Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10603295559627), received between Dec 19, 2025 and Mar 31, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Adverse Event Without Identified Device or Use Problem1
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Bone Fracture(s)1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DURING THE HUMERAL PREPARATION USING THE INTACT INHANCE INSTRUMENTATION SURGEON ADVANCED THE HUMERAL DRIVE SHAFT AND SWITCHED TO POWER ON REAM SETTING TO PREP THE HUMERUS. THE REAMER WOULD NOT ROTATE WHICH RESULTED IN A BENDING OF THE HUMERAL DRIVE SHAFT. THE HUMERAL REAMER BECAME STUCK IN THE HUMERAL HANDLE ALSO. SURGEON THEN SWITCHED TO THE NORMAL INHANCE TECHNIQUE AND COMPLETED THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742924 ↗
  • MalfunctionJan 20, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT SURGEON COULD NOT GET HUMERAL TEAMING SHAFT TO ALIGN WITH THE REAMER AND BEARING THRU THE HUMERAL FRAME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24117919 ↗
  • InjuryDec 19, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING IMPLANTATION THE STEMLESS COMPONENT ON THE HUMERAL FRAME AND HANDLE ON THE STEMLESS COMPONENT FRACTURED THE ANTERIOR HUMERUS LEADING TO USE OF A STEMMED COMPONENT.

    Patient problems: Bone Fracture(s); No Clinical Signs, Symptoms or Conditions

    MAUDE report 23853922 ↗