McKesson 16-N2558SRecalls & reported problems
Mckesson Medical-Surgical Inc. · NEEDLE, SAFETY 25GX5/8" (HINGED STYLE) · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10612479244655), received between Mar 27, 2025 and Mar 13, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Contamination1
- Decoupling1
- Fail-Safe Problem1
- Material Twisted/Bent1
Patient problems
- Needle Stick/Puncture2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2026Contamination
CUSTOMER REPORTED THAT THE NEEDLES ARE LEAVING A GRAY/SILVER METALLIC RESIDUE ON PATIENTS' SKIN. PRESENT BEFORE VIAL STOPPER PUNCTURE. NOT OBSERVED IN MOST OTHER LOTS OF SAME PRODUCT. REPRODUCIBLE ACROSS OBSERVERS AND SITES. CERTAIN NEEDLE LOTS EXHIBIT VISIBLE GRAY/DARK RESIDUE DIRECTLY OUT OF UNOPENED STERILE PACKAGING. WHEN WIPED WITH STERILE WHITE GAUZE OR TISSUE, A GRAY SMEAR IS PRODUCED. WHEN USED FOR INJECTION (SIMULATION AND PATIENT CARE), A VISIBLE DARK MARK IS AT THE INJECTION PATH. WHEN REPRODUCING IN ONE MEDIA WE WERE ABLE TO CUT INTO, THE DARK MARK TRACKED THROUGH THE MEDIA ALONG THE INJECTION PATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24589491 ↗MalfunctionFeb 17, 2026Break; Decoupling
CUSTOMER REPORTED THAT ,ON OCCASION THE NEEDLE COMES OFF (DOES NOT BREAK) AND IN ONE INCIDENT STAYS STUCK IN THE PATIENT. NO INJURY OCCURRED. ON OTHER OCCASIONS, THE NEEDLE HAS BROKEN AT THE HUB ALSO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24363705 ↗MalfunctionMay 16, 2025Material Twisted/Bent; Material Twisted/Bent
CUSTOMER REPORTED RETURNING THE NEEDLE DUE TO MALFUNCTION IN THE SAFETY FEATURE, WHICH DOES NOT CLOSE PROPERLY. ON (B)(6) 2025, THE CUSTOMER RESPONDED TO THE PROBING QUESTIONS AND CONFIRMED THAT THIS ISSUE WAS OBSERVED WITH A FEW NEEDLES, RESULTING IN AN EMPLOYEE INJURY. THEY FURTHER EXPLAINED THAT WHEN CLOSING OF THE SAFETY DEVICE, THE NEEDLE TENDS TO BEND AND PIERCE THROUGH THE ORANGE SAFETY COVER. UPDATE: THE CUSTOMER CONFIRMED THAT WHEN CLOSING OF THE SAFETY DEVICE, THE NEEDLE TENDS TO BEND AND PIERCE THROUGH THE ORANGE SAFETY COVER. THERE WAS ONE INCIDENT WHERE THIS RESULTED IN A NEEDLESTICK TO AN EMPLOYEE. NO TREATMENT WAS REQUIRED.
Patient problems: Needle Stick/Puncture
MAUDE report 22035481 ↗InjuryMar 27, 2025Fail-Safe Problem; Fail-Safe Problem
THE CUSTOMER REPORTED THAT THE NEEDLES EXPERIENCED A MALFUNCTION IN THE SAFETY FEATURE, WHICH DOES NOT CLOSE PROPERLY. THIS ISSUE CAUSES THE NEEDLE TO BEND AND PIERCE THROUGH THE SAFETY COVER. THERE WAS ONE INCIDENT WHERE THIS RESULTED IN A NEEDLESTICK TO AN EMPLOYEE.
Patient problems: Needle Stick/Puncture
MAUDE report 21709700 ↗
Other Needle, Hypodermic, Single Lumen devices
- BD Nano 2nd Gen 320550Becton Dickinson Pen Limited · Needle, hypodermic, single lumen2,044 reports
- BD SafetyGlide 305916Becton, Dickinson And Company · Needle, hypodermic, single lumen1,767 reports
- BD Nano Ultra-Fine 320122Becton Dickinson Pen Limited · Needle, hypodermic, single lumen601 reports
- N/A 305106Becton, Dickinson And Company · Needle, hypodermic, single lumen1 recall467 reports
- N/A 305110Becton, Dickinson And Company · Needle, hypodermic, single lumen418 reports
- BD Safe-Clip 328235Becton, Dickinson and Company Inc · Needle, hypodermic, single lumen268 reports
- BD Ultra-Fine 320109Becton Dickinson Pen Limited · Needle, hypodermic, single lumen229 reports
- BD AutoShield Duo 329515Becton, Dickinson And Company · Needle, hypodermic, single lumen208 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.