ATTUNE 2513-14-110Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION PRESS-FIT STEM TRIAL 14mm x 110mm · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10603295438816), received between Sep 11, 2020 and Aug 12, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Device-Device Incompatibility1

Patient problems

  • Bone Fracture(s)1
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2025Break; Break

    IT WAS REPORTED THAT THE THREADS OF THE 110X14 PRESS-FIT STEM TRIAL BROKE OFF INSIDE THE 37 MM TIBIAL SLEEVE BROACH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22784122 ↗
  • InjurySep 11, 2020Device-Device Incompatibility; Adverse Event Without Identified Device or Use Problem; Device-Device Incompatibility; Adverse Event Without Identified Device or Use Problem

    DURING A 2ND STAGE KNEE REVISION THE SURGEON REPORTED FRACTURING THE PROXIMAL TIBIA WHEN SEATING THE ATTUNE REVISION TIBIAL BROACH IN THE PROXIMAL TIBIA. THE SURGEON HAD ALREADY PREPARED THE PROXIMAL TIBIA BONE WITH A STEM PILOT/BROACH AND TIB TRAY HOWEVER WHEN HE EXTRACTED THE TRIAL CONSTRUCT (WITH RP ADAPTER AND SYSTEM HANDLE) THE TIB TRAY ROTATION WAS NOT LOCKED AND THEREFORE HE WAS REQUIRED TO RE-SEAT THE STEM TRIAL/BROACH INTO THE PREPARED BONE ENVELOPE AND LOCK OFF HIS RP TRAY ROTATION BEFORE EXTRACTING AND TRANSFERRING THIS TO THE DEFINITIVE IMPLANT/ASSEMBLING. HE COMMENTED THAT WHEN THE REGISTRAR RE-INSERTED THE BROACH INTO THE BONE SHE HIT IT TOO HARD WHICH HE THINKS CAUSED THE FRACTURE. IT WAS WORTH NOTING THAT THE REGISTRAR ALSO HAD GREAT DIFFICULTY EXTRACTING THE TRIAL STEM AND BROACH/TRAY CONSTRUCT ONCE THE ROTATION HAD BEEN LOCKED AND WHEN THE SURGEON TOOK OVER, HE ENDED UP REMOVING THE RP TRIAL TRAY AND ATTACHING THE BROACH HANDLE TO REMOVE THE BROACH (WHICH IS WHEN HE NOTICED THE SMALL FRACTURE IN THE PROX TIB) AND REPEATING THE ABOVE STEPS TO SET THE ROTATION. THE SURGERY CONTINUED AS PLANNED AND HE ADVISED THAT THIS FRACTURE WOULD NOT CHANGE THE OUTCOME OF THE SUR

    Patient problems: Bone Fracture(s); No Code Available

    MAUDE report 10521878 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page