INHANCE 6205-60-050Recalls & reported problems
Depuy (Ireland) · INHANCE SHOULDER SYSTEM ONE STEP PREP GLENOID REAMER - BOWTIE LARGE · Class 1
All models of INHANCE (202)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10603295539506), received between Jul 10, 2024 and Nov 14, 2025 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Failure to Cut4
- Entrapment of Device1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Insufficient Information2
- Foreign Body In Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 14, 2025Break; Break
IT WAS REPORTED THAT A TOOTH FROM A LARGE GLENOID BOW-TIE REAMER BROKE OFF WHILE PREPARING THE GLENOID BASEPLATE. ALL PIECES WERE RETRIEVED, SURGICAL TIME WAS NOT PROLONGED, AND THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23556575 ↗InjuryOct 28, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT PATIENT HAD HARD BONE AND WAS DIFFICULT REAMING. GLENOID WAS IMPLANTED. UPON X-RAY THERE WAS FOREIGN OBJECTS SURROUNDING THE GLENOID POST. THEY INSPECTED THE INSTRUMENTS AND NOTICED THAT THE TIPS OF THE CENTER DRILL FROM THE LG DRILL WERE SHEERED OFF. THEY HAVE BROKE DURING THE TEAMING PROCESS. AN ATTEMPT TO REMOVE THE BROKEN TIPS WAS NOT PERFORMED BECAUSE THE FINAL IMPLANT HAD BEEN INSERTED.
Patient problems: Foreign Body In Patient
MAUDE report 23401585 ↗MalfunctionSep 26, 2025Break; Break
IT WAS REPORTED THAT THE REAMER WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23156416 ↗MalfunctionMay 27, 2025Break; Break
IT WAS REPORTED THAT THE INSTRUMENTATION BROKE FROM THE SET. IT WAS NOTICED BEFORE THE CASE. A NEW SET WAS BROUGHT TO NOT DELAY THE CASE. THERE WAS NO DELAY. NO BROKEN PIECES LEFT ON FIELD OR PATIENT. NO SURGEON COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22088724 ↗MalfunctionMay 22, 2025Break; Break
IT WAS REPORTED THAT THE CUTTING TEETH OF THE CENTRAL PART OF THE LARGE BOWTIE REAMER BROKE OFF WHILE REAMING THE GLENOID. ALL PIECES WERE REMOVED. IT DID NOT NEGATIVELY AFFECT THE CASE AND THERE WAS NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22069374 ↗MalfunctionMay 20, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE GLENOID REAMER METAL TWISTED. NO HARM WAS DONE TO PATIENT, NO DELAY HAPPENED AS BACKUP TRAYS WERE AVAILABLE; NO BROKEN PIECES WERE LEFT IN PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22054815 ↗MalfunctionJan 15, 2025Break; Break
IT WAS REPORTED THAT THE TIP OF THE LARGE BOWTIE GLENOID REAMER BROKE WHILE REAMING THE GLENOID. FRAGMENTS WERE GENERATED AND WERE RETRIEVED. THE LARGE BLAZER FUSED ONTO THE CANNULATED HANDLE. THERE WAS NO MISHANDLING OF THE PRODUCT BEFORE OR AFTER ASSEMBLING THE TWO ITEMS. THERE WAS ABOUT A 20 MINUTE TIME DELAY TO TRY TO REMOVE THE BLAZER FROM THE CANNULATED HANDLE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21158356 ↗MalfunctionOct 30, 2024Break; Break
IT WAS REPORTED THAT THE TIP OF THE LARGE BOW TIE REAMER BROKE DURING SURGERY. IT WAS UNKNOWN IF THERE WAS ANY SURGICAL DELAY.
Patient problems: Insufficient Information
MAUDE report 20565383 ↗MalfunctionAug 13, 2024Break; Break
IT WAS REPORTED THAT THE TIP OF THE GLENOID REAMER BROKE INTO PIECES. ALL PIECES WHERE REMOVED FROM THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 19972219 ↗MalfunctionJul 10, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE INHANCE BOW TIE REAMERS WERE DULL. THERE WAS NO CASE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19716845 ↗MalfunctionJul 10, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE ENHANCE BOW TIE REAMERS WERE DULL. THERE WAS NO CASE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19716841 ↗MalfunctionJul 10, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE ENHANCE BOW TIE REAMERS WERE DULL. THERE WAS NO CASE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19716840 ↗MalfunctionJul 10, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT THE INHANCE BOW TIE REAMERS WERE DULL. THERE WAS NO CASE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19716012 ↗
Other DRIVER, PROSTHESIS devices
- ATTUNE 2544-01-005Depuy (Ireland) · DRIVER, PROSTHESIS1,383 reports
- EXPEDIUM VERSE 299704230Medos International Sàrl · DRIVER, PROSTHESIS875 reports
- NA 9200-10-029Depuy (Ireland) · DRIVER, PROSTHESIS745 reports
- MOSS 279712600Depuy Spine, LLC · DRIVER, PROSTHESIS706 reports
- NA 2598-07-440Depuy (Ireland) · DRIVER, PROSTHESIS665 reports
- NA 2598-07-570Depuy (Ireland) · DRIVER, PROSTHESIS505 reports
- VIPER 286745550Depuy Spine, LLC · DRIVER, PROSTHESIS263 reports
- SUMMIT 2570-05-100Depuy (Ireland) · DRIVER, PROSTHESIS225 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026