PDS PDP317HRecalls & reported problems

Ethicon, LLC · Antibacterial Violet Monofilament (Polydioxanone), Absorbable Surgical Suture · Class 2 · Implant

Quick summary

This Ethicon PDS absorbable suture is an antibacterial polydioxanone surgical thread. The 40 reports mainly describe the thread breaking or separating, including needle detachment, during procedures. They mention no deaths and 5 injuries. No recalls name this device. Reports are 0 in the last 12 months, compared with 7 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 10705031048034), received between Jun 25, 2018 and Oct 2, 2025 — classified as 5 injuries, 35 malfunctions.

The most frequently reported specific device problem is “Break” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 6 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break25
  • Material Separation12
  • Adverse Event Without Identified Device or Use Problem5
  • Difficult to Open or Remove Packaging Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • No Consequences Or Impact To Patient10
  • Fluid Discharge5
  • Impaired Healing5
  • Inflammation5
  • Wound Dehiscence5
  • Hernia1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2025Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN ANIMAL PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT SNAPPING SUTURE CAUSED PET TO HERNIATE THE ABDOMEN. SUTURE WAS BREAKING CAUSING THE PET TO HAVE SECONDARY SURGERY. DEVICE IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Hernia

    MAUDE report 23199387 ↗
  • MalfunctionMay 5, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PERITONEAL CLOSURE IN CESAPE (VIA LSK) PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE THREAD IMMEDIATELY BROKE UNDER TENSION (AT THE KNOT OR WITH CONTINUOUS SUTURES). DURING USE ON PATIENT. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED. IT COST 15 EXTRA MINUTES OF SURGERY DURATION. PROCEDURE FINISHED WITH SAME LIKE DEVICE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21960509 ↗
  • MalfunctionMay 5, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PERITONEAL CLOSURE IN CESAPE (VIA LSK) PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE THREAD IMMEDIATELY BROKE UNDER TENSION (AT THE KNOT OR WITH CONTINUOUS SUTURES). DURING USE ON PATIENT. NO PATIENT HARM NOR SIGNIFICANT DELAY REPORTED. IT COST 15 EXTRA MINUTES OF SURGERY DURATION. PROCEDURE FINISHED WITH SAME LIKE DEVICE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21960499 ↗
  • MalfunctionApr 17, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A SLEEVE GASTRECTOMY ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, AT THE TIME OF REINFORCING THE STAPLING LINE, THE SUTURE WAS PULLED AND THE NEEDLE FLEW OUT, THUS DISASSEMBLING THE THREAD NEEDLE. THE PROCEDURE SUCCESSFULLY COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21862399 ↗
  • MalfunctionApr 17, 2025Break; Break

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT A SURGICAL PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21861987 ↗
  • MalfunctionMar 18, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE SUTURE IS CUT DURING SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21633865 ↗
  • MalfunctionJan 28, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A SLEEVE GASTRECTOMY PROCEDURE ON (B)(6)2025 AND SUTURE WAS USED. DURING SUTURE REINFORCEMENT, THE THREAD IS RELEASED FROM THE NEEDLE, A NEW SUTURE IS PASSED OUT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21258228 ↗
  • MalfunctionOct 24, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE THREAD IS CUT REPEATEDLY WHEN USED DURING SURGERY. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20522570 ↗
  • MalfunctionOct 24, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE THREAD IS CUT REPEATEDLY WHEN USED DURING SURGERY. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20522558 ↗
  • MalfunctionOct 24, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE THREAD IS CUT REPEATEDLY WHEN USED DURING SURGERY. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20522410 ↗
  • MalfunctionOct 24, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE THREAD IS CUT REPEATEDLY WHEN USED DURING SURGERY. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20522403 ↗
  • MalfunctionOct 23, 2024Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE THREAD IS CUT REPEATEDLY WHEN USED DURING SURGERY, SURGEONS. THERE WERE NOT PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20514031 ↗
  • MalfunctionOct 23, 2024Break; Break

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE THREAD IS CUT REPEATEDLY WHEN USED DURING SURGERY, SURGEONS. THERE WERE NOT PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20513846 ↗
  • MalfunctionAug 20, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A SLEEVE GASTRECTOMY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE IS INTRODUCED THROUGH THE TROCAR. WHEN THE NEEDLE IS BEING RETRIEVED TO PASS THE FIRST STITCH, THE NEEDLE REMAINS MOUNTED ON THE NEEDLE HOLDER AND THE THREAD COMES APART. A NEW SUTURE IS PASSED TO CONTINUE THE PROCEDURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20001958 ↗
  • MalfunctionMar 28, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A SMALL BOWEL RESECTION ON (B)(6) 2024 AND SUTURE WAS USED. THE THREAD CAME LOOSE FROM THE NEEDLE AND THE THREAD BROKE DURING SUTURING. A NEW SUTURE WAS TAKEN AND PLACED. THE PROCEDURE WAS PROLONGED BY 8 MINUTES AS A RESULT OF THE EVENT. THERE WAS NO PATIENT CONSEQUENCE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18993828 ↗
  • MalfunctionMar 28, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A SMALL BOWEL RESECTION ON (B)(6) 2024 AND SUTURE WAS USED. THE THREAD CAME LOOSE FROM THE NEEDLE AND THE THREAD BROKE DURING SUTURING. A NEW SUTURE WAS TAKEN AND PLACED. THE PROCEDURE WAS PROLONGED BY 8 MINUTES AS A RESULT OF THE EVENT. THERE WAS NO PATIENT CONSEQUENCE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18991170 ↗
  • MalfunctionOct 12, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING SURGERY SEVERAL NEEDLES CAME APART FROM THE SUTURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17927406 ↗
  • MalfunctionOct 12, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT A GYN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE THEY WERE GOING THROUGH TOUGH SKIN AND A COUPLE OF THE NEEDLES BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17927360 ↗
  • InjuryOct 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ABDOMINOPLASTY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED ON THE ABDOMINAL FLAP DERMAL PLANE. IN ABOUT 3 WEEKS TO 2 MONTHS, THE PATIENT DEVELOPED A SMALL INCREASE IN VOLUME IN THE SCAR AREA, PAIN AND SUBSEQUENT LEAKAGE OF VISCOUS FLUID AND WOUND DESHICENCE IN THE LATERAL AREA WITHOUT TENSION. THE SUTURE WAS EXPOSED AND THE INFLAMMATION SUBSIDED AS THE SUTURE MATERIAL WAS REMOVED. THESE PHENOMENA WERE REPEATED MANY TIMES IN THE SAME PATIENT IN DIFFERENT AREAS OF THE OPERATIVE WOUND, HAVING TO REMOVE THE SUTURE MATERIAL REPEATEDLY. THE PATIENT WAS MANAGED WITH ADVANCED HEALING UNTIL CLOSURE.

    Patient problems: Fluid Discharge; Impaired Healing; Inflammation; Wound Dehiscence

    MAUDE report 17860647 ↗
  • InjuryOct 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ABDOMINOPLASTY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED ON THE ABDOMINAL FLAP DERMAL PLANE. IN ABOUT 3 WEEKS TO 2 MONTHS, THE PATIENT DEVELOPED A SMALL INCREASE IN VOLUME IN THE SCAR AREA, PAIN AND SUBSEQUENT LEAKAGE OF VISCOUS FLUID AND WOUND DEHISCENCE IN THE LATERAL AREA WITHOUT TENSION. THE SUTURE WAS EXPOSED AND THE INFLAMMATION SUBSIDED AS THE SUTURE MATERIAL WAS REMOVED. THESE PHENOMENA WERE REPEATED MANY TIMES IN THE SAME PATIENT IN DIFFERENT AREAS OF THE OPERATIVE WOUND, HAVING TO REMOVE THE SUTURE MATERIAL REPEATEDLY. THE PATIENT WAS MANAGED WITH ADVANCED HEALING UNTIL CLOSURE.

    Patient problems: Fluid Discharge; Impaired Healing; Inflammation; Wound Dehiscence

    MAUDE report 17860641 ↗
  • InjuryOct 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ABDOMINOPLASTY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED ON THE ABDOMINAL FLAP DERMAL PLANE. IN ABOUT 3 WEEKS TO 2 MONTHS, THE PATIENT DEVELOPED A SMALL INCREASE IN VOLUME IN THE SCAR AREA, PAIN AND SUBSEQUENT LEAKAGE OF VISCOUS FLUID AND WOUND DEHISCENCE IN THE LATERAL AREA WITHOUT TENSION. THE SUTURE WAS EXPOSED AND THE INFLAMMATION SUBSIDED AS THE SUTURE MATERIAL WAS REMOVED. THESE PHENOMENA WERE REPEATED MANY TIMES IN THE SAME PATIENT IN DIFFERENT AREAS OF THE OPERATIVE WOUND, HAVING TO REMOVE THE SUTURE MATERIAL REPEATEDLY. THE PATIENT WAS MANAGED WITH ADVANCED HEALING UNTIL CLOSURE.

    Patient problems: Fluid Discharge; Impaired Healing; Inflammation; Wound Dehiscence

    MAUDE report 17860639 ↗
  • InjuryOct 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ABDOMINOPLASTY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED ON THE ABDOMINAL FLAP DERMAL PLANE. IN ABOUT 3 WEEKS TO 2 MONTHS, THE PATIENT DEVELOPED A SMALL INCREASE IN VOLUME IN THE SCAR AREA, PAIN AND SUBSEQUENT LEAKAGE OF VISCOUS FLUID AND WOUND DEHISCENCE IN THE LATERAL AREA WITHOUT TENSION. THE SUTURE WAS EXPOSED AND THE INFLAMMATION SUBSIDED AS THE SUTURE MATERIAL WAS REMOVED. THESE PHENOMENA WERE REPEATED MANY TIMES IN THE SAME PATIENT IN DIFFERENT AREAS OF THE OPERATIVE WOUND, HAVING TO REMOVE THE SUTURE MATERIAL REPEATEDLY. THE PATIENT WAS MANAGED WITH ADVANCED HEALING UNTIL CLOSURE.

    Patient problems: Fluid Discharge; Impaired Healing; Inflammation; Wound Dehiscence

    MAUDE report 17860638 ↗
  • InjuryOct 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ABDOMINOPLASTY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. TWO MONTHS AFTER ABDOMINOPLASTY, FIRST, THE AREA IS INFLAMED AND CLOUDY SEROUS FLUID BEGINS TO EMERGE. THEN, THE SUTURE MATERIAL IS EXPOSED AND THE WOUND IS OPENED WITH DEHISCENCE. PROLONGED CURES, ANTIBIOTIC THERAPY, SUMMARIZATION, VACUUM HEALING ARE PERFORMED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Fluid Discharge; Impaired Healing; Inflammation; Wound Dehiscence

    MAUDE report 17860634 ↗
  • MalfunctionAug 8, 2023Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. SUTURE ARE BECOMING DETACHED FROM THE NEEDLE AFTER 3-4 THROWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17481810 ↗
  • MalfunctionAug 8, 2023Material Separation; Material Separation

    IT WAS REPORTED AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2023 AND SUTURE WAS USED. SUTURE ARE BECOMING DETACHED FROM THE NEEDLE AFTER 3-4 THROWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17481765 ↗

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →