CONMED 0036280Recalls & reported problems
Conmed Corporation · CONMED Suction Connecting Tubing, 6' long (1.8 m), 1/4" I.D. (6.4 mm) · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10653405000464), received between Apr 20, 2016 and Sep 24, 2021 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Delivered as Unsterile Product3
- Device Packaging Compromised3
- Unsealed Device Packaging3
Patient problems
- Not Applicable5
- No Clinical Signs, Symptoms or Conditions3
- No Patient Involvement1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2021Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, CATALOG # 0036280, LOT 202006265, FOR A POSSIBLE BREACH OF STERILITY DUE TO AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH ANY PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION IN JAPAN. THE COMPLETION OF THE DEVICE EVALUATION CONFIRMED AN INSUFFICIENT HEATSEAL; THEREFORE, THIS COMPLAINT MEETS THE CRITERIA FOR A REPORTABLE EVENT. DUE TO THE BREACH IN STERILITY, THIS REPORT IS BEING RAISED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12528341 ↗MalfunctionSep 1, 2021Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 0036280, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12408482 ↗MalfunctionAug 30, 2021Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, CATALOG # 0036280, LOT 202001145, FOR A POSSIBLE BREACH OF STERILITY DUE TO AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH ANY PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION IN JAPAN. THE COMPLETION OF THE DEVICE EVALUATION CONFIRMED AN INSUFFICIENT HEATSEAL; THEREFORE, THIS COMPLAINT MEETS THE CRITERIA FOR A REPORTABLE EVENT. DUE TO THE BREACH IN STERILITY, THIS REPORT IS BEING RAISED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12385543 ↗MalfunctionMar 28, 2017Device Packaging Compromised; Device Packaging Compromised
AS REPORTED BY THE DISTRIBUTOR IN (B)(4), ONE (1) 1/4" X 6' SUCTION TUBING PACKAGE WAS DISCOVERED TO HAVE AN "INSUFFICIENT HEATSEAL." THERE WAS NO PATIENT INVOLVEMENT IN THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER. THIS REPORT IS BEING FILED BASED ON THE POTENTIAL FOR INJURY WITH RECURRENCE.
Patient problems: Not Applicable
MAUDE report 6440363 ↗MalfunctionMar 28, 2017Device Packaging Compromised; Device Packaging Compromised
AS REPORTED BY THE DISTRIBUTOR IN (B)(6), THREE (3) 0036280 1/4" X 6' SUCTION TUBING PACKAGES WERE DISCOVERED TO HAVE AN "INSUFFICIENT HEATSEAL OR BLISTER PACK." THERE WAS NO PATIENT INVOLVEMENT IN THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER. THIS REPORT IS BEING FILED BASED ON THE POTENTIAL FOR INJURY WITH RECURRENCE.
Patient problems: Not Applicable
MAUDE report 6438466 ↗MalfunctionFeb 2, 2017Unsealed Device Packaging; Unsealed Device Packaging
THE DISTRIBUTOR IN (B)(4) REPORTED THAT DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, ONE (1) PACKAGE CONTAINING ONE (1) EACH 1/4" X 6' SUCTION TUBING WAS DISCOVERED WITH "INSUFFICIENT HEAT SEAL." IN THIS INSTANCE, THERE WAS NO PATIENT INVOLVEMENT WITH THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER.
Patient problems: Not Applicable
MAUDE report 6299011 ↗MalfunctionJan 13, 2017Unsealed Device Packaging; Unsealed Device Packaging
THE DISTRIBUTOR IN (B)(4) REPORTED THAT DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, 2 PACKAGES CONTAINING 1/4" X 6' SUCTION TUBING WAS DISCOVERED WITH "DEFECTIVE SEAL." THERE WAS NO PATIENT INVOLVEMENT WITH THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER. THIS REPORT IS BEING FILED BASED ON THE POTENTIAL FOR INJURY WITH RECURRENCE.
Patient problems: Not Applicable
MAUDE report 6248860 ↗MalfunctionDec 20, 2016Unsealed Device Packaging; Unsealed Device Packaging
THE DISTRIBUTOR IN (B)(4) REPORTED THAT DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, TWO (2) PACKAGES CONTAINING ONE (1) EACH 1/4" X 6' SUCTION TUBING WERE DISCOVERED WITH "INSUFFICIENT HEAT SEAL." IN THIS INSTANCE, THERE WAS NO PATIENT INVOLVEMENT WITH THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER.
Patient problems: Not Applicable
MAUDE report 6192087 ↗MalfunctionApr 20, 2016Device Packaging Compromised
THE DISTRIBUTOR IN JAPAN REPORTED THAT DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, TWO (2) PACKAGES CONTAINING 1/4" X 6' SUCTION TUBING WERE DISCOVERED WITH "INSUFFICIENT HEAT SEAL". BOTH OF THE RETURNED PACKAGES EXHIBITED FULL OPEN SEAL DISRUPTION IN THE FINAL MANUFACTURING SEALS. IN THIS INSTANCE, THERE WAS NO PATIENT INVOLVEMENT WITH THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER.
Patient problems: No Patient Involvement
MAUDE report 5587926 ↗
FDA clearance and classification
The FDA regulates product code BYY (tube, aspirating, flexible, connecting) under 21 CFR 880.6740 as a Class II device. 21 CFR 880.6740 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the CONMED 0036280
Has the CONMED 0036280 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10653405000464), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CONMED 0036280?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Apr 20, 2016 and Sep 24, 2021. In those reports, the most common specific device problems are “Delivered as Unsterile Product” (3), “Device Packaging Compromised” (3) and “Unsealed Device Packaging” (3); the most common patient problem is “Not Applicable” (5). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CONMED 0036280?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CONMED 0036280 FDA approved?
Devices of this type (product code BYY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Conmed Corporation.
Who makes the CONMED 0036280, and is it still sold?
The FDA's device database lists Conmed Corporation as the company responsible for it (brand CONMED, model 0036280). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other TUBE, ASPIRATING, FLEXIBLE, CONNECTING devices
- DeRoyal 71-C506Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall4 reports
- DeRoyal 71-C512Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall2 reports
- DeRoyal 71-C520Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall1 report
- DeRoyal 71-C501Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
- DeRoyal 71-C510Deroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
- DeRoyal 71-C7100NSDeroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
- DeRoyal 71-C6100NSDeroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
- DeRoyal 71-C5100NSDeroyal Industries, Inc. · TUBE, ASPIRATING, FLEXIBLE, CONNECTING1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page