Skip to content

ATTUNE 252305201Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION TIBIAL AUGMENT TRIAL LL/RM 5MM SIZE 5/6 · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10603295437253), received between Aug 5, 2019 and Apr 22, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Corroded” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Corroded8
  • Contamination /Decontamination Problem3
  • Device-Device Incompatibility3
  • Material Deformation1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Consequences Or Impact To Patient2
  • Insufficient Information1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Device-Device Incompatibility; Unintended Movement

    IT WAS REPORTED THAT DURING WORKSHOPS, THERE WERE DIFFICULTIES IN MATING THE FB TEST TIBIAL BASES WITH THE LL/MR WEDGE TESTS. ACCORDING TO THE REPORT, IN SOME CASES IT WAS IMPOSSIBLE TO GET THE WEDGE TO BE 100% FIXED TO THE TIBIAL TEST, LEAVING IT SLIGHTLY LOOSE WITH A LIGHT BETWEEN BOTH COMPONENTS. IN OTHER CASES, THE WEDGE SCREW WAS FUSED TO THE TIBIAL BASE, MAKING IT IMPOSSIBLE TO DISENGAGE. IT WAS REPORTED THAT IT WAS SUSPECTED THAT SOME OF THE PARTICIPANTS WERE BEING "ROUGH" WITH THE MATERIALS. HOWEVER, AFTER TESTING, AND EVEN THOUGH THE USER WAS CAREFUL, THE SAME THING HAPPENED. ADDITIONALLY, IT WAS REPORTED THAT THE ATTUNE KNEE SYSTEM REVISION ASSEMBLY OFFSET STABILIZER X2, BROKE WHEN THE DECOUPLING OF THE DEFINITIVE IMPLANTS WAS CARRIED OUT. IN BOTH CASES, THE DEVICE BROKE AFTER COUPLING AND UNCOUPLING APPROXIMATELY 10 TIMES. IT WAS FURTHER NOTED THAT THE USE OF THE DEVICE WAS CORRECT. IT WAS REPORTED THAT DUE TO THE SIZE OF THE PLASTIC BONES, THE DEVICE WORKED WITH TIBIAL BASES 4 AND 5. THE USER HAD PROBLEMS WITH USING THE 6. HOWEVER, IF NECESSARY, IT COULD BE USED. WAS THERE ANY HARM TO THE REPORTED PATIENT OR USER? NO. WAS THERE A SIGNIFICANT DELAY? NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24957096 ↗
  • MalfunctionAug 21, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED DURING A WORKSHOP IT WAS NOT POSSIBLE TO PROPERLY FIX THE THE TRIAL AUGMENTS IN THE TRIAL TIBIAE. IT SEEMS THAT SOMETHING IS WRONG WITH THE TREADS OF THE TRIAL. THEY ARE SOMEHOW COLD WELDED. THE INSTRUMENTS WERE DIFFICULT TO SCREW TOGETHER AND STOPPED WORKING BEFORE THEY ARE FIRMLY CONNECTED. AT THIS POINT, THE ITEMS COULD NO LONGER BE UNSCREWED. (B)(4) IS RELATED TO, (B)(4) - 10 IPS, (B)(4) - 5 IPS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22862466 ↗
  • MalfunctionMay 9, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE TIBIAL AUGMENTATION WEDGE IS ASSEMBLED TO THE TRIAL TRAY, A TRIAL IS PERFORMED, AND WHEN THE CHANGE IS MADE TO TEST WITH ANOTHER HEIGHT IN THE WEDGE, IT IS EVIDENT THAT IT IS COLD-FUSED.

    Patient problems: Insufficient Information

    MAUDE report 21993559 ↗
  • MalfunctionJul 1, 2024Corroded; Contamination /Decontamination Problem; Corroded; Contamination /Decontamination Problem

    ATTUNE REVISION GENERAL TIBIAL PREP TRAY CONTAMINATED WITH RUST ON SOME INSTRUMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643604 ↗
  • MalfunctionMar 28, 2024Corroded; Corroded

    IT WAS REPORTED THAT INSTRUMENTS WERE RUSTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18992233 ↗
  • MalfunctionOct 26, 2023Corroded; Contamination /Decontamination Problem; Corroded; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE FEMORAL TRIAL HAS STARTED TO RUST DUE TO INCORRECT DETERGENT USED AT THE HOSPITAL.WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED? UNKNOWN.IS THE PATIENT PART OF A CLINICAL STUDY? UNKNOWN.(B)(4).DEVICE PROPERTY OF? NONE.DEVICE IN POSSESSION OF? NONE.(B)(4).DEVICE PROPERTY OF? NONE.DEVICE IN POSSESSION OF ? NONE.(B)(4).DEVICE PROPERTY OF ? NONE.DEVICE IN POSSESSION OF? NONE.(B)(4).DEVICE PROPERTY OF ? NONE.DEVICE IN POSSESSION OF? NONE.(B)(4).DEVICE PROPERTY OF? NONE.DEVICE IN POSSESSION OF? NONE.(B)(4).DEVICE PROPERTY OF? NONE.DEVICE IN POSSESSION OF? NONE.(B)(4).DEVICE PROPERTY OF? NONE.DEVICE IN POSSESSION OF? NONE.(B)(4).DEVICE PROPERTY OF? NONE.DEVICE IN POSSESSION OF? NONE.(B)(4).DEVICE PROPERTY OF? NONE.DEVICE IN POSSESSION OF? NONE.(B)(4).DEVICE PROPERTY OF? NONE.DEVICE IN POSSESSION OF? NONE.BY CHECKING THIS BOX I CERTIFY THAT ALL INFORMATION THAT ARE KNOWN/AVAILABLE HAS BEEN DISCLOSED. IF ANY NEW INFORMATION WILL BE MADE AVAILABLE, THE ADDITIONAL INFORMATION WILL BE SUBMITTED THROUGH CST? TRUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18012198 ↗
  • MalfunctionSep 29, 2022Corroded; Corroded

    IT WAS REPORTED THAT THE DEVICES ARE RUSTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15508678 ↗
  • MalfunctionJul 22, 2022Corroded; Corroded

    IT WAS REPORTED THAT THE INSTRUMENTS ARE SHOWING RUST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15085638 ↗
  • MalfunctionJun 24, 2022Corroded; Corroded

    IT WAS REPORTED THAT THE DEVICE RUSTED AFTER WASHING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14799656 ↗
  • MalfunctionSep 8, 2021Contamination /Decontamination Problem; Contamination /Decontamination Problem

    STERILE SERVICES IDENTIFIED MULTIPLE INSTRUMENTS WITH RED RESIDUE PRESENT, MAINLY AFFECTING TIBIAL TRIALS. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12443968 ↗
  • MalfunctionAug 5, 2019Corroded; Corroded

    INSTRUMENTS WERE SENT IN WITH VARIOUS DEGREES OF STAINING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8861927 ↗
  • MalfunctionAug 5, 2019Corroded; Corroded

    INSTRUMENTS WERE SENT IN WITH VARIOUS DEGREES OF STAINING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8861926 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page