Skip to content

ATTUNE 2513-18-060Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION PRESS-FIT STEM TRIAL 18mm x 60mm · Class 1

All models of ATTUNE (1475)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10603295438892), received between Jul 3, 2018 and Apr 29, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Break” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break11

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Insufficient Information1
  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Break

    IT WAS REPORTED THAT DURING A TOTAL KNEE ARTHROPLASTY (TKA) SURGICAL PROCEDURE IT WAS OBSERVED THAT THE STEM TRIAL THREADS BROKE OFF AT THE SLEEVE JUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25031273 ↗
  • MalfunctionMay 4, 2023Break; Break

    IT WAS REPORTED STEM TRIAL BROKE AT BASE LEAVING THREADS IN BROACH TRIAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16867428 ↗
  • MalfunctionDec 15, 2022Break; Break

    IT WAS REPORTED THAT THREADED PART OF THE TIP OF STEM TRIAL BROKE OFF WHEN TAKING IT OFF BROACH TRIAL. NO PIECES WERE LEFT IN PATIENT. THERE WAS NO DELAY SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15990446 ↗
  • MalfunctionNov 16, 2022Break; Break

    IT WAS REPORTED THAT THE REVISION STEM TRIAL BROKE INSIDE OF THE BROACH WHILE BROACHING

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15807077 ↗
  • MalfunctionAug 2, 2022Break; Break

    IT WAS REPORTED THAT THE STEM TRIAL BROKE. THE THREADED PORTION JUST ABOVE THE STEM TRIAL SNAPPED. THE THREADED BROKEN PIECE REMAINED IN THE 29 SLEEVE. ALL PIECES ACCOUNTED FOR AND THE CASE WAS NOT DELAYED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15152198 ↗
  • MalfunctionJul 13, 2022Break; Break

    IT WAS REPORTED THAT THE 18MM STEM TRIAL THREADED CONNECTION WAS BROKEN. THIS INSTRUMENT WAS NOT USED IN SURGERY. IT WAS DISCOVERED DURING AN OFFICE IN SERVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15001886 ↗
  • MalfunctionMar 28, 2022Break; Break

    IT WAS REPORTED THAT THE18MM TRIAL STEM BROKE AT THREAD/PILOT JUNCTION. ALL PIECES WERE ACCOUNTED FOR. THERE IS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13928319 ↗
  • MalfunctionDec 20, 2021Break; Break

    ATTUNE REVISION 18MMX60MM STEM TRIAL BROKE OFF IN THE TIBIA WHILE BROACHING THE TIBIA TO ACCEPT A TIBIAL SLEEVE. THE STEM TRIAL BROKE RIGHT WHERE THE MALE THREADS MEET THE TRIAL ITSELF WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? YES, IF YES, NUMBER OF MINUTES: 1, WAS PROCEDURE SUCCESSFULLY COMPLETED? YES, WERE FRAGMENTS GENERATED? YES, IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? YES.

    Patient problems: Insufficient Information

    MAUDE report 13044314 ↗
  • MalfunctionMar 26, 2021Break; Break

    IT WAS REPORTED THAT THE THREADS FROM THE 18X60 ATTUNE REVISION KNEE TRIAL BROKE OFF INTO THE 45 TIBIAL BROACH. ALL PARTS WERE ACCOUNTED FOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11574721 ↗
  • MalfunctionSep 22, 2020Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE THREADS BROKE OFF THE BODY OF THE STEM AND THIS SHOULD BE IDENTIFIED WHEN THE PRODUCT IS RETURNED.

    Patient problems: No Code Available

    MAUDE report 10566860 ↗
  • MalfunctionJul 3, 2018Break; Break

    IT WAS REPORTED THAT DURING REVISION THE STEM DRILL BROKE OFF THE BROACH. THE STEM TRIAL WAS STUCK IN THE PATIENTS TIBIA AND IT WAS RETRIEVED WITHOUT HAVING ANY CAUSE OR DAMAGE TO PATIENT. ALL PIECES WERE RETRIEVED FROM THE PATIENT. SURGICAL DELAY OF 20-30 MINUTES.

    Patient problems: Not Applicable

    MAUDE report 7658879 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates