Skip to content

CADD 21-7390-24Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of CADD (123)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10610586027505), received between Jul 19, 2019 and Jul 22, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K040636); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 9, 2022Class IOpen, Classified

    There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms

    Z-0963-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Connection Problem1
  • Improper Flow or Infusion1
  • Leak/Splash1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Information2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2024Break; Leak/Splash; Use of Device Problem; Break; Leak/Splash; Use of Device Problem

    IT WAS REPORTED THAT PCA TUBING WAS FOUND TO BE LEAKING AT CONNECTION SITE WITH IV FLUIDS. THE ORIGINAL INTENDED PROCEDURE WAS IV INFUSION. THE DEVICE HAD FAILED. IT BROKE AND THE CUSTOMER COULD NOT GET IT TO WORK, OR THE DEVICE HAD STOPPED WORKING. THE DEVICE DID NOT DO WHAT IT WAS SUPPOSED TO DO. THE DEVICE WAS HARD TO USE. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19801321 ↗
  • MalfunctionJun 24, 2024

    PCA TUBING FOUND TO BE LEAKING AT CONNECTION SITE WITH IV FLUIDS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19593525 ↗
  • MalfunctionJul 19, 2019Improper Flow or Infusion; Improper Flow or Infusion

    INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL CADD ADMINISTRATION TUBING WAS DISCONNECTED AT POINT OF CONNECTION SITE PROXIMAL TO THE BLUE/YELLOW LOCK. THE TUBING CONNECTION SITE WAS FOUND TO BE INTACT UPON REGISTERED NURSE ASSESSMENT AND SITE SPONTANEOUSLY DISCONNECTED ON PATIENT. NO ADVERSE EFFECTS REPORTED.

    Patient problems: No Information

    MAUDE report 8809811 ↗
  • MalfunctionJul 19, 2019Connection Problem; Connection Problem

    INFORMATION WAS RECEIVED INDICATING THAT THE REGISTERED NURSE FOUND THE PATIENT'S EPIDURAL TUBING DISCONNECTED AT THE CONNECTION SITE. THE SITE WAS APPROXIMATELY DISCONNECTED FOR AN HOUR. THERE WAS NO REPORTED INJURY TO THE PATIENT.

    Patient problems: No Information

    MAUDE report 8806927 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K040636. One submission often covers many models, so check the product and lot details in each record.

  • Z-2522-2025 · Aug 6, 2025 · Open, Classified
    Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper included. 50mL, Product Code 21-7001-24
  • Z-1767-2017 · Mar 17, 2017 · Terminated
    CADD Administration Set, 21-7339-24, Rx Only, Sterile EO.
  • Z-1831-2016 · Apr 22, 2016 · Terminated
    CADD Administration Sets with Flow Stop, Product Reorder No: 21-7321-01, 21-7321-24, 21-7322-01, 21-7322-24, 21-7323-24, 21-7324-01, 21-7324-24, 21-7333-24, 21-7336-01, 21-7336-24
  • Z-1286-2015 · Feb 2, 2015 · Terminated
    CADD" Medication Cassette Reservoir, 50mL, REF 21-7001-24 & 21-7301-24, Sterile EO, Rx only.
  • Z-0879-2008 · Dec 14, 2007 · Terminated
    Smiths CADD Yellow Medication Cassette Reservoirs with Flow Stop, Clamp and Female Luer, 100 ml, Reorder No. 21-7300-24, Nonvented stopper included, Sterile, Smiths Medical, St. Pa

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates