Skip to content

ATTUNE 2546-00-415Recalls & reported problems

Depuy (Ireland) · ATTUNE KNEE SYSTEM REVISION SYSTEM HEX ATTACHMENT · Class 1

All models of ATTUNE (1475)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Attune hex attachment is a driver piece for the Attune knee revision system. Reports mostly describe the attachment seizing or cold-welding to handles or trial parts, with stripped threads, spinning, jamming or breaking. There are 25 reports naming its identifier, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 10603295451280), received between Jul 19, 2018 and Oct 21, 2025 — classified as 25 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility17
  • Break5
  • Material Deformation5
  • Mechanical Jam4
  • Material Twisted/Bent3
  • Device Slipped1
  • Difficult to Remove1
  • Illegible Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • No Consequences Or Impact To Patient7
  • Insufficient Information3
  • No Code Available1

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2025Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT IT WAS NOT POSSIBLE TO PROPERLY FIX THE TRIAL AUGMENTS IN THE TRIAL TIBIAE. THE THREADS OF THE TRIAL APPEAR TO BE COLD WELDED. TWO ADDITIONAL DEVICES WERE FOUND TO BE AFFECTED DURING ANALYSIS OF RETURNED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23351428 ↗
  • MalfunctionOct 21, 2025No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THAT IT WAS NOT POSSIBLE TO PROPERLY FIX THE TRIAL AUGMENTS IN THE TRIAL TIBIAE. THE THREADS OF THE TRIAL APPEAR TO BE COLD WELDED. TWO ADDITIONAL DEVICES WERE FOUND TO BE AFFECTED DURING ANALYSIS OF RETURNED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23349662 ↗
  • MalfunctionMay 22, 2025Device-Device Incompatibility; Illegible Information; Device-Device Incompatibility; Illegible Information

    IT WAS REPORTED THAT WHILE DISASSEMBLING THE HEX ATTACHMENT FROM THE BLACK HANDLE SCREWDRIVER, THE HEX ATTACHMENT WAS COLD WELDED ONTO THE BLACK HANDLE. THERE WAS NO PATIENT HARM, NO DELAY IN THE CASE. NOTHING BROKEN FROM THE TWO INSTRUMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22069625 ↗
  • MalfunctionFeb 3, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT STAFF HAD A SCREW DRIVER HANDLE THAT WAS JUST SPINNING AND A DRIVER THAT WAS STRIPPED. THERE WAS NO DELAY OR PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21295408 ↗
  • MalfunctionOct 11, 2024Break; Break

    IT WAS REPORTED THE INSTRUMENT WAS BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20427483 ↗
  • MalfunctionAug 21, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE HANDLES ARE STUCK TOGETHER.

    Patient problems: Insufficient Information

    MAUDE report 20020854 ↗
  • MalfunctionMar 18, 2024Break; Break

    IT WAS REPORTED THAT HEX SCREWDRIVER WAS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18920887 ↗
  • MalfunctionFeb 19, 2024Device-Device Incompatibility; Device-Device Incompatibility

    BREAKAGE OF SMALL ATTUNE REVISION SCREWDRIVER. SHALT WONT DECOUPLE FROM HANDLE. NO DELAY IN SURGICAL TIME. NO ADVERSE IMPACT ON PATIENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18730513 ↗
  • MalfunctionFeb 12, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DRIVER AND DRIVER HANDLE ARE STUCK TOGETHER, NO ADVERSE EVENTS NO DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18691279 ↗
  • MalfunctionJan 18, 2024Device-Device Incompatibility; Device-Device Incompatibility

    BROKEN ATTUNE REVISION SCREWDRIVER AND HEX, STUCK IN FIXING WON¿T RELEASE NO DELAY ON SURGERY NOT BROKEN INTO TWO OR MORE PIECES NO IMPACT ON PATIENT LOT NUMBER UNKNOWN, NOT SUITABLE TO BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18534647 ↗
  • MalfunctionJan 12, 2024Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION RECEIVED: A. WAS THERE ANY SURGICAL DELAY RELATED TO THE EVENT? THIS WAS FOUND "AFTER THE CASE". THERE WERE NO DETAILS ABOUT THE OPERATIVE PORTION OF THE PROCEDURE BECAUSE THE SCREWDRIVER PERFORMED NORMALLY DURING THAT TIME. IT WAS ONLY AFTER THAT THERE WAS AN ISSUE. B. WAS THERE ANY PATIENT CONSEQUENCE RELATED TO THE EVENT? NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18501665 ↗
  • MalfunctionDec 13, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SCREWDRIVER SHAFT WILL NOT DISENGAGE FROM HANDLE.

    Patient problems: Insufficient Information

    MAUDE report 18317684 ↗
  • MalfunctionNov 1, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE TORQUE WRENCH AND HEX SHAFT ARE STUCK TOGETHER AND CANNOT GET APART.

    Patient problems: Insufficient Information

    MAUDE report 18046116 ↗
  • MalfunctionAug 9, 2023Break; Mechanical Jam; Naturally Worn; Break; Mechanical Jam; Naturally Worn

    ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE SCREWDRIVER SHAFT WAS STUCK INTO THE TORQUE HANDLE, ALSO THE SCREWDRIVER TIP HAS CROSS THREADED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17491962 ↗
  • MalfunctionMar 15, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT A SCREW THAT SHOULD COME APART WILL NOT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16547439 ↗
  • MalfunctionMar 14, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT THE SCREW DRIVER STUCK TOGETHER. IT WAS DAMAGED DURING SURGERY. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16539466 ↗
  • MalfunctionJun 2, 2021Material Deformation; Material Deformation

    ITEMS DAMAGED: THIS WAS DISCOVERED UPON OPENING THE DEMO SET PRIOR TO TRAINING. THE 4MM OFFSET GUIDE BECAME CROSS THREADED IN ONE OF THE FE GUIDES, THE HEX ATTACHMENT (1.3) WAS BENT AS THIS WAS ATTEMPTED TO BE UNSCREWED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11920507 ↗
  • MalfunctionMar 11, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THEY HAD A SUCCESSFUL ATTUNE REVISION IMPLANT SURGERY. WHEN I CAME BACK TO THE HOSPITAL THE NEXT DAY, TO PUT INSTRUMENTS BACK IN THEIR TRAYS, I WAS HAVING TROUBLE WITH ONE. I AM NOT SURE WHY OR HOW IT HAPPENED BUT THE HEX DRIVER IS STUCK IN THE TORQUE HANDLE. THEY CANNOT BE SEPARATED. THEY WERE WORKING CORRECTLY DURING THE PROCEDURE. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11465637 ↗
  • MalfunctionAug 28, 2020Device-Device Incompatibility; Material Deformation; Material Twisted/Bent; Device-Device Incompatibility; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE HEX HEAD OF SCREWDRIVERS ARE ROUNDED AND NOT HOLDING THE SCREWS WELL. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10466524 ↗
  • MalfunctionAug 28, 2020Device-Device Incompatibility; Material Deformation; Material Twisted/Bent; Device-Device Incompatibility; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE HEX HEAD OF SCREWDRIVERS ARE ROUNDED AND NOT HOLDING THE SCREWS WELL. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10466496 ↗
  • MalfunctionJul 24, 2020Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam

    IT WAS REPORTED THAT THE SCREWDRIVER AND HEX DRIVER ARE STUCK TOGETHER SOMEHOW AND WONT DISASSEMBLE. NO SURGICAL DELAY.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 10318361 ↗
  • MalfunctionMar 12, 2020Device Slipped; Device-Device Incompatibility; Material Twisted/Bent; Unintended Movement; Physical Resistance/Sticking; Device Slipped; Device-Device Incompatibility; Material Twisted/Bent; Unintended Movement; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE HEX DRIVER AND HANDLE WILL NOT SEPARATE. THE HEX DRIVER FREELY SPINS IN HANDLE, THE HANDLE IS STRIPPED OUT. BOTH INSTRUMENTS NEED RAPLACED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9827722 ↗
  • MalfunctionMar 2, 2020Difficult to Remove; Device-Device Incompatibility; Mechanical Jam; Difficult to Remove; Device-Device Incompatibility; Mechanical Jam

    IT WAS REPORTED THAT THE TORQUE SCREW DRIVER AND HANDLE BECAME LODGED TOGETHER AND WON¿T COME APART. IT DIDN¿T DELAY THE SURGERY. NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9775992 ↗
  • MalfunctionJul 20, 2018Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE SCREW INSIDE TIBIAL TRIAL STRIPPED OUT. HEAD TO SCREWDRIVER ROUNDED OFF. CAN'T GET SCREW OUT OF TIBIA.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7706007 ↗
  • MalfunctionJul 19, 2018Break; Break

    IT WAS REPORTED THAT THIS PART WERE BROKEN DURING A CADAVER LAB TRAINING ON (B)(6) 2018. RECEIVED BROKEN PART FROM WAREHOUSE ON 06/26/2018. THE 2 PARTS JAMMED TOGETHER AND CANNOT SEPARATE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7704690 ↗

Other DRIVER, PROSTHESIS devices

All DRIVER, PROSTHESIS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page