ETHICON M656GRecalls & reported problems
Ethicon, LLC · Monofilament 316L Stainless Steel, Nonabsorbable Surgical Suture · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10705031045507), received between Nov 29, 2018 and Jun 14, 2024 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
10 other recalls cover devices cleared under the same FDA submission (K170767, K946173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Material Separation3
- Break2
Patient problems
- No Consequences Or Impact To Patient7
- Bone Fracture(s)1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryJun 14, 2024Break; Break
IT WAS REPORTED THAT A PATIENT SUFFERED A RIGHT PATELLAR FRACTURE DUE TO A FALL INJURY AND UNDERWENT AN OPEN REDUCTION AND INTERNAL FIXATION OF RIGHT PATELLAR FRACTURE UNDER GENERAL ANESTHESIA ON (B)(6) 2024 AND SUTURE WAS USED. PRIOR TO THE SURGERY, THE PATIENT WAS INFORMED THAT THE RIGHT KNEE JOINT SHOULD NOT BE EXCESSIVELY FLEXED OR MOVED, AND THAT CRUTCHES SHOULD BE USED WHEN MOVING ON THE GROUND TO MAINTAIN AN EXTENDED POSITION OF THE RIGHT KNEE JOINT. THE RIGHT THIGH SHOULD NOT BE SUBJECTED TO FORCE MOVEMENTS. THE PATIENT MAY HAVE MISUNDERSTOOD POSTOPERATIVE ACTIVITIES, AND MAY NOT BE ACCUSTOMED TO USING THE RESTROOM IN BED. WHEN USING THE RESTROOM ON THE GROUND, THEY MAY HAVE A PARTNER IN THEIR 70S SUPPORTING THEM. AS A RESULT, IT IS IMPOSSIBLE TO ENSURE THAT THEIR RIGHT THIGH DOES NOT EXERT EXCESSIVE FORCE DURING ACTIVITIES, LEADING TO PREMATURE AND EXCESSIVE MOVEMENT . ON THE AFTERNOON OF THE 5TH DAY AFTER SURGERY, THE PATIENT FELT A POPPING SOUND IN THE RIGHT KNEE JOINT ACCOMPANIED BY PAIN. ON THE 6TH DAY AFTER SURGERY, A FOLLOW-UP X-RAY WAS ISSUED. DUE TO WEATHER CONDITIONS, THE PATIENT TOOK A X-RAY BEFORE THE 7TH DAY AFTER SURGERY, WHICH SHOWED THAT THE TENSION BAND STE
Patient problems: Bone Fracture(s)
MAUDE report 19538892 ↗MalfunctionDec 17, 2020Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN OPEN REDUCTION AND INTERNAL FIXATION FOR COMMINUTED FRACTURE OF LEFT PATELLA PROCEDURE ON (B)(6) 2018 AND SUTURE WAS USED. THE PATIENT RECOVERED WELL. THEN REHABILITATION EXERCISE WAS RECOMMENDED. ON (B)(6) 2018 AT THE TIME OF REEXAMINATION, A STRAND OF SUTURE WAS FOUND BROKEN. THE PATIENT WAS ADVISED TO REDUCE THE INTENSITY OF REHABILITATION EXERCISE. WHEN THE FRACTURE HEALED, THE SURGERY OF REMOVAL OF INTERNAL FIXATION DEVICE FOR LEFT PATELLA FRACTURE WAS PERFORMED ON (B)(6) 2020. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11030803 ↗MalfunctionDec 18, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. THE NEEDLE PULLED OFF THE SUTURE DURING USE. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. CHANGED ANOTHER ONE TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9490971 ↗MalfunctionDec 18, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. THE NEEDLE PULLED OFF THE SUTURE DURING USE. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. CHANGED ANOTHER ONE TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9490963 ↗MalfunctionApr 26, 2019Material Separation; Material Separation
IT WAS REPORTED THAT THE PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. THE NEEDLE PULLED OFF THE SUTURE DURING USE. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. CHANGED ANOTHER ONE TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8556185 ↗MalfunctionNov 29, 2018Detachment of Device or Device Component
NEEDLE PULL OFF. IT OCCURRED NEEDLE PULL OFF ISSUE DURING SEWING STERNUM WITH FIRST STITCH IN THE CLOSING THE STERNUM IN EXTRACARDIAC SURGERY. 4 STEEL SUTURES HAS THE SIMILAR ISSUE IN THIS ONE PATIENT, PLEASE HAVE A DEEP INVESTIGATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8117649 ↗MalfunctionNov 29, 2018Detachment of Device or Device Component
IT WAS REPORTED THAT A PATIENT UNDERWENT A STERNUM CLOSURE IN AN EXTRACARDIAC SURGERY PROCEDURE ON (B)(6) 2018 AND A STEEL SUTURE WAS USED. DURING SURGERY, NEEDLE PULL OFF ISSUE WHILE SEWING STERNUM WITH FIRST STITCH. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8117645 ↗MalfunctionNov 29, 2018Detachment of Device or Device Component
IT WAS REPORTED THAT A PATIENT UNDERWENT A STERNUM CLOSURE IN AN EXTRACARDIAC SURGERY PROCEDURE ON (B)(4) 2018 AND A STEEL SUTURE WAS USED. DURING SURGERY, NEEDLE PULL OFF ISSUE WHILE SEWING STERNUM WITH FIRST STITCH. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8117644 ↗MalfunctionNov 29, 2018Detachment of Device or Device Component
IT WAS REPORTED THAT A PATIENT UNDERWENT A STERNUM CLOSURE IN AN EXTRACARDIAC SURGERY PROCEDURE ON (B)(6) 2018 AND A STEEL SUTURE WAS USED. DURING SURGERY, NEEDLE PULL OFF ISSUE WHILE SEWING STERNUM WITH FIRST STITCH. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8117642 ↗
FDA clearance and classification
Cleared through Special 510(k) K170767 on Nov 1, 2017 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Surgical Stainless Steel Suture, Stainless Steel Suture”. The FDA review took 232 days. FDA record ↗
Cleared through 510(k) K946173 on Jan 9, 1995 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Stainless Steel, Gut, Poly (Glycolide/L-Lactide) Polyamide, Polypropylene, Poly (Ethylene Terephthalate) And Silk”. The FDA review took 21 days. FDA record ↗
The FDA regulates product code GAQ (suture, nonabsorbable, steel, monofilament and multifilament, sterile) under 21 CFR 878.4495 as a Class II device. 21 CFR 878.4495 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K170767, K946173. One submission often covers many models, so check the product and lot details in each record.
- Z-0284-2027 · Sep 23, 2026 · Open, ClassifiedETHILON Nylon Suture, Product Code 697G
- Z-0153-2027 · Sep 8, 2026 · Open, ClassifiedPERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H
- Z-1516-2025 · Mar 7, 2025 · Open, ClassifiedETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
- Z-1097-2025 · Dec 20, 2024 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
- Z-2486-2020 · Jun 10, 2020 · Open, ClassifiedPERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
- Z-1841-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
- Z-1842-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
- Z-0747-2019 · Dec 18, 2018 · TerminatedPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
- Z-1839-2015 · May 8, 2015 · TerminatedEthicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
- Z-0492-2015 · Oct 17, 2014 · TerminatedEthicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Common questions about the ETHICON M656G
Has the ETHICON M656G been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705031045507), based on the FDA recall file dated Nov 27, 2026. However, 10 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ETHICON M656G?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Nov 29, 2018 and Jun 14, 2024. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (4), “Material Separation” (3) and “Break” (2); the most common patient problem is “Bone Fracture(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ETHICON M656G?
None of the reports are classified as deaths, 1 as an injury and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ETHICON M656G FDA approved?
It was cleared through 510(k) K946173 on Jan 9, 1995; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the ETHICON M656G MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the ETHICON M656G, and is it still sold?
The FDA's device database lists Ethicon, LLC as the company responsible for it (brand ETHICON, model M656G). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE devices
- ETHICON M660GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE100 reports
- ETHICON M650GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE67 reports
- ETHIBOND EXCEL X425HEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall12 reports
- Steel 88862224-49Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall3 reports
- ETHICON M655GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE46 reports
- ETHIBOND EXCEL MB66GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE40 reports
- Steel 88862424-89Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE39 reports
- ETHICON M649GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE35 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.