Skip to content

EMPHASYS 4819-22-090Recalls & reported problems

Depuy (Ireland) · EMPHASYS MOBILE BEARING TRIAL HEAD EXTRACTOR 22.225-28mm · Class 1

All models of EMPHASYS (109)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295528340), received between Dec 17, 2025 and Sep 15, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Break

    It was reported that during a primary hip surgery the supervising ortho consultant was using a press to disassemble the dual mobility system. They used the piece intended for a 22.2/28mm head trial extraction to remove a 36mm head instead of the 36mm adaptor. During this process one of the small tabs that aligned with the grooves of the liner trial snapped off the construct. No patient or user harm was reported. There was no significant delay. No parts remained in situ of the of patient. The part broke over the working trolley. The small piece of plastic that broke off the instrument was disposed of by the surgeon at the time. The remainder of the part remained with SSD Hospital. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50774984 ↗
  • MalfunctionMay 19, 2026Break

    IT WAS REPORTED THAT DURING DECONTAMINATION AFTER SURGERY, IT WAS NOTICED THAT A PIECE HAD BROKE OFF OF THE TRIAL HEAD REMOVAL TOOL. THE DEVICE WAS NOT USED DURING THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25216362 ↗
  • MalfunctionApr 13, 2026Break; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT WHEN USING THE INSTRUMENT ,IT WAS DROPPED ON THE FLOOR AND ONE OF THE FINS OF THE HEAD EXTRACTOR BROKE OFF. ENDED UP DESTROYING THE HEAD TRIAL IN THE PROCESS OF TRYING TO CHANGE HEADS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24861202 ↗
  • MalfunctionDec 17, 2025Break; Break

    IT WAS REPORTED THAT THE BROKEN TONGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23831826 ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates