ATTUNE 252309101Recalls & reported problems
Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION TIBIAL AUGMENT TRIAL LM/RL 5MM SIZE 9/10 · Class 1
All models of ATTUNE (1475)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295437307), received between Oct 31, 2023 and Jun 28, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Corroded” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Corroded2
- Improper or Incorrect Procedure or Method1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 28, 2024Corroded; Corroded
IT WAS REPORTED THAT THE ATTUNE REVISION INSTRUMENTS ARE CONTAMINATED WITH RUST. THIS WAS PICKED UP POST CLEANING BY CSSD STAFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19633690 ↗MalfunctionOct 31, 2023Corroded; Improper or Incorrect Procedure or Method; Corroded; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE FEMORAL TRIAL HAS STARTED TO RUST DUE TO INCORRECT DETERGENT USED AT THE HOSPITAL. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. PATIENT STATUS/ OUTCOME / CONSEQUENCES . , WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY : UNKNOWN, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4). DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NON
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18039852 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page