INHANCE 7000-30-100Recalls & reported problems
Depuy (Ireland) · INHANCE SHOULDER SYSTEM RESECTION GUIDE PIN (3 PACK) diameter 3mm x 100mm · Class 1
All models of INHANCE (202)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295538530), received between Jan 19, 2023 and Jan 20, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Device-Device Incompatibility3
- No Apparent Adverse Event2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 20, 2026Device-Device Incompatibility
IT WAS REPORTED THAT THE PIN STUCK IN RESECTION GUIDE. ANOTHER WAS AVAILABLE AND THE CASE PROCEEDED AS PLANNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24123540 ↗MalfunctionJun 12, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
ADDITIONAL INFORMATION RECEIVED. A. CAN YOU PLEASE CLARIFY WHAT DO YOU MEAN BY DAMAGED? WAS IT WORN, CRACKED, BROKEN INTO PIECES, BENT, STRIPPED, CROSS THREADED OR ANY DEVICE INTERACTION. WHEN THE PIN WAS INSERTED INTO THE RESECTION GUIDE UNDER POWER, IT CREATED METAL SPIRAL FRAGMENTS AS IF THE RESECTION GUIDE HOLE WAS TOO SMALL FOR THE PIN. THE PIN LOOKS MACHINED TO A SMALLER DIAMETER. C. CAN YOU PLEASE CONFIRM THE QUANTITY OF THE RETENTION PIN? 3 FROM THE SAME PACKAGE. - 1 DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17106889 ↗MalfunctionJun 12, 2023Break; No Apparent Adverse Event; Break; No Apparent Adverse Event
IT WAS REPORTED THAT THIN MACHINED STRANDS OF METAL WERE GETTING DISCHARGED FROM THE PIN - GUIDE INTERFACE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17106836 ↗MalfunctionJun 12, 2023Break; No Apparent Adverse Event; Break; No Apparent Adverse Event
IT WAS REPORTED THAT THIN MACHINED STRANDS OF METAL WERE GETTING DISCHARGED FROM THE PIN - GUIDE INTERFACE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17106829 ↗MalfunctionJan 19, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE RESECTION GUIDE PIN GOT STUCK (COLDWELDED) IN RESECTION SILHOUETTE GUIDE. UNKNOWN SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16209824 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page