Skip to content

INHANCE 7000-30-100Recalls & reported problems

Depuy (Ireland) · INHANCE SHOULDER SYSTEM RESECTION GUIDE PIN (3 PACK) diameter 3mm x 100mm · Class 1

All models of INHANCE (202)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10603295538530), received between Jan 19, 2023 and Jan 20, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Device-Device Incompatibility3
  • No Apparent Adverse Event2

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 20, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT THE PIN STUCK IN RESECTION GUIDE. ANOTHER WAS AVAILABLE AND THE CASE PROCEEDED AS PLANNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24123540 ↗
  • MalfunctionJun 12, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    ADDITIONAL INFORMATION RECEIVED. A. CAN YOU PLEASE CLARIFY WHAT DO YOU MEAN BY DAMAGED? WAS IT WORN, CRACKED, BROKEN INTO PIECES, BENT, STRIPPED, CROSS THREADED OR ANY DEVICE INTERACTION. WHEN THE PIN WAS INSERTED INTO THE RESECTION GUIDE UNDER POWER, IT CREATED METAL SPIRAL FRAGMENTS AS IF THE RESECTION GUIDE HOLE WAS TOO SMALL FOR THE PIN. THE PIN LOOKS MACHINED TO A SMALLER DIAMETER. C. CAN YOU PLEASE CONFIRM THE QUANTITY OF THE RETENTION PIN? 3 FROM THE SAME PACKAGE. - 1 DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17106889 ↗
  • MalfunctionJun 12, 2023Break; No Apparent Adverse Event; Break; No Apparent Adverse Event

    IT WAS REPORTED THAT THIN MACHINED STRANDS OF METAL WERE GETTING DISCHARGED FROM THE PIN - GUIDE INTERFACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17106836 ↗
  • MalfunctionJun 12, 2023Break; No Apparent Adverse Event; Break; No Apparent Adverse Event

    IT WAS REPORTED THAT THIN MACHINED STRANDS OF METAL WERE GETTING DISCHARGED FROM THE PIN - GUIDE INTERFACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17106829 ↗
  • MalfunctionJan 19, 2023Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE RESECTION GUIDE PIN GOT STUCK (COLDWELDED) IN RESECTION SILHOUETTE GUIDE. UNKNOWN SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16209824 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page