Skip to content

ATTUNE 2547-06-004Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION DISTAL FEMORAL AUGMENT TRIAL 16MM X SIZE 6 · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
22
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 10603295439394), received between Aug 24, 2018 and May 13, 2025 — classified as 22 malfunctions.

The most frequently reported specific device problem is “Corroded” (20 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 10 reportsMay 2025: 6 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Corroded20
  • Contamination /Decontamination Problem4
  • Material Discolored4
  • Appropriate Device Problem Term/Code Not Available2
  • Device Contamination with Chemical or Other Material2

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • No Consequences Or Impact To Patient6

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2025Corroded; Material Discolored; Corroded; Material Discolored

    IT WAS REPORTED THAT THE PRODUCT(S) HAS RUSTING AND DISCOLORATION. THE ISSUE WAS IDENTIFIED DURING LOAN KIT INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22011787 ↗
  • MalfunctionMay 13, 2025Corroded; Material Discolored; Corroded; Material Discolored

    IT WAS REPORTED THAT THE THE PRODUCT(S) HAS RUSTING AND DISCOLORATION. THE ISSUE WAS IDENTIFIED DURING LOAN KIT INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22011314 ↗
  • MalfunctionMay 12, 2025Corroded; Material Discolored; Corroded; Material Discolored

    IT WAS REPORTED THAT DURING LOAN KIT INSPECTION IT WAS IDENTIFIED THAT THE PRODUCT SHOWS RUSTING AND DISCOLORATION ON PRODUCT INFORMATION. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22005157 ↗
  • MalfunctionMay 12, 2025Corroded; Material Discolored; Corroded; Material Discolored

    IT WAS REPORTED THAT DURING LOAN KIT INSPECTION IT WAS IDENTIFIED THAT THE PRODUCT SHOWS RUSTING AND DISCOLORATION ON PRODUCT INFORMATION. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22005149 ↗
  • MalfunctionMay 8, 2025Corroded; Contamination /Decontamination Problem; Corroded; Contamination /Decontamination Problem

    IT WAS REPORTED THAT DURING LOAN KIT INSPECTION IT WAS NOTICED THAT THE AUGMENTS ARE RUSTING AROUND THE MAGNET AREA AND CANNOT BE CLEANED EFFECTIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21986840 ↗
  • MalfunctionMay 8, 2025Corroded; Contamination /Decontamination Problem; Corroded; Contamination /Decontamination Problem

    IT WAS REPORTED THAT DURING LOAN KIT INSPECTION IT WAS NOTICED THAT THE AUGMENTS ARE RUSTING AROUND THE MAGNET AREA AND CANNOT BE CLEANED EFFECTIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21986709 ↗
  • MalfunctionApr 25, 2025Corroded; Corroded

    ATUN REV DST F AUGTRL 16XSZ7-8 AND ATTUNE REV DST F AUGTRL 16XSZ6 HAVE POTENTIAL SURFACE RUST. THE PRODUCT DAMAGE DOCUMENTED IN PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION BY THE LOAN KIT TECHNICIAN. THE EVENT AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE WAS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21911458 ↗
  • MalfunctionApr 25, 2025Corroded; Corroded

    ATUN REV DST F AUGTRL 16XSZ7-8 AND ATTUNE REV DST F AUGTRL 16XSZ6 HAVE POTENTIAL SURFACE RUST. THE PRODUCT DAMAGE DOCUMENTED IN PC HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION BY THE LOAN KIT TECHNICIAN. THE EVENT AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE WAS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21911256 ↗
  • MalfunctionApr 24, 2025Corroded; Corroded

    IT WAS REPORTED THAT DURING LOAN KIT INSPECTION ON AN UNKNOWN DATE IN 2025, IT WAS IDENTIFIED THAT FEMORAL TRIALS HAVE POTENTIAL SURFACE RUST. THERE WAS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21903300 ↗
  • MalfunctionApr 24, 2025Corroded; Corroded

    IT WAS REPORTED THAT DURING LOAN KIT INSPECTION ON AN UNKNOWN DATE IN 2025, IT WAS IDENTIFIED THAT FEMORAL TRIALS HAVE POTENTIAL SURFACE RUST. THERE WAS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21903298 ↗
  • MalfunctionApr 24, 2025Corroded; Corroded

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE ATUN REV DST F AUGTRL 16XSZ7-8 AND ATTUNE REV DST F AUGTRL 16XSZ6 HAVE POTENTIAL SURFACE RUST. THIS WAS IDENTIFIED DURING LOAN KIT INSPECTION BY THE LOAN KIT TECHNICIAN. THERE ARE NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898749 ↗
  • MalfunctionApr 24, 2025Corroded; Corroded

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE ATUN REV DST F AUGTRL 16XSZ7-8 AND ATTUNE REV DST F AUGTRL 16XSZ6 HAVE POTENTIAL SURFACE RUST. THIS WAS IDENTIFIED DURING LOAN KIT INSPECTION BY THE LOAN KIT TECHNICIAN. THERE ARE NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898747 ↗
  • MalfunctionApr 23, 2025Corroded; Corroded

    ATUN REV DST F AUGTRL 16XSZ7-8 AND ATTUNE REV DST F AUGTRL 16XSZ6 POSSIBLE SURFACE RUST SURROUNDING MAGNETS. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIES DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT AND ALERT DATE ARE THE DATE THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21893439 ↗
  • MalfunctionApr 23, 2025Corroded; Corroded

    ATUN REV DST F AUGTRL 16XSZ7-8 AND ATTUNE REV DST F AUGTRL 16XSZ6 POSSIBLE SURFACE RUST SURROUNDING MAGNETS. THE PRODUCT DAMAGE DOCUMENTED IN THIS PC HAS BEEN IDENTIFIES DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT AND ALERT DATE ARE THE DATE THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE. NO FURTHER INFORMATION CAN BE OBTAINED AS THE CASE WAS NOT REPORTED BY THE CUSTOMER..

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21893407 ↗
  • MalfunctionApr 10, 2025Corroded; Contamination /Decontamination Problem; Corroded; Contamination /Decontamination Problem

    IT WAS REPORTED THAT A TARNISHING WAS FOUND ON THE TOP AND BOTTOM OF THE TRIALS. IT WAS UNABLE TO REMOVE THROUGH STANDARD CLEANING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21816852 ↗
  • MalfunctionApr 10, 2025Corroded; Contamination /Decontamination Problem; Corroded; Contamination /Decontamination Problem

    IT WAS REPORTED THAT A TARNISHING WAS FOUND ON THE TOP AND BOTTOM OF THE TRIALS. IT WAS UNABLE TO REMOVE THROUGH STANDARD CLEANING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21816847 ↗
  • MalfunctionOct 14, 2019Corroded; Corroded

    IT WAS REPORTED THAT THE INSTRUMENTS APPEAR TO HAD RUST ON THEM. NOTHING LEFT IN PATIENT. NASHVILLE LOANER 4 ATTUNE REVISION GENERAL FEMORAL PREP CONTINUED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9185143 ↗
  • MalfunctionOct 14, 2019Corroded; Corroded

    IT WAS REPORTED THAT THE INSTRUMENTS APPEAR TO HAD RUST ON THEM. NOTHING LEFT IN PATIENT. NASHVILLE LOANER 4 ATTUNE REVISION GENERAL FEMORAL PREP CONTINUED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9185131 ↗
  • MalfunctionJan 25, 2019Corroded; Device Contamination with Chemical or Other Material; Corroded; Device Contamination with Chemical or Other Material

    IN DEPUY LOANER DEPARTMENT METAL DEBRIS WAS SEEN ON FEMUR AUGMENTS OF THE ATTUNE REVISION SETS. ALL AUGMENTS WERE AFFECTED BUT DIFFERENT DEGREES. THE INSTRUMENTATION WAS READY PACKED FOR SHIPMENT. THE INSTRUMENTATION WAS SHIPPED A TOTAL OF 8 TIMES AND WAS THOROUGHLY CLEANED 3 TIMES WITH ALCOHOL BY AN ORTHOKIT OPERATOR. IN ADDITION, THERE ARE NOTICEABLE RUST MARKS, PROBABLY FROM THE ENGRAVING. THIS WAS SEEN INTERNALLY; NO PATIENT INVOLVEMENT, NO SURGERY DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8279495 ↗
  • MalfunctionJan 25, 2019Corroded; Device Contamination with Chemical or Other Material; Corroded; Device Contamination with Chemical or Other Material

    IN DEPUY LOANER DEPARTMENT METAL DEBRIS WAS SEEN ON FEMUR AUGMENTS OF THE ATTUNE REVISION SETS. ALL AUGMENTS WERE AFFECTED BUT DIFFERENT DEGREES. THE INSTRUMENTATION WAS READY PACKED FOR SHIPMENT. THE INSTRUMENTATION WAS SHIPPED A TOTAL OF 8 TIMES AND WAS THOROUGHLY CLEANED 3 TIMES WITH ALCOHOL BY AN ORTHOKIT OPERATOR. IN ADDITION, THERE ARE NOTICEABLE RUST MARKS, PROBABLY FROM THE ENGRAVING. THIS WAS SEEN INTERNALLY; NO PATIENT INVOLVEMENT, NO SURGERY DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8279466 ↗
  • MalfunctionAug 24, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    TRIAL SHOWS MAGNETIC/METAL WEAR BEFORE USE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7814095 ↗
  • MalfunctionAug 24, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    TRIAL SHOWS MAGNETIC/METAL WEAR BEFORE USE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7814050 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates