ATTUNE 2504-40-501Recalls & reported problems
Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION BROACH ADAPTOR SIZE 4-7 · Class 1
All models of ATTUNE (1475)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10603295438267), received between Oct 7, 2019 and Sep 22, 2023 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Twisted/Bent5
- Device-Device Incompatibility2
- Naturally Worn2
- Material Deformation1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient4
- No Code Available1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2023Naturally Worn; Unintended Movement; Naturally Worn; Unintended Movement
ADDITIONAL INFORMATION RECEIVED. A. WAS THERE A SURGICAL DELAY? IF YES, WHAT IS THE DURATION OF THE DELAY? THERE WAS NO SURGICAL DELAY. B. WAS THERE ANY ADVERSE CONSEQUENCES THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT? THE PATIENT WAS NOT AFFECTED. C. PLEASE CONFIRM IF THE BOLT WAS STRIPPED OR NOT. THE BOLT WAS STRIPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17798037 ↗MalfunctionOct 1, 2021Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE THREADS ARE STRIPPED AND NEEDS REPLACING. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12562610 ↗MalfunctionAug 10, 2021Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED THAT THE SET SCREW ON THE BROACH ADAPTER IS DAMAGED AND WILL NO LONGER STAY IN PLACE. THREADS ON THE STEM TRIAL ARE STRIPPED AND WILL NO LONGER ATTACH TO BROACH OR BOSS ADAPTER. THIS DID NOT OCCUR DURING SURGERY. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12302100 ↗MalfunctionApr 19, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE THREADS ON THE BROACH ADAPTER WERE STRIPPED. A REPLACEMENT IS NEEDED TO BE SENT TO THE CHARLESTON WAREHOUSE. NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11685725 ↗MalfunctionJan 6, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE SCREW THREADS ARE STRIPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11123043 ↗MalfunctionAug 4, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE THREADS OF THE BROACH ADAPTER ARE STRIPPED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10364321 ↗MalfunctionJul 24, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE THREADS WAS STRIPPED. NEEDS REPLACING. NO SURGICAL DELAY. THERE WERE NO ADVERSE EVENTS OR PATIENT CONSEQUENCES REPORTED.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10321415 ↗MalfunctionJun 30, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE ATTUNE REV BROACH ADAPTOR WAS CROSS-THREADED. NO SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10213514 ↗MalfunctionOct 7, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT WHEN ATTACHING THE ATTUNE REVISION FEMORAL BROACH ADAPTOR IT WOULDN¿T FULLY LOCK IN WHEN USING THE SCREWDRIVER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9162087 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page