CADD recallZ-0963-2023 · Smiths Medical ASD Inc. · initiated Dec 9, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There are two potential issues with CADD Infusion System Infusion Sets: (1) Lack of Delivery or Underdelivery related to Tubing Occlusion and (2) False "No Disposable Attached (NDA)" Alarms
FDA root cause category: Under Investigation by firm
What the company is doing
The firm issued an URGENT MEDICAL DEVICE CORRECTION notice on 12/09/2022 by mail. Actions for Clinicians and Patients: . The pump may not adequately detect the cassette before or during an infusion due to this issue, and an alarm will be triggered. If a pump displays an NDA alarm, the user can try to resolve the alarm by repositioning the CADD Medication Cassette Reservoir while connected to the pump, repositioning the reservoir by disconnecting from the pump and reattaching it to the pump, or replacing the reservoir. . Or the user can remove the reservoir from the pump and push the plastic ridge highlighted in the circle towards the arch on the reservoir. . For patients requiring life sustaining medications, we recommend use of alternative CADD infusion sets. Contact pharmacy to discuss availability of alternative CADD infusion sets. Due to availability and specific patient situations, clinicians may consider switching patients to an alternative pump. . If the user cannot resolve the NDA alarm, replace the cassette reservoir, though the issues may recur if that product is also affected by this recall. . Clinicians, share this letter with your homecare patients and educate them about the actions above. Complete and return the Response Form via fax at 866-338-4443 or email to [email protected]. . DISTRIBUTORS: Immediately forward this notice to your customers them. Have them contact Sedgwick 866-562-5931(M-F, 8am-5pm ET) to obtain a response form. Global Complaint Management, [email protected] 1-(866)-216-8806, Report adverse events or product complaints: Technical Assistance (800)-258-5361 . Pharmacists: . Immediately identify affected products in your possession. Separate and label them to avoid use for life sustaining medications. . For use with life sustaining medications, contact Smiths Medical customer service (800)-258-5361 for info on obtaining alternative CADD infusion sets. On 06/12/2023, another notification
Recall details
- Recall number
- Z-0963-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 9, 2022
- Posted by the FDA
- Jan 30, 2023
- Recalling firm
- Smiths Medical ASD Inc.
- Quantity
- 45285351 units
- Product code
- FPA
- FDA submissions
- K040636, K170982
- Distribution
- Worldwide Distribution: US (nationwide) and OUS countries of: CH, CA, ES, FR, AT, BE, IT, DK, NO, EE, DE, NL, AU, GB, SE, PL, SA, IN, IL, PT, CZ, IE, NZ, LU, SI, ZA, BM, SG, MQ, CL, LB, MY, HU, JP, TW, and CN
Product description
CADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24 3) REF 21-7302-24, 4) REF 21-7321-24, 5) REF 21-7322-24, 6) REF 21-7323-24, 7) REF 21-7324-24, 8) REF 21-7333-24, 9) REF 21-7336-24, 10) REF 21-7339-24, 11) REF 21-7343-24, 12) REF 21-7349-24, 13) REF 21-7359-24, 14) REF 21-7383-24, 15) REF 21-7390-24, 16) REF 21-7391-24, 17) REF 21-7394-24, 18) REF 21-7395-24, 19) REF 21-7301-24JP, 20) REF 21-7302-24JP, 21) REF 21-7321-24JP, 22) REF 21-7322-24JP, 23) REF 21-7359-24JP, 24) REF 21-7394-24JP
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
1) REF 21-7300-24, UDI/DI 30610586027196, Lot numbers: 3630772 through 4334107; 2) REF 21-7301-24, UDI/DI 30610586027219, Lot numbers: 3630747 through 4334119; 3) REF 21-7302-24, UDI/DI 30610586027233, Lot numbers: 3617363 through 4334138; 4) REF 21-7321-24, UDI/DI 30610586029633, Lot numbers: 3773534 through 4308545; 5) REF 21-7322-24, UDI/DI 30610586029640, Lot numbers: 3776375 through 4334324; 6) REF 21-7323-24, UDI/DI 061058603961, Lot numbers: 3776373 through 4334325; 7) REF 21-7324-24, UDI/DI 30610586029657, Lot numbers: 3773527 through 4334328; 8) REF 21-7333-24, UDI/DI 30610586043233, Lot numbers: 3776362 through 3984144; 9) REF 21-7336-24, UDI/DI 30610586027295, Lot numbers: 3776360 through 4025381; 10) REF 21-7339-24, UDI/DI 30610586027301, Lot numbers: 3780565 through 4009665; 11) REF 21-7343-24, UDI/DI 35019517191824, Lot numbers: 3965344 through 4334338; 12) REF 21-7349-24, UDI/DI 35019517184239, Lot numbers: 3926579 through 4339160; 13) REF 21-7359-24, UDI/DI 30610586027400, Lot numbers: 3776315 through 4334339; 14) REF 21-7383-24, UDI/DI 30610586027479, Lot numbers: 3780549 through 3971523; 15) REF 21-7390-24, UDI/DI 30610586027509, Lot numbers: 3780548 through 4334346; 16) REF 21-7391-24, UDI/DI 30610586027516, Lot numbers: 3773276 through 4334351; 17) REF 21-7394-24, UDI/DI 30610586027530, Lot numbers: 3774739 through 4339178; 18) REF 21-7395-24, UDI/DI 30610586027547, Lot numbers: 3808536 through 4308562 19) REF 21-7301-24JP, UDI/DI N/A , Lot numbers: 3630747 through 4334113; 20) REF 21-7302-24JP, UDI/DI N/A , Lot numbers: 3617363 through 4329630; 21) REF 21-7321-24JP, UDI/DI N/A , Lot numbers: 3773534 through 4308545; 22) REF 21-7322-24JP, UDI/DI N/A , Lot numbers: 3776375 through 4334324; 23) REF 21-7359-24JP, UDI/DI N/A , Lot numbers: 3776315 through 4334339; 24) REF 21-7394-24JP, UDI/DI N/A , Lot numbers: 3774739 through 4339178.
Devices named in this recall
17 devices in the FDA device registry whose identifier appears in the recall's code information.
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- CADD 21-7394-24Icu Medical, Inc. · Set, administration, intravascular2 recalls2 deaths586 reports
- CADD 21-7301-24Icu Medical, Inc. · Set, administration, intravascular4 recalls467 reports
- CADD 21-7322-24Icu Medical, Inc. · Set, administration, intravascular2 recalls391 reports
- CADD 21-7300-24Icu Medical, Inc. · Set, administration, intravascular4 recalls121 reports
- CADD 21-7324-24Icu Medical, Inc. · Set, administration, intravascular2 recalls155 reports
- CADD 21-7349-24Icu Medical, Inc. · Set, administration, intravascular2 recalls135 reports
- CADD 21-7359-24Icu Medical, Inc. · Set, administration, intravascular2 recalls110 reports
- CADD 21-7391-24Icu Medical, Inc. · Set, administration, intravascular2 recalls77 reports
- CADD 21-7343-24Icu Medical, Inc. · Set, administration, intravascular2 recalls65 reports
- CADD 21-7383-24Icu Medical, Inc. · Set, administration, intravascular2 recalls63 reports
- CADD 21-7333-24Icu Medical, Inc. · Set, administration, intravascular2 recalls24 reports
- CADD 21-7321-24Icu Medical, Inc. · Set, administration, intravascular1 recall45 reports
- CADD 21-7336-24Icu Medical, Inc. · Set, administration, intravascular1 recall30 reports
- CADD 21-7339-24Icu Medical, Inc. · Set, administration, intravascular1 recall25 reports
- CADD 21-7395-24Icu Medical, Inc. · Set, administration, intravascular1 recall5 reports
- CADD 21-7390-24Icu Medical, Inc. · Set, administration, intravascular1 recall4 reports
Product lines
Other devices under the same FDA submission
18 devices cleared under K040636, K170982 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CADD 21-2111-0402-51Icu Medical, Inc. · Pump, infusion, pca4 recalls1,535 reports
- CADD 21-2111-0401-51Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION4 recalls461 reports
- CADD 21-2112-0402-51Icu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery3 recalls469 reports
- CADD 21-2112-0402-51LIcu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery3 recalls129 reports
- CADD 21-2111-0402-51LIcu Medical, Inc. · Pump, infusion, pca3 recalls127 reports
- CADD 2110Icu Medical, Inc. · Pump, infusion, pca4 recalls4 reports
- CADD 2110Icu Medical, Inc. · Pump, infusion, pca4 recalls0 reports
- CADD 21-2111-0402-78Icu Medical, Inc. · Neuraxial Administration Set - Intrathecal delivery3 recalls5 reports
- CADD 21-2112-0402-78Icu Medical, Inc. · ACCESSORIES, PUMP, INFUSION3 recalls0 reports
- CADD 21-2112-0401-51Icu Medical, Inc. · Pump, infusion, pca1 recall97 reports
- CADD 2110Icu Medical, Inc. · Pump, infusion, pca2 recalls1 report
- CADD 21-7323-24Icu Medical, Inc. · Set, administration, intravascular1 recall21 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91392).
- Z-0962-2023 · Open, ClassifiedCADD Administration Set, List Numbers: a) REF 21-7346-24; b) REF 21-7363-24
- Z-0964-2023 · Open, ClassifiedCADD Administration Set, List Numbers: a) REF 21-7308-24, b) REF 21-7309-24, c) REF 21-7310-24, d) REF 21-7308-24JP, e) REF 21-7309-24JP. Computerized Ambulatory Delivery De…
- Z-0965-2023 · Open, ClassifiedCADD Administration Set, List Numbers: a) REF 21-7600-24, b) REF 21-7609-24, c) REF 21-7624-24, d) REF 21-7649-24, e) REF 21-7600-24JP, f) REF 21-7609-24JP, g) REF 21-7624-2…
- Z-0966-2023 · Open, ClassifiedCADD Administration Set, List Numbers: 1) REF 21-7300-24, 2) REF 21-7301-24, 3) REF 21-7302-24, 4) REF 21-7308-24, 5) REF 21-7309-24, 6) REF 21-7310-24, 7) REF 21-7322-24, …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page