Skip to content

DELTA XTEND 2307-74-101Recalls & reported problems

Depuy (Ireland) · DELTA XTEND 145 RSA REAM GUIDE INSERTER · Class 1

All models of DELTA XTEND (194)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Delta Xtend reamer guide inserter is a surgical instrument used during reverse shoulder replacement. Many of the 82 reports describe the handle failing to attach or staying loose, and some describe the handle breaking. The reports include no injuries and no deaths, and reports fell from 14 in the previous 12 months to 5 in the last 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
82
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 82 adverse event reports that list this exact device (DI 10603295518792), received between Jan 5, 2021 and Mar 23, 2026 — classified as 82 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (52 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility52
  • Unintended Movement30
  • Break14
  • Material Deformation5
  • Structural Problem5
  • Difficult to Open or Close4
  • Naturally Worn2
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions66
  • Insufficient Information16

Recent adverse event reports

The 25 most recent of 82 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2026Break

    IT WAS REPORTED THAT THE BROACH HANDLE IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24666833 ↗
  • MalfunctionMar 2, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT HANDLE DOESN'T ATTACH TO THE REAMER POST GUIDES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24484729 ↗
  • MalfunctionJan 22, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT EPYPHISEAL REAMER HANDLE BECAME LOOSE. IT¿S NOT HOLDING THE GUIDE STEADY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24144142 ↗
  • MalfunctionNov 12, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THERE WAS NO TENSION ON THE LEVER WHEN CLAMPING ONTO THE REAMING GUIDE POST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23534808 ↗
  • MalfunctionNov 4, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE HANDLE ON THE XTND 145 RSA REAM GUID INSRTR DEVICE WAS LOOSE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23466132 ↗
  • MalfunctionOct 28, 2025Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement

    IT WAS REPORTED THAT DURING A PRIMARY REVERSE TOTAL SHOULDER, THE SCRUB TECH NOTED THAT THE PROXIMAL REAMING GUIDE HANDLE¿S LATCH WAS LOOSE SUCH THAT THE REAMING GUIDE WOULD NOT STAY ENGAGED IN THE HANDLE. THERE WERE NO BROKEN PIECES. THERE WERE NO DELAYS OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23402369 ↗
  • MalfunctionOct 21, 2025Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement

    IT WAS REPORTED THAT HANDLE HAS BECOME LOOSE AND DOES NOT HOLD THE EPIPHYSIS REAMER GUIDE STEADY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350673 ↗
  • MalfunctionAug 20, 2025Unintended Movement; Unintended Movement

    WHILE USING THE GLOBAL UNIT SYSTEM FOR A REVERSE TOTAL SHOULDER, THE REAMER GUIDE INSERTER WAS DISCOVERED TO BE NOT FUNCTIONING CORRECTLY AS THE LOCKING MECHANISM SEEMED LOOSE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22850274 ↗
  • MalfunctionAug 6, 2025Structural Problem; Structural Problem

    IT WAS REPORTED THAT EPIPHYSEAL BROACH HANDLE DID NOT LOCK ONTO BROACH. THERE WAS NO DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22731111 ↗
  • MalfunctionJun 25, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT HANDLE DOES NOT LATCH PROPERLY. IT DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332010 ↗
  • MalfunctionJun 20, 2025Device-Device Incompatibility; Device-Device Incompatibility

    ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT THE LATCH OF THE HANDLE APPEARED BROKEN AS THE HANDLE WOULD NOT STAY IN A SECURE POSITION WHEN COMPLETELY CLOSED. NO ADVERSE REACTIONS. INSTRUMENT WAS REMOVED AND ASSIGNED A COMPLAINT NUMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22296452 ↗
  • MalfunctionApr 21, 2025Break; Break

    IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING SHOULDER REPAIR. FACILITY REPORTS THAT THEY BREAK THIS INSTRUMENT IN NORMAL USE AT LEAST ONCE A MONTH.

    Patient problems: Insufficient Information

    MAUDE report 21874122 ↗
  • MalfunctionApr 21, 2025Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A SHOULDER REPAIR SURGERY AND THE INSTRUMENT WAS BROKEN.

    Patient problems: Insufficient Information

    MAUDE report 21871770 ↗
  • MalfunctionApr 3, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE REAMER GUIDE HANDLE WAS LOOSE. THE ISSUE DID NOT HAPPEN IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21763736 ↗
  • MalfunctionMar 28, 2025Difficult to Open or Close; Unintended Movement; Difficult to Open or Close; Unintended Movement

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A SHOULDER REPAIR SURGERY, AND EPIPHYSIS REAMER HANDLE FROM THE 145 EPI LATERAL TRAY WAS LOOSE AND DID NOT CLOSE PROPERLY. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 21719683 ↗
  • MalfunctionMar 17, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE HANDLE DOESN'T STAY LATCHED. THERE IS A RISK OF THE HANDLE OPENING AND DROPPING THE REAMER GUIDE. SURGICAL DELAY WAS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 21621334 ↗
  • MalfunctionMar 10, 2025Device-Device Incompatibility; Unintended Movement; Device-Device Incompatibility; Unintended Movement

    IT WAS REPORTED THAT THIS PATIENT HAS A LEFT TOTAL SHOULDER. IT WAS DISCOVERED AFTER REASSEMBLING THE INSTRUMENTS, THAT THE HUMERAL STEM ROTATION GUIDE BROACH HANDLE DID NOT HOLD THE TRIAL PROPERLY AND WAS LOOSE.

    Patient problems: Insufficient Information

    MAUDE report 21560597 ↗
  • MalfunctionJan 31, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE LOCKING MECHANISM ON THE REAM GUIDE INSERTER IS LOOSE AND POPS OPEN UNDER VERY LITTLE PRESSURE. THE PRODUCT DAMAGE DOCUMENTED HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE ARE NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21281838 ↗
  • MalfunctionNov 7, 2024Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL SHOULDER REPLACEMENT ON (B)(6) 2024. WHILE INSERTING THE REAMER GUIDE, SOMETHING IN THE REAMER GUIDE INSERTER BROKE AND THE GUIDE WOULD WOBBLE BACK AND FORTH. STAFF REMOVED REAMER GUIDE BY HAND. THERE WAS NO SIGNIFICANT DELAY. NOTHING WAS LEFT INSIDE OF THE PATIENT AND NO HARM WAS CAUSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20636278 ↗
  • MalfunctionOct 28, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT PRIOR TO THE CASE BEGINNING, BEFORE THE PATIENT WAS BROUGHT BACK TO THE OR IT WAS DISCOVERED THAT THE 145 DEGREE REAMER GUIDE INSERTER DID NOT RETAIN THE REAMER EPI GUIDE SUFFICIENTLY. A SECOND SET WAS OPENED FOR THE REPLACEMENT. THIS DID NOT IMPACT THE DURATION OF THE CASE AND THERE WAS IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20544539 ↗
  • MalfunctionOct 11, 2024Unintended Movement; Unintended Movement

    IT WAS REPORTED DURING A SHOULDER REPAIR ON (B)(6) 2024 THAT INSTRUMENT BECAME LOOSE WHEN TRYING TO CAPTURE THE EPI REAMER GUIDE.

    Patient problems: Insufficient Information

    MAUDE report 20428726 ↗
  • MalfunctionMay 28, 2024Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE EPI 145 REAMER HANDLE IS LOOSE.

    Patient problems: Insufficient Information

    MAUDE report 19404723 ↗
  • MalfunctionMay 28, 2024Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE EPI STEM GUIDE INSERTER WAS LOOSE.

    Patient problems: Insufficient Information

    MAUDE report 19404635 ↗
  • MalfunctionMar 15, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE145 RSA REAMER GUIDE HANDLE WAS LOOSE AND CAUSED DELAYS INTR-OP TO MAKE ATTEMPTS TO ENSURE PROPER CONTACT AND TECHNIQUE OF REAMER PLACEMENT WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? YES, IF YES, NUMBER OF MINUTES: 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18911472 ↗
  • MalfunctionFeb 8, 2024Device-Device Incompatibility; Naturally Worn; Device-Device Incompatibility; Naturally Worn

    IT WAS REPORTED THAT THE HANDLE TO BROACH CONNECTION APPEARS TO BE WORN SO WOULD NOT TIGHTEN COMPLETELY. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18669582 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page