ATTUNE 2513-14-110Recalls & reported problems
Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION PRESS-FIT STEM TRIAL 14mm x 110mm · Class 1
All models of ATTUNE (1475)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295438816), received between Sep 11, 2020 and Aug 12, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Device-Device Incompatibility1
Patient problems
- Bone Fracture(s)1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 12, 2025Break; Break
IT WAS REPORTED THAT THE THREADS OF THE 110X14 PRESS-FIT STEM TRIAL BROKE OFF INSIDE THE 37 MM TIBIAL SLEEVE BROACH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22784122 ↗InjurySep 11, 2020Device-Device Incompatibility; Adverse Event Without Identified Device or Use Problem; Device-Device Incompatibility; Adverse Event Without Identified Device or Use Problem
DURING A 2ND STAGE KNEE REVISION THE SURGEON REPORTED FRACTURING THE PROXIMAL TIBIA WHEN SEATING THE ATTUNE REVISION TIBIAL BROACH IN THE PROXIMAL TIBIA. THE SURGEON HAD ALREADY PREPARED THE PROXIMAL TIBIA BONE WITH A STEM PILOT/BROACH AND TIB TRAY HOWEVER WHEN HE EXTRACTED THE TRIAL CONSTRUCT (WITH RP ADAPTER AND SYSTEM HANDLE) THE TIB TRAY ROTATION WAS NOT LOCKED AND THEREFORE HE WAS REQUIRED TO RE-SEAT THE STEM TRIAL/BROACH INTO THE PREPARED BONE ENVELOPE AND LOCK OFF HIS RP TRAY ROTATION BEFORE EXTRACTING AND TRANSFERRING THIS TO THE DEFINITIVE IMPLANT/ASSEMBLING. HE COMMENTED THAT WHEN THE REGISTRAR RE-INSERTED THE BROACH INTO THE BONE SHE HIT IT TOO HARD WHICH HE THINKS CAUSED THE FRACTURE. IT WAS WORTH NOTING THAT THE REGISTRAR ALSO HAD GREAT DIFFICULTY EXTRACTING THE TRIAL STEM AND BROACH/TRAY CONSTRUCT ONCE THE ROTATION HAD BEEN LOCKED AND WHEN THE SURGEON TOOK OVER, HE ENDED UP REMOVING THE RP TRIAL TRAY AND ATTACHING THE BROACH HANDLE TO REMOVE THE BROACH (WHICH IS WHEN HE NOTICED THE SMALL FRACTURE IN THE PROX TIB) AND REPEATING THE ABOVE STEPS TO SET THE ROTATION. THE SURGERY CONTINUED AS PLANNED AND HE ADVISED THAT THIS FRACTURE WOULD NOT CHANGE THE OUTCOME OF THE SUR
Patient problems: Bone Fracture(s); No Code Available
MAUDE report 10521878 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.