Skip to content

POOLE 0035040Recalls & reported problems

Conmed Corporation · POOLE Suction Instrument with Removable Sheath · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10653405000389), received between Oct 10, 2016 and Mar 12, 2026 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Delivered as Unsterile Product9
  • Device Packaging Compromised3
  • Unsealed Device Packaging2
  • Detachment of Device or Device Component1
  • Device Damaged Prior to Use1
  • Packaging Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Not Applicable4
  • No Patient Involvement1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 12, 2026Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0035040, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24581006 ↗
  • MalfunctionSep 12, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0035040, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23037563 ↗
  • MalfunctionMar 17, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    UPDATE 19FEB25: THE DEVICE WAS RETURNED AND EVALUATED. THE COMPLETION OF THE DEVICE EVALUATION CONFIRMED AN INSUFFICIENT HEATSEAL; THEREFORE, THIS COMPLAINT HAS BEEN REASSESSED AND MEETS THE CRITERIA FOR A REPORTABLE EVENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE. THE DISTRIBUTOR REPORTED THAT THE 0035040 DEVICE HAD AN INSUFFICIENT HEATSEAL FOUND DURING INCOMING INSPECTION. NO PATIENT INVOLVEMENT OCCURRED. THE SHIPMENT ASSOCIATED WITH THE RGA FOR THIS DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION AND IT HAS NOT BEEN CONFIRMED THAT AN INSUFFICIENT HEATSEAL CAUSING A STERILITY BREACH HAS OCCURRED. BASED ON THE INFORMATION AVAILABLE, AND THE DECISION TREE RESULTS, THIS COMPLAINT DOES NOT MEET THE DEFINITION OF A REPORTABLE EVENT. UPON THE RECEIPT OF THE DEVICE, AN EVALUATION WILL BE CONDUCTED AND IF A BREACH OF STERILITY IS CONFIRMED THE REPORTING DECISION WILL BE REASSESSED AT THAT TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21617297 ↗
  • MalfunctionJan 17, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0035040, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21178367 ↗
  • MalfunctionSep 15, 2023Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0035040, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17754231 ↗
  • MalfunctionJan 26, 2023Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0035040, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16247464 ↗
  • MalfunctionJan 9, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT THE DEVICE, 0035040, POOLE SUCTION INSTRUMENT (50/CS), WAS BEING USED DURING AN EMERGENCY LAPAROTOMY PROCEDURE ON (B)(6) 2022 WHEN IT WAS REPORTED, ¿WE EXPERIENCED SOME ISSUES WITH 3 OF THIS PRODUCT OVERNIGHT DURING AN EMERGENCY LAPAROTOMY. TWO (2) CAME APART WITHIN THE OPEN ABDOMEN, AND A THIRD BROKE DURING ASSEMBLY PRIOR TO USE. I HAVE BEEN ABLE TO CONFIRM THESE ARE ALL FROM THE SAME BATCH/LOT NUMBER.¿. THE PROCEDURE WAS REPORTED AS HAVING BEEN COMPLETED WITHOUT ANY INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THE COMPONENTS WERE RETRIEVED FROM THE PATIENT BY HAND CAUSING A 5¿10-MINUTE DELAY. IT WAS ADVISED THAT THE PATIENT IS ¿ABSOLUTELY FINE¿. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16120164 ↗
  • MalfunctionJul 25, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0035040, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15092544 ↗
  • MalfunctionMar 8, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0035040, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13695097 ↗
  • MalfunctionFeb 18, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0035040, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13558172 ↗
  • MalfunctionMay 15, 2017Device Packaging Compromised; Device Packaging Compromised

    AS REPORTED BY THE DISTRIBUTOR IN (B)(6), ONE (1) POOLE SUCTION INSTRUMENT WITH REMOVABLE SHEATH WAS DISCOVERED TO HAVE AN "INSUFFICIENT HEATSEAL OR BLISTER PACK." THERE WAS NO PATIENT INVOLVEMENT IN THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER. THIS REPORT IS BEING FILED BASED ON THE POTENTIAL FOR INJURY WITH RECURRENCE.

    Patient problems: Not Applicable

    MAUDE report 6566605 ↗
  • MalfunctionApr 26, 2017Unsealed Device Packaging; Unsealed Device Packaging

    THE DISTRIBUTOR IN (B)(6) REPORTED THAT DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, TWO (2) PACKAGES, CONTAINING ONE EACH POOLE WITH REMOVABLE SHEATH, WERE DISCOVERED WITH "THE SEAL OF THE STERILE POUCH WAS BREACHED." IN THIS INSTANCE, THERE WAS NO PATIENT INVOLVEMENT WITH THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER. THIS REPORT IS BEING FILED BASED ON THE POTENTIAL FOR INJURY WITH RECURRENCE.

    Patient problems: Not Applicable

    MAUDE report 6523353 ↗
  • MalfunctionApr 4, 2017Device Packaging Compromised; Device Packaging Compromised

    AS REPORTED BY THE DISTRIBUTOR IN (B)(4), ONE (1) POOLE SUCTION INSTRUMENT WAS DISCOVERED TO HAVE AN "INSUFFICIENT HEATSEAL OR BLISTER PACK." THERE WAS NO PATIENT INVOLVEMENT IN THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER. THIS REPORT IS BEING FILED BASED ON THE POTENTIAL FOR INJURY WITH RECURRENCE.

    Patient problems: Not Applicable

    MAUDE report 6457087 ↗
  • MalfunctionMar 21, 2017Unsealed Device Packaging; Unsealed Device Packaging

    AS REPORTED BY THE DISTRIBUTOR IN (B)(4), DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, 1 PACKAGE CONTAINING A POOLE SUCTION INSTRUMENT WAS DISCOVERED TO HAVE "INSUFFICIENT HEATSEAL." THERE WAS NO PATIENT INVOLVEMENT IN THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER. THIS REPORT IS BEING FILED BASED ON THE POTENTIAL FOR INJURY WITH RECURRENCE.

    Patient problems: Not Applicable

    MAUDE report 6421917 ↗
  • MalfunctionOct 10, 2016Device Damaged Prior to Use; Device Packaging Compromised; Packaging Problem

    THE DISTRIBUTOR IN (B)(4) REPORTED THAT DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, 2 PACKAGES, EACH CONTAINING ONE 10 FRENCH FRAZIER SUCTION INSTRUMENT, WAS DISCOVERED WITH "INSUFFICIENT HEAT SEAL." THERE WAS NO PATIENT INVOLVEMENT WITH THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER.

    Patient problems: No Patient Involvement

    MAUDE report 6015323 ↗

FDA clearance and classification

The FDA regulates product code BYY (tube, aspirating, flexible, connecting) under 21 CFR 880.6740 as a Class II device. 21 CFR 880.6740 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the POOLE 0035040

Has the POOLE 0035040 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10653405000389), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the POOLE 0035040?

The FDA's MAUDE database lists 15 adverse event reports for this device, received between Oct 10, 2016 and Mar 12, 2026. In those reports, the most common specific device problems are “Delivered as Unsterile Product” (9), “Device Packaging Compromised” (3) and “Unsealed Device Packaging” (2); the most common patient problem is “Not Applicable” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the POOLE 0035040?

None of the reports are classified as deaths, 0 as injuries and 15 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the POOLE 0035040 FDA approved?

Devices of this type (product code BYY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Conmed Corporation.

Who makes the POOLE 0035040, and is it still sold?

The FDA's device database lists Conmed Corporation as the company responsible for it (brand POOLE, model 0035040). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TUBE, ASPIRATING, FLEXIBLE, CONNECTING devices

All TUBE, ASPIRATING, FLEXIBLE, CONNECTING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page