Skip to content

PDSRecalls & reported problems across all models

Ethicon, LLC · 224 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

7 open recalls name at least one PDS model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
2,635
Deaths reported
1
Injuries reported
170
Recalls
10

We found 224 models of PDS by Ethicon, LLC in the FDA's device registry (GUDID). Together they are named in 2,635 adverse event reports in the FDA's MAUDE database, received between Nov 27, 2017 and Sep 30, 2026 — classified as 1 death, 170 injuries, 2,464 malfunctions.

The most frequently reported specific device problem is “Break” (1,694 reports).

10 FDA recall records name at least one of these models by its device identifier, 7 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 19 reportsDec 2023: 23 reportsJan 2024: 33 reportsFeb 2024: 37 reportsMar 2024: 54 reportsApr 2024: 40 reportsMay 2024: 21 reportsJun 2024: 46 reportsJul 2024: 13 reportsAug 2024: 36 reportsSep 2024: 32 reportsOct 2024: 30 reportsNov 2024: 45 reportsDec 2024: 33 reportsJan 2025: 51 reportsFeb 2025: 18 reportsMar 2025: 32 reportsApr 2025: 28 reportsMay 2025: 37 reportsJun 2025: 47 reportsJul 2025: 19 reportsAug 2025: 31 reportsSep 2025: 11 reportsOct 2025: 29 reportsNov 2025: 28 reportsDec 2025: 40 reportsJan 2026: 46 reportsFeb 2026: 24 reportsMar 2026: 35 reportsApr 2026: 33 reportsMay 2026: 21 reportsJun 2026: 11 reportsJul 2026: 5 reportsAug 2026: 20 reportsSep 2026: 28 reportsOct 2026: 0 reports20242025202654
Last 36 months. The most recent month may be incomplete.

PDS recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Mar 7, 2025Open, Classified

    Identified curing issues with the silicone during the needles manufacturing process.

    Z-1514-2025 · Root cause: Process control · Recall details · FDA record ↗

  2. Mar 7, 2025Open, Classified

    Identified curing issues with the silicone during the needles manufacturing process.

    Z-1513-2025 · Root cause: Process control · Recall details · FDA record ↗

  3. May 10, 2024Open, Classified

    Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .

    Z-2147-2024 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  4. May 10, 2024Open, Classified

    Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .

    Z-2146-2024 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗

  5. Aug 10, 2023Open, Classified

    Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.

    Z-2617-2023 · Root cause: Under Investigation by firm · names 22 of these models · Recall details · FDA record ↗

  6. Aug 10, 2023Open, Classified

    Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.

    Z-2616-2023 · Root cause: Under Investigation by firm · names 8 of these models · Recall details · FDA record ↗

  7. Nov 16, 2022Open, Classified

    knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage

    Z-0480-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Jan 10, 2018Terminated

    Three (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.

    Z-1340-2018 · Root cause: Process control · Recall details · FDA record ↗

  9. Jan 10, 2018Terminated

    Three (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.

    Z-1339-2018 · Root cause: Process control · Recall details · FDA record ↗

  10. Jan 10, 2018Terminated

    Three (3) lots were distributed with the incorrect expiry date on the label. The lots were labeled with an expiry date of five (5) years from date of manufacture instead of the correct expiry date of two (2) years from date of manufacture.

    Z-1338-2018 · Root cause: Process control · Recall details · FDA record ↗

Most reported PDS problems

Device problems

  • Break1,694
  • Material Separation707
  • Adverse Event Without Identified Device or Use Problem104
  • Component Misassembled28
  • Device Contaminated During Manufacture or Shipping25
  • Device Markings/Labelling Problem25
  • Detachment Of Device Component24
  • Defective Component20

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,802
  • No Consequences Or Impact To Patient591
  • Wound Dehiscence107
  • Unspecified Infection47
  • Insufficient Information35
  • Not Applicable29
  • Foreign Body In Patient27
  • Pain23

PDS models

224 models, 100 per page, most reported first. Each model has its own page with its reports, recalls and device details.

Recent PDS adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Ethicon, LLC products

All Ethicon, LLC devices →

Sources and updates