Skip to content

McKesson 16-N2558SRecalls & reported problems

Mckesson Medical-Surgical Inc. · NEEDLE, SAFETY 25GX5/8" (HINGED STYLE) · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10612479244655), received between Mar 27, 2025 and Mar 13, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Contamination1
  • Decoupling1
  • Fail-Safe Problem1
  • Material Twisted/Bent1

Patient problems

  • Needle Stick/Puncture2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2026Contamination

    CUSTOMER REPORTED THAT THE NEEDLES ARE LEAVING A GRAY/SILVER METALLIC RESIDUE ON PATIENTS' SKIN. PRESENT BEFORE VIAL STOPPER PUNCTURE. NOT OBSERVED IN MOST OTHER LOTS OF SAME PRODUCT. REPRODUCIBLE ACROSS OBSERVERS AND SITES. CERTAIN NEEDLE LOTS EXHIBIT VISIBLE GRAY/DARK RESIDUE DIRECTLY OUT OF UNOPENED STERILE PACKAGING. WHEN WIPED WITH STERILE WHITE GAUZE OR TISSUE, A GRAY SMEAR IS PRODUCED. WHEN USED FOR INJECTION (SIMULATION AND PATIENT CARE), A VISIBLE DARK MARK IS AT THE INJECTION PATH. WHEN REPRODUCING IN ONE MEDIA WE WERE ABLE TO CUT INTO, THE DARK MARK TRACKED THROUGH THE MEDIA ALONG THE INJECTION PATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24589491 ↗
  • MalfunctionFeb 17, 2026Break; Decoupling

    CUSTOMER REPORTED THAT ,ON OCCASION THE NEEDLE COMES OFF (DOES NOT BREAK) AND IN ONE INCIDENT STAYS STUCK IN THE PATIENT. NO INJURY OCCURRED. ON OTHER OCCASIONS, THE NEEDLE HAS BROKEN AT THE HUB ALSO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24363705 ↗
  • MalfunctionMay 16, 2025Material Twisted/Bent; Material Twisted/Bent

    CUSTOMER REPORTED RETURNING THE NEEDLE DUE TO MALFUNCTION IN THE SAFETY FEATURE, WHICH DOES NOT CLOSE PROPERLY. ON (B)(6) 2025, THE CUSTOMER RESPONDED TO THE PROBING QUESTIONS AND CONFIRMED THAT THIS ISSUE WAS OBSERVED WITH A FEW NEEDLES, RESULTING IN AN EMPLOYEE INJURY. THEY FURTHER EXPLAINED THAT WHEN CLOSING OF THE SAFETY DEVICE, THE NEEDLE TENDS TO BEND AND PIERCE THROUGH THE ORANGE SAFETY COVER. UPDATE: THE CUSTOMER CONFIRMED THAT WHEN CLOSING OF THE SAFETY DEVICE, THE NEEDLE TENDS TO BEND AND PIERCE THROUGH THE ORANGE SAFETY COVER. THERE WAS ONE INCIDENT WHERE THIS RESULTED IN A NEEDLESTICK TO AN EMPLOYEE. NO TREATMENT WAS REQUIRED.

    Patient problems: Needle Stick/Puncture

    MAUDE report 22035481 ↗
  • InjuryMar 27, 2025Fail-Safe Problem; Fail-Safe Problem

    THE CUSTOMER REPORTED THAT THE NEEDLES EXPERIENCED A MALFUNCTION IN THE SAFETY FEATURE, WHICH DOES NOT CLOSE PROPERLY. THIS ISSUE CAUSES THE NEEDLE TO BEND AND PIERCE THROUGH THE SAFETY COVER. THERE WAS ONE INCIDENT WHERE THIS RESULTED IN A NEEDLESTICK TO AN EMPLOYEE.

    Patient problems: Needle Stick/Puncture

    MAUDE report 21709700 ↗

FDA clearance and classification

The FDA regulates product code FMI (needle, hypodermic, single lumen) under 21 CFR 880.5570 as a Class II device. 21 CFR 880.5570 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the McKesson 16-N2558S

Has the McKesson 16-N2558S been recalled?

No FDA recall record in our data names this exact device identifier (DI 10612479244655), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the McKesson 16-N2558S?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Mar 27, 2025 and Mar 13, 2026. In those reports, the most common specific device problems are “Break” (1), “Contamination” (1) and “Decoupling” (1); the most common patient problem is “Needle Stick/Puncture” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the McKesson 16-N2558S?

None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the McKesson 16-N2558S, and is it still sold?

The FDA's device database lists Mckesson Medical-Surgical Inc. as the company responsible for it (brand McKesson, model 16-N2558S). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Needle, Hypodermic, Single Lumen devices

All Needle, Hypodermic, Single Lumen devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page