INHANCE 7205-30-301Recalls & reported problems
Depuy (Ireland) · INHANCE SHOULDER SYSTEM SIZER HANDLE · Class 1
All models of INHANCE (202)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10603295538592), received between Apr 24, 2023 and Jun 2, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Device-Device Incompatibility2
- Material Deformation2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 2, 2026Device-Device Incompatibility
IT WAS REPORTED THAT THE GLENOID PIN WOULD NOT EASILY SLIDE THROUGH THE SIZER HANDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25365760 ↗MalfunctionJun 2, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE PIECES GOT STUCK TOGETHER. THE PIN WOULD NOT PULL OUT OF SIZER HANDLE AFTER ATTEMPTING TO PUT IT INTO GLENOID THROUGH GUIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22122669 ↗MalfunctionApr 24, 2023Break; Material Deformation; Break; Material Deformation
THE SIZER HANDLE FOR THE ENHANCE SHOULDER SYSTEM HAD A BENT TAB ON THE CURVED END OF THE DEVICE AND WHEN THE HUMERAL AND GLENOID PINS WERE INSERTED DURING THE PROCEDURE, PLASTIC SHARDS WERE SEEN COMING OFF THE DEVICE. THE CONCERN WAS COULD THESE PLASTIC SHARDS/DEBRIS GET INTO THE OPEN SHOULDER JOINT AND NOT BE FOUND OR REMOVED THROUGH IRRIGATION THAT COULD CAUSE POST OPERATIVE COMPLICATIONS. THERE WAS A THREE MINUTE SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16796365 ↗MalfunctionApr 24, 2023Break; Material Deformation; Break; Material Deformation
THE SIZER HANDLE FOR THE INHANCE SHOULDER SYSTEM HAD A BENT TAB ON THE CURVED END OF THE DEVICE AND WHEN THE HUMERAL AND GLENOID PINS WERE INSERTED DURING THE PROCEDURE, PLASTIC SHARDS WERE SEEN COMING OFF THE DEVICE. THE CONCERN WAS COULD THESE PLASTIC SHARDS/DEBRIS GET INTO THE OPEN SHOULDER JOINT AND NOT BE FOUND OR REMOVED THROUGH IRRIGATION THAT COULD CAUSE POST OPERATIVE COMPLICATIONS. THERE WAS A TWO MINUTE SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16796300 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page