Pneupac 122002Recalls & reported problems

Icu Medical, Inc. · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10610586046056), received between Dec 1, 2019 and Aug 21, 2025 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Flow or Under Infusion3
  • Loose or Intermittent Connection2
  • Improper Flow or Infusion1
  • Increase in Pressure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Hypoventilation1
  • Low Oxygen Saturation1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 21, 2025Loose or Intermittent Connection; Insufficient Flow or Under Infusion; Loose or Intermittent Connection; Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT DURING TRANSPORT OF A CARDIAC PATIENT, THEY CONNECTED THE RESPIRATORY CIRCUIT TO THE ENDOTRACHEAL TUBE (ETT) AND NOTICED THE PATIENT WAS NOT RECEIVING ADEQUATE BREATHS. UPON INSPECTION, THEY FOUND THAT THE RUBBER BAND AROUND THE POSITIVE END-EXPIRATORY PRESSURE (PEEP) ADAPTER WAS LOOSE AND OUT OF PLACE. THEY TESTED FOUR DIFFERENT CIRCUITS BEFORE IDENTIFYING ONE THAT FUNCTIONED PROPERLY AND ALLOWED THE PATIENT TO BREATH EFFECTIVELY. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22864536 ↗
  • MalfunctionAug 21, 2025Loose or Intermittent Connection; Insufficient Flow or Under Infusion; Loose or Intermittent Connection; Insufficient Flow or Under Infusion

    IT WAS REPORTED THAT DURING TRANSPORT OF A CARDIAC PATIENT, THEY CONNECTED THE RESPIRATORY CIRCUIT TO THE ENDOTRACHEAL TUBE (ETT) AND NOTICED THE PATIENT WAS NOT RECEIVING ADEQUATE BREATHS. UPON INSPECTION, THEY FOUND THAT THE RUBBER BAND AROUND THE POSITIVE END-EXPIRATORY PRESSURE (PEEP) ADAPTER WAS LOOSE AND OUT OF PLACE. THEY TESTED FOUR DIFFERENT CIRCUITS BEFORE IDENTIFYING ONE THAT FUNCTIONED PROPERLY AND ALLOWED THE PATIENT TO BREATH EFFECTIVELY. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22861459 ↗
  • InjuryJan 5, 2021Insufficient Flow or Under Infusion; Improper Flow or Infusion; Insufficient Flow or Under Infusion; Improper Flow or Infusion

    IT WAS REPORTED THAT MRI TRANSPORT VENTILATOR CIRCUIT BECAME DISCONNECTED IN PACKAGE AND WAS REASSEMBLED INCORRECTLY AT THE TIME OF USE. WHEN INCORRECTLY ASSEMBLED, THE CIRCUIT WAS ATTACHED TO THE PATIENT CONNECTION AND THE PATIENT WAS CONNECTED WHERE THE CIRCUIT SHOULD BE, RESULTING IN BREATHS BEING DELIVERED OUT THE EXHALATION PORT. THE PATIENT REQUIRED BRIEF RESUSCITATION MEASURES. THE PATIENT RECOVERED WITHOUT FURTHER INCIDENT.

    Patient problems: Low Oxygen Saturation

    MAUDE report 11119704 ↗
  • InjuryDec 1, 2019Increase in Pressure; Increase in Pressure

    IT WAS REPORTED THAT THE PATIENT WAS PLACED ON AN MRI VENTILATOR AND, THE VENTILATOR STARTED PEAK PRESSURING RIGHT AWAY. THE OPERATOR ATTEMPTED TO TROUBLESHOOT THE VENTILATOR BUT COULD NOT FIND THE SOURCE OF THE HIGH PRESSURE. IT WAS FURTHER REPORTED THAT THE PATIENT WAS RETURNED TO THE ROOM AND VENTILATED WITH AN AMU BAG. THE PATIENT WAS MONITORED USING A TRANSPORT MONITOR. INFORMATION PERTAINING TO THE CURRENT HEALTH STATUS OF THE PATIENT, OR IF THE ISSUE WAS ONGOING OR RESOLVED, WAS UNKNOWN.

    Patient problems: Hypoventilation

    MAUDE report 9392897 ↗

Other Ventilator, emergency, powered (resuscitator) devices

All Ventilator, emergency, powered (resuscitator) devices →