Skip to content

ATTUNE 252307203Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION TIBIAL AUGMENT TRIAL LL/RM 15MM SIZE 7/8 · Class 1

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10603295437178), received between Aug 2, 2019 and Aug 22, 2025 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Corroded” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Corroded8
  • Contamination /Decontamination Problem2
  • Device-Device Incompatibility1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 22, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT IT WAS NOT POSSIBLE TO PROPERLY FIX THE TRIAL AUGMENTS IN THE TRIAL TIBIAE. THE THREADS OF THE TRIAL APPEAR TO BE COLD WELDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22873598 ↗
  • MalfunctionJun 28, 2024Corroded; Corroded

    IT WAS REPORTED THAT THE ATTUNE REVISION INSTRUMENTS ARE CONTAMINATED WITH RUST. THIS WAS PICKED UP POST CLEANING BY CSSD STAFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19633736 ↗
  • MalfunctionMar 28, 2024Corroded; Corroded

    IT WAS REPORTED THAT INSTRUMENTS WERE RUSTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18992291 ↗
  • MalfunctionOct 26, 2023Corroded; Contamination /Decontamination Problem; Corroded; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE FEMORAL TRIAL HAS STARTED TO RUST DUE TO INCORRECT DETERGENT USED AT THE HOSPITAL. WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: IS THE PATIENT PART OF A CLINICAL STUDY (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. (B)(4). DEVICE PROPERTY OF: NONE. DEVICE IN POSSESSION OF: NONE. BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST. TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18012265 ↗
  • MalfunctionSep 29, 2022Corroded; Corroded

    IT WAS REPORTED THAT THE DEVICES ARE RUSTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15508313 ↗
  • MalfunctionJul 13, 2022Corroded; Corroded

    IT WAS REPORTED THAT THE INSTRUMENTS ARE SHOWING RUST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15002495 ↗
  • MalfunctionJun 24, 2022Corroded; Corroded

    IT WAS REPORTED THAT THE DEVICE RUSTED AFTER WASHING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14799618 ↗
  • MalfunctionSep 8, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    STERILE SERVICES IDENTIFIED MULTIPLE INSTRUMENTS WITH RED RESIDUE PRESENT, MAINLY AFFECTING TIBIAL TRIALS. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12443986 ↗
  • MalfunctionAug 2, 2019Corroded; Corroded

    INSTRUMENTS WERE SENT IN WITH VARIOUS DEGREES OF STAINING. LOAN KIT REFERENCE: ZZINAUK0855.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8856052 ↗
  • MalfunctionAug 2, 2019Corroded; Corroded

    INSTRUMENTS WERE SENT IN WITH VARIOUS DEGREES OF STAINING. LOAN KIT REFERENCE: (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8856050 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.