Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295438809), received between Sep 12, 2018 and Sep 17, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Entrapment of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2025Break; Break
WHILE BROACHING FOR THE ATTUNE REVISION TIBIAL SLEEVE, THE 12X60 SHEARED OFF THE 29MM TIBIAL SLEEVE. MANAGED TO FISH THE STEM OUT WITH A LONG GRASPER SO VERY LITTLE DELAY TO THE SURGERY AND NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23072970 ↗MalfunctionJul 27, 2021Break; Break
12X60 ATTUNE REVISION STEM SHEARED OFF A 45MM TIBIAL BROACH, LEAVING BOTH ITEMS UNUSABLE AND NEEDING REPLACEMENT. SURGEON WAS BROACHING THE TIBIA WHEN IT HAPPENED. NO REAL DELAY TO THE CASE BUT HE DID HAVE TO RETRIEVE THE BROKEN STEM TRIAL FROM THE TIBIAL CANAL. EVERYTHING WAS REMOVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12235482 ↗InjuryNov 23, 2020Break; Entrapment of Device; Break; Entrapment of Device
THE CONEXIÓN BETWEEN SLEEVE AND STEM WAS BROKEN AND THE STEM WAS LEFT INSIDE THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 10885118 ↗MalfunctionSep 12, 2018Break; Break
IT WAS REPORTED THAT THE STEM TRIAL FRACTURED AT THE THREADED JUNCTION OF THE PART DURING REMOVAL OF THE BROACH IN THE TIBIA ON A SAWBONE. A SALES REPRESENTATIVE, WHO WAS UNDERGOING TRAINING THROUGH THE ATTUNE REVISION TRAINING DEMONSTRATION, OBSERVED THE FAILURE WHILE COMPLETING THE TASK. IT WAS CONFIRMED THAT THE STEM TRIAL WAS THE OLD DESIGN, WHICH WAS RELEASED WITH THE EARLY LAUNCHES OF THE ATTUNE REVISION KNEE SYSTEM (PRIOR TO THE STEM TRIALS BEING OPTIMIZED FOR USE WITH BROACHING). IT SHOULD BE NOTED THAT THE STEM TRIAL DIAMETER, IN RELATION TO THE TIBIAL CANAL, WAS INTENTIONALLY UNDERSIZED DURING THE DEMONSTRATION. THE UNDER-SIZING OF THE STEM TRIAL, COULD HAVE LED TO THE PART BEING FURTHER OFF-AXIS IN RELATION TO THE LINE OF IMPACTION, THEREFORE INCREASING THE OVERALL STRESS ON THE THREADED JUNCTION IF IT WERE TO BE SLIGHTLY LOOSE FROM THE MATING INTERFACE (BROACH). THE PART WAS RETRIEVED FROM THE SAWBONE. NO PATIENT INTERACTION AS IT OCCURRED DURING A TRAINING DEMONSTRATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7868061 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.