INTUITION 2525-30-200Recalls & reported problems
Depuy (Ireland) · INTUITION FOR HP PARTIAL FEMORAL FINISHING BLOCK SIZE - LEFT MEDIAL/RIGHT LATERAL 2 · Class 1
All models of INTUITION (23)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10603295544531), received between May 5, 2025 and May 5, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Incorrect Measurement” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2025Incorrect Measurement; No Apparent Adverse Event; Incorrect Measurement; No Apparent Adverse Event
IT WAS REPORTED THAT ON APRIL 11, 2025, THE PERSON IN CHARGE CHECKED THE "PRODUCTS AND" FOUND THAT THERE WERE DIFFERENCES BETWEEN THE PRODUCTS AND THE CAD MODELS. SPECIFICALLY, THE DIFFERENCES ARE AS FOLLOWS: CHISEL ACTUALLY GOES APPROXIMATELY 1MM DEEPER THAN IT¿S OBSERVED IN NX MODELS FOR ALL FINISHING GUIDES LM SZ 1-4 ON MY END. SIZE 2 LM SHOWN BELOW. NO SURGICAL DELAY. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21963460 ↗
Other INSTRUMENT, CUTTING, ORTHOPEDIC devices
- SIGMA LCS 9505-01-230Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC95 reports
- ATTUNE 2544-00-522Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC58 reports
- SIGMA LCS 9505-01-229Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC56 reports
- SIGMA 2024-40-100Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC43 reports
- ATTUNE 2545-00-158Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC41 reports
- ATTUNE 2545-00-156Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC41 reports
- SIGMA 2024-40-200Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC40 reports
- ATTUNE 2545-00-155Depuy (Ireland) · INSTRUMENT, CUTTING, ORTHOPEDIC39 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page