McKesson 306616Recalls & reported problems
Mckesson Medical-Surgical Inc. · Safety Hypodermic Needles GLIDE Thin Wall 25G x 1 in (50/bx, 10/cs) · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This McKesson safety hypodermic needle is a single-use needle used for injections. Many of the 113 reports describe the needle pulling out of its hub during or after an injection, and one describes a needle left in a patient's arm. The reports include 1 injury and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 112
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 112 adverse event reports that list this exact device (DI 10612479262598), received between Jun 8, 2023 and May 6, 2026 — classified as 1 injury, 111 malfunctions.
The most frequently reported specific device problem is “Complete Blockage” (95 reports).
4 other recalls cover devices cleared under the same FDA submission (K951254); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Complete Blockage95
- Break9
- Air/Gas in Device1
- Contamination1
- Defective Component1
- Detachment of Device or Device Component1
- Failure to Deliver1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions96
- Insufficient Information14
- Foreign Body In Patient1
- Needle Stick/Puncture1
Recent adverse event reports
The 25 most recent of 112 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2026Break
IT WAS REPORTED THAT BD NEEDLE SFTYGLD 25X1 RB MCK NEEDLE PULLED OUT OF HUB. VERBATIM: COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. RECEIVED THE FOLLOWING PRODUCT COMPLAINT. THIS COMMUNICATION IS INTENDED FOR YOUR QUALITY ASSURANCE DEPARTMENT. THE PURPOSE OF THIS NOTIFICATION IS TO PROVIDE YOUR COMPANY WITH INFORMATION CONCERNING A CUSTOMER COMPLAINT SO THAT, IF NECESSARY, ACTION CAN BE TAKEN TO ENSURE THAT QUALITY STANDARDS ARE BEING MET. PLEASE INVESTIGATE AND PROVIDE, WITH THE OUTCOME/ CORRECTIVE ACTION FOR THE BELOW LISTED COMPLAINT. PLEASE SEND AN ACKNOWLEDGEMENT UPON RECEIPT OF THIS NOTIFICATION. ALL INVESTIGATION RESULTS AND RESPONSES SHOULD BE SENT TO IT IS OUR GOAL TO CLOSE ALL COMPLAINTS WITHIN 60 DAYS. COMPLAINT NUMBER (B)(4). REPORT DATE 4/20/2026 FACTORY/MANUFACTURER MFR REORDER # (B)(4). PRODUCT NAME NEEDLE, SFTY 192-N251S PREVENTSG ORG 25GX1" LOT / SERIAL NUMBER (B)(6). PRODUCT CONCERN JUST GAVE A VACCINE WITH THE FOLLOWING NEEDLE ATTACHED, INJECTED ALL AS USUAL, WENT TO PULL IT FROM THE PATIENTS ARM AND THE NEEDLE CAME COMPLETELY OUT OF THE HOUSING, REMAINING IN THE ARM. SAMPLE AVAILABILITY PLEASE HAVE YOUR QA TEAM LOOK INTO THIS AND PROVIDE A WRITTEN RESPONSE REGARDING
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25101779 ↗MalfunctionMay 4, 2026Break
NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25070999 ↗MalfunctionMar 18, 2026Break
THE FOLLOWING WAS PROVIDED BY THE INITIAL REPORTER: PRODUCT: 306616 - BD NEEDLE SFTYGLD 25X1 RB MCK - NEEDLE PULLED OUT OF HUB. UNIT OF MEASURE: QUANTITY: 1. LOT #: 5093367. PRODUCT PICK-UP INFO: ADDITIONAL DESCRIPTION NOTES: DONT SEND SAME AFFECTED LOT NUMBER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24628016 ↗MalfunctionJan 27, 2026Break; Material Integrity Problem
IT WAS REPORTED THAT BD NEEDLE SFTYGLD 25X1 RB MCK NEEDLE PULLED OUT OF HUB. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. WHEN ENGAGING SAFETY NEEDLE COVER AFTER ADMINISTERING VACCINE, NEEDLE POPPED OFF AND SCRATCHED NURSE IN THIGH.
Patient problems: Needle Stick/Puncture
MAUDE report 24188991 ↗MalfunctionJan 20, 2026Break
IT WAS REPORTED THAT BD NEEDLE SFTYGLD 25X1 RB MCK NEEDLE PULLED OUT OF HUB. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. THIS NEEDLE CAME DETACHED FROM THE LEUER LOCK AND WAS IN THE PATIENT'S ARM. (B)(6) 2026: ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. IT WAS UNCLEAR IF THE PATIENT RECEIVED THE FULL VACCINE OR NOT SINCE THE NEEDLE CAME OUT AND SOME FLUID LEAKED AFTERWARDS BUT IT WAS ASSUMED THAT THEY RECEIVED THE MAJORITY OF THE VACCINE. NO INJURY NOTED, RN REMOVED NEEDLE THAT WAS LEFT IN PATIENT'S ARM. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN FORMAT DD-MMM-YYYY? 22-12-2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24117632 ↗MalfunctionJan 5, 2026Break
IT WAS REPORTED THAT THE BD NEEDLE SFTYGLD 25X1 RB MCK NEEDLE PULLED OUT OF HUB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL#: 306616, BATCH#: 5093369. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. SAFETY HYPODERMIC NEEDLE GLIDE 25G X 1IN. (LOT # 5093369) WHILE REMOVING THE NEEDLE AFTER ADMINISTERING THE INJECTION THE HUB BROKE OFF THE NEEDLE. THE NEEDLE HAD TO BE REMOVED SEPARATELY. ADDITIONAL INFORMATION PROVIDED: 1. CAN YOU PLEASE PROVIDE AN EXACT DATE OF THE EVENT IN FORMAT DD-MMM-YYYY? FRIDAY 12/12/2025. 2. COULD YOU PLEASE CONFIRM IF THERE ARE ANY SAMPLES AVAILABLE TO SEND TO BD FOR INVESTIGATION? NO, WE DO NOT, WE DISCARDED EVERYTHING. WE JUST HAVE THE PICTURES THAT WERE INCLUDED. 3. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. NO, NO HARM OR INJURY WERE HAPPENED. WHEN TAKING THE INJECTION OUT THE HUB BROKE OFF, AND THE NEEDLE STAYED IN THE PATIENT. WE WERE ABLE TO SUCCESSFUL REMOVE THE NEEDLE WITHOUT ANY HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23957586 ↗MalfunctionDec 31, 2025Break; Break
IT WAS REPORTED THAT THE BD NEEDLE SFTYGLD 25X1 RB MCK PULLED OUT OF THE HUB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL # 306616 BATCH # 5045756. RCC RECEIVED A COMPLAINT VIA EMAIL. AFTER ADMINISTERING THE INJECTION INTO THE DELTOID OF A PATIENT, THE MEDICAL ASSISTANT ATTEMPTED TO REMOVE THE NEEDLE FROM THE ARM TO PUT IT INTO SHARPS. THE NEEDLE BECAME DETACHED FROM THE LUER LOCK AND SAFETY, LEAVING THE NEEDLE IN THE PATIENTS ARM AND THE PREFILLED SYRINGE AND LUER LOCK PORTION OF THE SYRINGE INTACT. SHE WAS ABLE TO REMOVE THE NEEDLE FROM THE PATIENT. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23936956 ↗InjuryMay 22, 2025Break; Break
IT WAS REPORTED THAT THE BD NEEDLE SFTYGLD 25X1 RB MCK WAS BROKEN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL # 306616. BATCH # UNKNOWN. VERBATIM: ACCOUNT GAVE AN HPV SHOT TO A YOUNG PATIENT OVER A YEAR AGO. THE PATIENT RECENTLY CAME BACK IN AND IT WAS DISCOVERED A NEEDLE IS IN THE PATIENTS ARM AND HAS BEEN WORKING ITS WAY DOWN THROUGH THE MUSCLE. THE ACCOUNT DOES NOT KNOW IF ITS THE NEEDLE FROM THE HPV SHOT OR WHEN THE INCIDENT OCCURRED, BUT THE ACCOUNT WOULD LIKE TO SPEAK TO SOMEONE AT BD REGARDING THE EVENT. CUSTOMER RESPONSE ON 13-MAY-2025. 1. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT? (B)(6) 2024. 2. CAN YOU PLEASE PROVIDE THE MATERIAL AND LOT NUMBER ASSOCIATED WITH REPORTED ISSUE? PREVENT SG SAFETY HYPODERMIC NEEDLE GLIDE NON-PYROGENIC STERILE 25G X 1INCH. MFR: 192-N251S. I CANNOT CONFIRM THE LOT NUMBER DUE TIME THAT HAS PAST FROM THE EVENT. THE CURRENT LOT NUMBER WE HAVE IN STOCK IN CLINIC IS 0141231. 3. PLEASE DESCRIBE ANY ADDITIONAL, UNPLANNED MEDICAL OR SURGICAL INTERVENTION (E.G., ADDITIONAL/ALTERNATIVE MEDICATIONS, IMAGING, ETC.) REQUIRED TO AVOID PATIENT HARM. NEEDLE WAS DISCOVERED VIA AN XRAY RECENTLY DUE TO A SEPARATE, UNRELATED INJU
Patient problems: Foreign Body In Patient
MAUDE report 22074688 ↗MalfunctionMar 28, 2025Break; Break
MATERIAL # 306616 BATCH # 5010901 IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE PULLED OUT OF THE HUB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED COMPLAINT NUMBER (B)(4). REPORT DATE - 03.11.25 MFR REORDER # (B)(4). PRODUCT NAME - NEEDLE, SFTY 192-N251S PREVENTS ORG 25GX1" LOT / SERIAL NUMBER - (B)(6). PRODUCT CONCERN - NEEDLE DETACHED/CAME OFF HUB WHILE IN THE PATIENT'S ARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21718702 ↗MalfunctionDec 8, 2024Complete Blockage; Complete Blockage
MATERIAL # 306616, LOT # 2202914. IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE WAS CLOGGED / BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. VACCINE WOULD NOT ADVANCE IN NEEDLE. LOT NUMBER - 2202914. ADDITIONAL INFORMATION PROVIDED: 1) PLEASE PROVIDE DATE OF EVENT? 2)ANY SAMPLE AVAILABLE FOR INVESTIGATION? IF YES, ARE YOU ABLE TO PROVIDE THE ADDRESS OF THE FACILITY FOR US TO SHIP THE RETURN LABEL? 3). PLEASE SHARE THE CAPTURED PHOTO OF THE EVENT IF AVAILABLE. 4) ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. 1. 11/1/2024. 2. WE HAVE THE BOX OF NEEDLES, BUT THE USED NEEDLE WAS PLACED IN THE SHARPS CONTAINER AFTERWARDS. 3. I DO NOT HAVE A PHOTO OF THE INCIDENT OR THE NEEDLE. I DO HAVE A PHOTO OF THE BOX THE NEEDLE CAME FROM. 4. NO INJURY OR ADVERSE EVENT TO PATIENT OR HEALTHCARE PROFESSIONAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20875238 ↗MalfunctionAug 15, 2024Complete Blockage; Complete Blockage
NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19997075 ↗MalfunctionJun 9, 2024Complete Blockage; Complete Blockage
MATERIALS#: (B)(4) , BATCH#: 2202914. IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE WAS CLOGGED/BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "VACCINE WOULD NOT PUSH THROUGH THE NEEDLES." PRODUCT DETAILS - NEEDLE, SFTY 192-N251S PREVENTSG ORG 25GX1" LOT - 2202914. ADDITIONAL INFORMATION PROVIDED: ARE YOU ABLE TO PROVIDE THE DATES OF THE EVENTS IN THE FORMAT OF MM-DD-YYYY? IF UNKNOWN, CAN STATE UNKNOWN. THE DATE THE FAULTY NEEDLES WERE DISCOVERED WAS (B)(6) 2024. HOW WAS THE PATIENT OUTCOME? ARE THERE ANY CLINICAL SIGNS, HEALTH CONSEQUENCES OR IMPACT? THERE WERE NO HEALTH CONSEQUENCES TO ANY PATIENTS. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTH CARE PROFESSIONAL? NO ADVERSE EVENTS OR SERIOUS INJURIES OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19495388 ↗MalfunctionJun 9, 2024Complete Blockage; Complete Blockage
MATERIAL# 306616. BATCH# 3300424. IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE CLOGGED/BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE NEEDLE PLUGS HALFWAY THROUGH GIVING AN INJECTION." VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. COMPLAINT NUMBER (B)(4). REPORT DATE 5/24/24. FACTORY/MANUFACTURER BD. MFR REORDER # (B)(4). PRODUCT NAME NEEDLE, SFTY 192-N251S PREVENTSG ORG 25GX1" LOT / SERIAL NUMBER (B)(6). PRODUCT CONCERN: THE NEEDLE PLUGS HALF WAY THROUGH GIVING AN INJECTION. ADDITIONAL INFORMATION PROVIDED 1. WHAT IS THE DATE OF EVENT? (B)(6) 2024. 2. PLEASE SHARE THE MATERIAL NUMBER. LOT NUMBER 3300424. 3. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. NO HARM BUT THE NEEDLE BECAME PLUGGED HALFWAY THROUGH GIVING ANY INJECTION SO THE NEEDLE HAD TO BE CHANGED AND PATIENT THEN HAD TO BE REPOKED TO FINISH ADMINISTERING THE INJECTION. 4. ANY SAMPLE AVAILABLE FOR INVESTIGATION? IF YES, ARE YOU ABLE TO PROVIDE THE ADDRESS OF THE FACILITY FOR US TO SHIP THE RETURN LABEL? NEEDLES WERE RETURNED TO OUR AMES PURCHASING DEPT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19495360 ↗MalfunctionMay 14, 2024Complete Blockage; Complete Blockage
NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Insufficient Information
MAUDE report 19315675 ↗MalfunctionApr 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
MATERIAL#306616 BATCH# UNKNOWN. IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE PULLED OUT OF THE HUB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE ENTIRE NEEDLE PULLED OUT OF THE HUB AND WAS LEFT INSIDE A INFANTS LEG." VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. NEEDLE, SFTY 192-N231S PREVENTSG BLU 23GX1". THE ENTIRE NEEDLE PULLED OUT OF THE HUB AND WAS LEFT INSIDE A INFANTS LEG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19164118 ↗MalfunctionApr 11, 2024Defective Component; Defective Component
IT WAS REPORTED THAT THE BD SAFETYGLIDE SAFETY MECHANISM WAS DIFFICULT TO ACTIVATE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY CUSTOMER THAT THE PRODUCT NEEDLE AND SAFETY MECHANISM IS FLIMSY. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. PRODUCT NEEDLE AND SAFETY MECHANISM IS FLIMSY. STAFF FEELS STRONGLY THESE WILL INCREASE CHANCES OF NEEDLE STICK WHEN TRYING TO INITIATE SAFETY. LOT#: 3292199 PRODUCT: NEEDLE, SFTY 192-N251S PREVENTSG ORG 25GX1".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19089580 ↗MalfunctionApr 9, 2024Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE WAS CLOGGED/BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "I RECEIVED A COMPLAINT ON THE BELOW PRODUCT. WHEN THE NEEDLE IS LUERED ON, IT IS REQUIRING EXTREME PRESSURE TO PRIME OUT AIR IN NEEDLE. THERE APPEARS TO BE SOMETHING BLOCKING OR IN SYRINGE, POSSIBLY A SMALL PIECE OF PLASTIC. THIS HAS HAPPENED WITH MULTIPLE LOTS. PLEASE ADVISE WHAT WE SHOULD DO. IS THIS JUST A BAD LOT?" VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. THERE APPEARS TO BE SOMETHING BLOCKING OR IN SYRINGE, POSSIBLY A SMALL PIECE OF PLASTIC. LOT#: 3300424. PRODUCT: NEEDLE, SFTY 192-N251S PREVENTSG ORG 25GX1".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19076284 ↗MalfunctionFeb 20, 2024Complete Blockage; Complete Blockage
NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18739253 ↗MalfunctionFeb 20, 2024Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE WAS CLOGGED/BLOCKED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER: "THE MEDICATION DOES NOT COME OUT OF THE NEEDLES. LIKE THEY ARE BLOCKED." COMPLAINT NUMBER (B)(4) REPORT DATE 1/23/2024 FACTORY/MANUFACTURER BD MFR REORDER # (B)(4) PRODUCT NAME NEEDLE, SFTY 192-N251S PREVENTSG ORG 25GX1" LOT / SERIAL NUMBER (B)(6) PRODUCT CONCERN THE MEDICATION DOES NOT COME OUT OF THE NEEDLES. LIKE THEY ARE BLOCKED. CUSTOMER NAME (B)(6) CLINICS (B)(6) (B)(6) CUSTOMER CONTACT/EMAIL/PHONE (B)(6) CLINICS (B)(6) (B)(6)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18738981 ↗MalfunctionFeb 20, 2024Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE WAS CLOGGED/BLOCKED. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER: "THE MEDICATION DOES NOT COME OUT OF THE NEEDLES. LIKE THEY ARE BLOCKED." COMPLAINT NUMBER (B)(4). REPORT DATE 1/23/24 FACTORY/MANUFACTURER BD MFR REORDER # 192-N251S PRODUCT NAME NEEDLE, SFTY 192-N251S PREVENTSG ORG 25GX1" LOT / SERIAL NUMBER (B)(6). PRODUCT CONCERN THE MEDICATION DOES NOT COME OUT OF THE NEEDLES. LIKE THEY ARE BLOCKED. (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18738839 ↗MalfunctionFeb 1, 2024Complete Blockage; Complete Blockage
NO ADDITIONAL INFORMATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18628412 ↗MalfunctionFeb 1, 2024Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE WAS CLOGGED/BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "WE USE THE MCKESSON 25GX1" SAFETY HYPODERMIC NEEDLES (#306616). WE'VE BEEN HAVING SOME ISSUES WITH THESE; WHEN INJECTING THE MEDICATION, IT'S AS THOUGH THE NEEDLE BECOMES BLOCKED. SOMETIMES WE INJECT 0.2ML BEFORE IT BLOCKS UP, SOMETIMES IT'S 0.5ML, THAT PART IS RANDOM. BUT WHEN THIS HAPPENS, WE HAVE TO POKE THE PATIENT A SECOND TIME. ANY RECOMMENDATION? HAVE YOU HEARD ABOUT THIS FROM ANYONE ELSE? I KNOW THIS IS THE 2ND OR 3RD BOX THAT WE'VE HAD THIS ISSUE WITH, SO I'M NOT SURE WHERE TO GO FROM HERE." ADDITIONAL INFORMATION RECEIVED ON 01/17/24: 1) MCKESSON SAFETY HYPODERMIC NEEDLES GLIDE, THIN WALL 25G X 1 IN MFR# 306616, LOT 3255801 *THERE WAS A PREVIOUS BOX ALSO, BUT WE WORKED OUR WAY THROUGH THAT ONE, SO I DO NOT HAVE THE LOT FROM THAT BOX. 2) I DO NOT HAVE AN EXACT DATE OF WHEN THIS OCCURRED. IT'S BEEN HAPPENING OVER THE LAST MONTH, EXCEPT IT DID HAPPEN TO (B)(6) YESTERDAY, (B)(6) 2024. 3) WE DO NOT HAVE A "PHYSICAL SAMPLE" AVAILABLE. DUE TO THIS BEING A NEEDLE, IT'S DISPOSED INTO A BIOHAZARD CONTAINER. IF WE NEED TO KEEP A NEEDLE, YOU WILL NEED T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18628410 ↗MalfunctionJan 31, 2024Complete Blockage; Complete Blockage
MATERIAL#: 306616, LOT#: 2188470. IT WAS REPORTED BY THE CUSTOMER REPORTED THAT NEEDLES WILL NOT ALLOW VACCINE TO BE PUSHED THRU THEM. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. NEEDLES WILL NOT ALLOW VACCINE TO BE PUSHED THRU THEM. 192-N251S. PRODUCT NAME: NEEDLE, SFTY 192-N251S PREVENTSG ORG 25GX1". COMPLAINT NUMBER: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18613833 ↗MalfunctionDec 27, 2023Complete Blockage; Complete Blockage
IT WAS REPORTED THAT THE BD SAFETYGLIDE NEEDLE WAS CLOGGED/BLOCKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NO INJURIES REPORTED. CUSTOMER STATES SOME OF THE NEEDLES DO NOT HAVE ANY HOLES. THEY ARE SOLID." PRODUCT#: 306616. LOT#: 2202914.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18405924 ↗MalfunctionDec 16, 2023Complete Blockage; Complete Blockage
MATERIAL#: UNKNOWN BATCH#: UNKNOWN. IT WAS REPORTED BY CUSTOMER THAT THE SOLUTION IN THE SYRINGE WILL NOT PASS THROUGH THE NEEDLE. NEEDLE CLOGGED. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. COMPLAINT NUMBER: DM-23-1177. REPORT DATE: 11/29/23. FACTORY/MANUFACTURER: BD. MFR REORDER#: 192-N251S. PRODUCT NAME : NEEDLE, SFTY 192-N251S PREVENTSG ORG 25GX1". LOT / SERIAL NUMBER: (B)(6). PRODUCT CONCERN: THE SOLUTION IN THE SYRINGE WILL NOT PASS THROUGH THE NEEDLE. NEEDLE CLOGGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18341656 ↗
FDA clearance and classification
Cleared through 510(k) K951254 on Oct 7, 1995 (decision: Substantially Equivalent). Submitted by Bd Becton Dickinson Vacutainer Systems Preanalytic as “Safetyglide Needle”. The FDA review took 200 days. FDA record ↗
The FDA regulates product code FMI (needle, hypodermic, single lumen) under 21 CFR 880.5570 as a Class II device. 21 CFR 880.5570 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K951254. One submission often covers many models, so check the product and lot details in each record.
- Z-2889-2020 · Jul 28, 2020 · TerminatedBD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack
- Z-0315-2018 · Jul 5, 2017 · TerminatedSafetyGlide Needle, 22GA 1-1/2in, Catalog n. 301701, manufactured by BD, sold in Centurion Convenience Kits. Centurion component code BD17O1NS.
- Z-3116-2017 · Jun 13, 2017 · TerminatedBD Needle 22GA 1-1/2in SafetyGlide(TM) Non Sterile, Bulk, REF 301701 The BD SafetyGlide(TM) needle is intended to be used for general purpose injection, aspiration of fluids from
- Z-2748-2017 · Sep 16, 2016 · TerminatedBD SafetyGlide" Injection Needle Catalog #305917 The SafetyGlide " needle is intended to be used for general purpose injection, aspiration of fluids from parts of the body below th
Common questions about the McKesson 306616
Has the McKesson 306616 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10612479262598), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the McKesson 306616?
The FDA's MAUDE database lists 112 adverse event reports for this device, received between Jun 8, 2023 and May 6, 2026. In those reports, the most common specific device problems are “Complete Blockage” (95), “Break” (9) and “Air/Gas in Device” (1); the most common patient problems are “Foreign Body In Patient” (1) and “Needle Stick/Puncture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the McKesson 306616?
None of the reports are classified as deaths, 1 as an injury and 111 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the McKesson 306616 FDA approved?
It was cleared through 510(k) K951254 on Oct 7, 1995. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the McKesson 306616, and is it still sold?
The FDA's device database lists Mckesson Medical-Surgical Inc. as the company responsible for it (brand McKesson, model 306616). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Needle, Hypodermic, Single Lumen devices
- BD Nano 2nd Gen 320550Becton Dickinson Pen Limited · Needle, hypodermic, single lumen1,892 reports
- BD SafetyGlide 305916Becton, Dickinson And Company · Needle, hypodermic, single lumen1,768 reports
- BD Nano Ultra-Fine 320122Becton Dickinson Pen Limited · Needle, hypodermic, single lumen566 reports
- N/A 305106Becton, Dickinson And Company · Needle, hypodermic, single lumen1 recall465 reports
- N/A 305110Becton, Dickinson And Company · Needle, hypodermic, single lumen418 reports
- BD Safe-Clip 328235Becton, Dickinson and Company Inc · Needle, hypodermic, single lumen268 reports
- BD Ultra-Fine 320109Becton Dickinson Pen Limited · Needle, hypodermic, single lumen208 reports
- BD AutoShield Duo 329515Becton, Dickinson And Company · Needle, hypodermic, single lumen200 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page