Pneupac 122003Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Pneupac (23)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10610586046063), received between Mar 14, 2017 and Mar 14, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Connection Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Connection Problem1
Patient problems
- Bradycardia1
- Hyperventilation1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 14, 2017Connection Problem; Connection Problem
IT WAS REPORTED THAT A PNEUPAC¿ PATIENT CIRCUIT WAS IN USE WITH A VENTILATED PATIENT ON A PORTABLE VENTILATOR WHEN IT WAS OBSERVED THAT THE CIRCUITRY TUBING AND THE PORTABLE VENTILATOR WERE CONNECTED INCORRECTLY. THE DEVICE ISSUE OCCURRED IN A HOSPITAL DURING A COMPUTERIZED TOMOGRAPHY (CT) SCAN. IT WAS NOTED THAT IT WAS UNCLEAR WHICH END OF THE TUBING SHOULD BE CONNECTED TO A GIVEN VALVE. THE DEVICE HAD TO BE DISCONNECTED, WHICH CAUSED THE PATIENT TO DECOMPENSATE DURING TRANSPORT. THE PATIENT BEGAN "'GASPING' FOR BREATHS" AND DEVELOPED BRADYCARDIA OF 30-40BPM. THE PATIENT WAS "MANUALLY [BAGGED]" AND ALSO PLACED ON A 100% FIO2 PORTABLE VENTILATOR AND RETURNED TO AN INTENSIVE CARE UNIT ROOM, WHERE THE PATIENT WAS RECONNECTED TO A VENTILATOR. NO PERMANENT INJURY WAS REPORTED. THE EVENT WAS CONSIDERED RESOLVED.
Patient problems: Bradycardia; Hyperventilation
MAUDE report 6403560 ↗
Other Ventilator, emergency, powered (resuscitator) devices
- Pneupac P310NUSIcu Medical, Inc. · Ventilator, emergency, powered (resuscitator)4 deaths1,035 reports
- Pneupac 120003Icu Medical, Inc. · Ventilator, emergency, powered (resuscitator)2 deaths523 reports
- Neo-Tee T-Piece Resuscitator 10-50805Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)3 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50811Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls3 reports
- Neo-Tee T-Piece Resuscitator 10-50809Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50810Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50832Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50808Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page