THE POSTERIOR MEDIAL BONE CHIP HE MEASURED 15MM BUT THE PLAN CALLED FOR 12.1MM RESECTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8775733 ↗
Other Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing devices
- ATTUNE 1506-10-005Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1506-10-004Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing442 reports
- ATTUNE 1518-20-038Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing3 deaths393 reports
- LCS SIGMA 1294-33-130Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 recall310 reports
- ATTUNE 1518-20-035Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing342 reports
- P.F.C. SIGMA 96-0102Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing1 death319 reports
- ATTUNE 1518-20-041Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing308 reports
- ATTUNE 1506-10-007Depuy (Ireland) · Prosthesis, knee, patellofemorotibial, semi-constrained, metal/polymer, mobile bearing284 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.