EMPHASYS 4819-22-090Recalls & reported problems
Depuy (Ireland) · EMPHASYS MOBILE BEARING TRIAL HEAD EXTRACTOR 22.225-28mm · Class 1
All models of EMPHASYS (109)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295528340), received between Dec 17, 2025 and Sep 15, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Improper or Incorrect Procedure or Method1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Break
It was reported that during a primary hip surgery the supervising ortho consultant was using a press to disassemble the dual mobility system. They used the piece intended for a 22.2/28mm head trial extraction to remove a 36mm head instead of the 36mm adaptor. During this process one of the small tabs that aligned with the grooves of the liner trial snapped off the construct. No patient or user harm was reported. There was no significant delay. No parts remained in situ of the of patient. The part broke over the working trolley. The small piece of plastic that broke off the instrument was disposed of by the surgeon at the time. The remainder of the part remained with SSD Hospital. All available information has been disclosed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50774984 ↗MalfunctionMay 19, 2026Break
IT WAS REPORTED THAT DURING DECONTAMINATION AFTER SURGERY, IT WAS NOTICED THAT A PIECE HAD BROKE OFF OF THE TRIAL HEAD REMOVAL TOOL. THE DEVICE WAS NOT USED DURING THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25216362 ↗MalfunctionApr 13, 2026Break; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT WHEN USING THE INSTRUMENT ,IT WAS DROPPED ON THE FLOOR AND ONE OF THE FINS OF THE HEAD EXTRACTOR BROKE OFF. ENDED UP DESTROYING THE HEAD TRIAL IN THE PROCESS OF TRYING TO CHANGE HEADS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24861202 ↗MalfunctionDec 17, 2025Break; Break
IT WAS REPORTED THAT THE BROKEN TONGS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23831826 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026