Neo-Tee T-Piece Resuscitator, for pediatric emergency… recallZ-0821-2025 · Mercury Enterprises, Inc. dba Mercury Medical · initiated Dec 10, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Mercury Enterprises, Inc. dba Mercury Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.
FDA root cause category: Process change control
What the company is doing
On 12/10/2024, "URGENT: FIELD SAFETY NOTICE" letters were mailed and emailed to medical facilities and distributors who were asked to do the following: Medical Facilities/Distributors: 1) Users should cease use and distribution of affected product and immediately quarantine the affected product. 2) Contact firm's customer service department at [email protected] (Domestic) and [email protected] (International) to assist you on how to send the affected product back to the firm for product replacement or credit. 3) Please notify all personnel subject to using these devices. 4) Complete and return the response form via email to [email protected] In addition, distributors were asked to do the following: 5) Provide a copy of this recall notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you. If you have any questions regarding this action, contact the firm at 800-237-6418/727-573-0088, or e-mail at [email protected]
Recall details
- Recall number
- Z-0821-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 10, 2024
- Posted by the FDA
- Jan 10, 2025
- Recalling firm
- Mercury Enterprises, Inc. dba Mercury Medical
- Quantity
- 37370
- Product code
- BTL
- FDA submissions
- K093913
- Distribution
- Worldwide Distribution: US (nationwide) to states of: PA, FL, IN, WI, ME, SC, WY, ID, MT, TX, MI, LA, MS, AR, WV, MD, KY, NJ, WA, CA, GA, MO, MA, NC, NV, TN, VA, OK, ND, MN, AL, HI, UT, NY, AZ, IL, OH, IA, NE, CT, AK, OR and OUS (foreign) to countries of: Hungary, Switzerland, Hong Kong, Spain, Canada, Italy, United Kingdom, United Arab Emirates, Saudi Arabia, France, Netherlands
Product description
Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, 1050841, 1050842
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Part/UDI-DI/Lot: 1050805/10641043508053, 30641043508057/2421450805, 2423550805, 2423650805, 2426750805; 1050808/10641043508084, 30641043508088/2426450808, 2426750808, 2426850808, 2429050808, 2429350808; 1050809/10641043508091, 30641043508095/2426450809, 2426750809, 2429050809, 2429150809, 2429350809, 2429750809, 2429850809, 2430050809, 2430250809; 1050810 /10641043508107, 30641043508101/2426750810, 2429350810, 2429750810, 2429850810, 2430250810; 1050811/10641043508114, 30641043508118/2426550811, 2426650811, 2426750811, 2429050811, 2429150811, 2429250811, 2429350811, 2429450811, 2429750811, 2429850811, 2429950811, 2430250811; 1050814/10641043508145, 30641043508149/2426450814, 2426750814, 2429050814, 2429750814, 2430250814; 1050832/10641043508329, 30641043508323/2429050832, 242985083; 1050839/10641043508398, 30641043508392/2430350839; 1050840/10641043508404, 30641043508408/2429050840, 2429750840; 1050841/10641043508411, 30641043508415/2429050841; 1050842/10641043508428, 30641043508422/2426750842
Devices named in this recall
11 devices in the FDA device registry whose identifier appears in the recall's code information.
- Neo-Tee T-Piece Resuscitator 10-50805Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)3 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50811Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls3 reports
- Neo-Tee T-Piece Resuscitator 10-50808Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50809Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50810Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50832Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50814Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50839Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50840Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50841Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50842Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
Other devices under the same FDA submission
25 devices cleared under K093913 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Neo-Tee T-Piece Resuscitator 10-50801Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50802Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50803Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50804Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50806Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50807Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50812Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50815Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50816Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50817Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50818Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
- Neo-Tee T-Piece Resuscitator 10-50819Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page