Skip to content

ATTUNE 2513-16-060Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION PRESS-FIT STEM TRIAL 16mm x 60mm · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10603295438861), received between Mar 6, 2020 and Jan 2, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Device-Device Incompatibility2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information2
  • No Consequences Or Impact To Patient2
  • No Code Available1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 2, 2026Break; Device-Device Incompatibility

    [SURGEON] - KNEE REVISION - [HOSPITAL] : WHILE PERFORMING TIBIAL BROACHING, THE 16 × 60 PRESS-FIT STEM TRIAL THAT HAD BEEN ATTACHED TO THE DISTAL END OF THE SIZE 37 TIBIAL BROACH SNAPPED AND REMAINED TRAPPED IN THE TIBIAL CANAL WITH NO REMAINING THREADS. THE STEM WAS FORTUNATELY RETRIEVED USING A KOCHER. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? UNKNOWN, WAS PROCEDURE SUCCESSFULLY COMPLETED? UNKNOWN, WERE FRAGMENTS GENERATED? UNKNOWN, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN, PATIENT STATUS/ OUTCOME / CONSEQUENCES , WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED: UNKNOWN, IS THE PATIENT PART OF A CLINICAL STUDY UNKNOWN, (B)(4), DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE, (B)(4) DEVICE PROPERTY OF NONE, DEVICE IN POSSESSION OF NONE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23949746 ↗
  • MalfunctionDec 4, 2024Break; Break

    IT WAS REPORTED THAT THE PATIENT HAD AN INFECTED LEFT TOTAL KNEE. THE INFECTION IS GONE, SO THE SURGEON REMOVED THE TEMPORARY SPACER AND IMPLANTED DEFINITIVE COMPONENTS. DURING BROACHING OF THE TIBIA, THE STEM TRIAL BROKE OFF THE END OF THE BROACH AND WAS STUCK IN THE TIBIAL CANAL. IT TOOK 5 MINUTES TO GET THE BROKEN TRIAL STEM OUT OF THE CANAL AND TO GET THE THREADED PIECE OUT OF THE BROACH SO THE BROACH COULD BE USED AGAIN. DOI: (B)(6) 2024. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT KNEE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20845336 ↗
  • MalfunctionJun 3, 2024Break; Break

    ATTUNE REVISION PRESS FIT STEM TRIAL BROKE OFF FROM TIBIAL BROACH DURING TIBIAL BROACHING.

    Patient problems: Insufficient Information

    MAUDE report 19447406 ↗
  • MalfunctionOct 19, 2022Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT TRIAL IS BROKEN. THERE WAS 20 MINUTES SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15632766 ↗
  • MalfunctionOct 15, 2020Break; Break

    ADDITIONAL INFORMATION RECEIVED STATES THAT THE MALE PART OF THE STEM TRIAL THAT THREADS INTO THE BROACH IS WHAT BROKE.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10684488 ↗
  • MalfunctionMar 6, 2020Break; Break

    IT WAS REPORTED THAT THE STEM TRIAL "MALE" ADAPTER SNAPPED OFF INSIDE OF THE 29 MM TIBIAL BROACH. BOTH THE STEM TRIAL AND THE BROACH WERE RETRIEVED. INCIDENT HAPPENED DURING TIBIAL BROACHING. PATIENT WAS NOT INJURED, NO DELAY IN SURGERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9801480 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates