PDS PDP513GRecalls & reported problems
Ethicon, LLC · Antibacterial Violet Monofilament (Polydioxanone), Absorbable Surgical Suture · Class 2 · Implant
All models of PDS (222)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon PDS absorbable surgical suture is a violet monofilament thread used to close tissue during surgery. Its 43 reports mostly describe the suture breaking during use, with a few mentioning the needle detaching or the thread separating. No deaths were reported; one injury was. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 43
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 10705031048485), received between Apr 10, 2018 and Apr 30, 2026 — classified as 1 injury, 42 malfunctions.
The most frequently reported specific device problem is “Break” (36 reports).
9 other recalls cover devices cleared under the same FDA submission (K061037, K212380); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break36
- Material Separation4
- Adverse Event Without Identified Device or Use Problem1
- Component Incompatible1
- Defective Component1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions27
- No Consequences Or Impact To Patient15
- Inflammation1
- Pain1
Recent adverse event reports
The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 30, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN SURGERY, THE SUTURE WAS BROKEN DURING THE SECOND LIGATION OF THE DERMAL SUTURE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25046269 ↗MalfunctionFeb 27, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. THE SUTURE WAS BROKEN DURING USE, AND IT OCCURRED 2 TIMES IN A ROW. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24475303 ↗MalfunctionFeb 27, 2026Break
AND SUTURE WAS USED. THE SUTURE WAS BROKEN DURING USE, AND IT OCCURRED 2 TIMES IN A ROW. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24475289 ↗MalfunctionJun 16, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE SUTURE WAS BROKEN DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22213852 ↗MalfunctionNov 11, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE DOCTOR FELT THE NEEDLE TIP WAS NOT SHARP ENOUGH DURING SUTURING. WHEN THE NEEDLE WAS REMOVED, IT WAS FOUND THAT THE NEEDLE TIP WAS MISSING AND GONE. A VISUAL SEARCH WAS CONDUCTED, BUT THE NEEDLE TIP WAS NOT FOUND, SO X-RAYS WERE TAKEN. AS A RESULT, THE NEEDLE TIP WAS SUCCESSFULLY FOUND ON THE DRAPE OUTSIDE OF THE PATIENT¿S BODY. IT WAS NOT A CONTROL RELEASE NEEDLE. THE PRODUCT WAS USED ON SUBCUTANEOUS DERMIS. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20655038 ↗MalfunctionSep 23, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN OB-GYN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING USE IN AN UNKNOWN OB-GYN SURGERY, THE NEEDLE AND SUTURE DETACHED EASILY WHEN TENSION WAS APPLIED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20279001 ↗MalfunctionFeb 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18744151 ↗MalfunctionJan 25, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE WAS BROKEN DURING USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18579309 ↗MalfunctionAug 8, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2023 AND A SUTURE WAS USED. DURING SURGERY, WHEN THE NEEDLE-SUTURE WAS PASSING THROUGH THE SKIN, THE SUTURE WAS BROKEN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17484954 ↗MalfunctionJul 1, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. THE SUTURE BROKE DURING INTERRUPTED SUTURING ON THE DERMIS. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14885676 ↗MalfunctionMar 23, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN PAPILLECTOMY PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. THE SUTURE WAS BROKEN DURING SUTURING. NO FRAGMENT WAS LEFT IN THE PATIENT¿S BODY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13861433 ↗MalfunctionMar 14, 2022Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE LOOKED BENT. UPON THE VALUATION OF THE RETURNED SAMPLE, A FRACTURE WAS OBSERVED AT THE TIP OF THE NEEDLE. ONE SIDE OF THE NEEDLE WAS RECEIVED, THE MATING FRACTURE SURFACE WAS NOT PROVIDED FOR THIS EVALUATION. A SCANNING ELECTRON MICROSCOPE WAS USED TO EXAMINE THE FRACTURE SURFACES AND SURROUNDING AREA OF THE NEEDLE. THE FRACTURE SURFACES WERE EXAMINED IN MULTIPLE LOCATIONS TO DETERMINE THE FRACTURE MODE. THE EVALUATION OF (B)(4) REVEALED THE FRACTURE WAS COMPOSED OF MICROVOID COALESCENCE, WHICH IS EVIDENCE OF A DUCTILE FRACTURE MODE. THIS WAS A DUCTILE FRACTURE. THE EVIDENCE OF THIS EXAMINATION INDICATES THAT THE BREAKAGE OCCURRED AT THE TIP OF THE NEEDLE DURING USE DUE TO TENSILE OVERLOAD. THERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13764587 ↗MalfunctionMar 12, 2022Defective Component; Defective Component
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE NEEDLE TIP BROKE, ON (B)(6) 2022 ADDITIONAL INFORMATION RECEIVED CONFIRMING THAT THE NEEDLE DID NOT BROKE. UPON EVALUATION OF THE RETURNED SAMPLE, A FRACTURE WAS OBSERVED AT THE TIP OF THE NEEDLE. ONE SIDE OF THE NEEDLE WAS RECEIVED, THE MATING FRACTURE SURFACE WAS NOT PROVIDED FOR THIS EVALUATION. A SCANNING ELECTRON MICROSCOPE WAS USED TO EXAMINE THE FRACTURE SURFACES AND SURROUNDING AREA OF THE NEEDLE. THE FRACTURE SURFACES WERE EXAMINED IN MULTIPLE LOCATIONS TO DETERMINE THE FRACTURE MODE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13743785 ↗MalfunctionJan 14, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE DURING INTERRUPTED SUTURING. NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13257809 ↗MalfunctionJan 14, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE DURING INTERRUPTED SUTURING. NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13257805 ↗MalfunctionJan 4, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING PROCEDURE THE NURSE SET THE NEEDLE IN THE NEEDLE HOLDER AND LEFT IT THERE, BUT WHEN TRYING TO USE IT, THE TIP OF THE NEEDLE HAD BROKEN. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13161015 ↗MalfunctionSep 21, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE 3 CM FROM THE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12508427 ↗MalfunctionSep 3, 2021Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN SURGERY ON UNKNOWN DATE AND THE SUTURE WAS USED. IT WAS REPORTED THAT SUTURE BREAKAGE OCCURRED DURING CONTINUOUS SUTURING. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12422990 ↗MalfunctionAug 3, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN OTOLARYNGOLOGY HEAD AND NECK SURGERY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12268455 ↗MalfunctionAug 3, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN OTOLARYNGOLOGY HEAD AND NECK SURGERY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12268447 ↗MalfunctionAug 3, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN OTOLARYNGOLOGY HEAD AND NECK SURGERY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12268430 ↗MalfunctionJun 24, 2021Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A CESAREAN SURGERY ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, AFTER STARTING THE DERMAL SUTURE, THE SUTURE BROKE AROUND THE SWAGE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12057605 ↗InjuryJun 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PLASTIC SURGERY ON UNKNOWN DATE AND THE SUTURE WAS USED. AFTER SURGERY, THE PATIENT EXPERIENCED PAINFUL INFLAMMATION. IT WAS REPORTED THAT LOCAL NEGATIVE PRESSURE THERAPY, VAC THERAPY, OR PICO THERAPY CANNOT BE APPLIED ANYWHERE, BECAUSE WHEN ABDOMINAL SUTURE WAS PERFORMED THERE WAS AN INFLAMMATORY REACTION WITH SEVERE PAIN AFTER SURGERY. MRI ALSO CONFIRMED INFLAMMATION. THE PATIENT IS STILL HOSPITALIZED AT THE DEPARTMENT OF PLASTIC SURGERY. THE PATIENT IS SUSPECTED TO HAVE INTRACTABLE DISEASE. THE POSSIBILITY OF INTRACTABLE DISEASE IS HIGH AND CURRENTLY THE PATIENT VISITED VARIOUS MEDICAL DEPARTMENTS FOR EXAMINATIONS. AT PRESENT, THE PATIENT IS BEING TREATED AT THE PLASTIC SURGERY DEPARTMENT BUT THE CAUSE IS ALSO UNKNOWN AT THE THORACIC SURGERY DEPARTMENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Inflammation; Pain
MAUDE report 11941208 ↗MalfunctionApr 23, 2021Break; Break
IT WAS REPORTED BY A SALES REP THAT, DURING A TOTAL LAPAROSCOPIC HYSTERECTOMY, WHEN CLOSING THE SURGICAL WOUND, THE NEEDLE WAS BROKEN AT THE SWAGE PART. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. AFFILIATE REFERENCE NUMBER IS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11718683 ↗MalfunctionMar 25, 2021Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN SURGERY IN 2021 AND THE SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE BREAKAGE OCCURRED DURING SKIN CLOSURE. THERE WERE NO PATIENT CONSEQUENCES. ANOTHER LIKE SUTURE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11573019 ↗
FDA clearance and classification
Cleared through 510(k) K212380 on Feb 22, 2022 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “PDS Plus Antibacterial (Polydioxanone) Sterile Synthetic Absorable Surgical Suture”. The FDA review took 204 days. FDA record ↗
Cleared through 510(k) K061037 on Jul 14, 2006 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Pds Plus Antibacterial Suture”. The FDA review took 91 days. FDA record ↗
The FDA regulates product code NEW (suture, surgical, absorbable, polydioxanone) under 21 CFR 878.4840 as a Class II device. 21 CFR 878.4840 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K061037, K212380. One submission often covers many models, so check the product and lot details in each record.
- Z-2147-2024 · May 10, 2024 · Open, ClassifiedPDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in oph
- Z-1571-2023 · Mar 31, 2023 · Open, ClassifiedPDS Plus Antibacterial (polydioxanone) Suture - PDS PLUS CLR 27IN(70CM) USP3-0(M2) S/A PC-25 PRIME. Intended for use in general soft tissue approximation, including use in pediatri
- Z-1135-2019 · Mar 12, 2019 · TerminatedVICRYL PLUS Sutures: VICRYL PLUS SUTURE 18"(45CM) 3-0 VIO VICRYL PLUS SUTURE 18"(45CM) 0 UND VICRYL PLUS SUTURE 18"(45CM) 0 VIO VICRYL PLUS SUTURE 18"(45CM) 0 VIO) VICRYL PLUS SU
- Z-2533-2018 · Jun 8, 2018 · TerminatedPDS Plus Antibacterial Suture-PDS PLUS Undyed Suture 45cm (18") 5-0 Single Armed P-3 Prime Needle Product Code: PDP493G (GTIN 10705031203945)
- Z-2532-2018 · Jun 8, 2018 · TerminatedPDS Plus Antibacterial (Polydioxanone) Suture-PDS PLUS Undyed Suture 45cm (18") 4-0 Single Armed PS-4 Prime Needle Product Code: PDP507G (GTIN 10705031203983)
- Z-1340-2018 · Jan 10, 2018 · TerminatedPDS Plus Antibacterial (Polydioxanone) Suture Size 0 PDS Plus Violet 36 Single Armed CTB-1 Needle
- Z-1339-2018 · Jan 10, 2018 · TerminatedPDS Plus Antibacterial (Polydioxanone) Suture Size 1 PDS PLUS Violet 27 Single Armed OS-4 Needle
- Z-1338-2018 · Jan 10, 2018 · TerminatedPDS Plus Antibacterial (Polydioxanone) Suture Size 0 PDS Plus Violet 36 Single Armed CT-1 Needle
- Z-1314-2009 · Feb 2, 2009 · TerminatedPDS Plus Antibacterial (polydioxanone) Suture Sutures are intended for general soft tissue approximation and/or ligation.
Common questions about the PDS PDP513G
Has the PDS PDP513G been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705031048485), based on the FDA recall file dated Nov 27, 2026. However, 9 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the PDS PDP513G?
The FDA's MAUDE database lists 43 adverse event reports for this device, received between Apr 10, 2018 and Apr 30, 2026. In those reports, the most common specific device problems are “Break” (36), “Material Separation” (4) and “Component Incompatible” (1); the most common patient problems are “Inflammation” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PDS PDP513G?
None of the reports are classified as deaths, 1 as an injury and 42 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PDS PDP513G FDA approved?
It was cleared through 510(k) K061037 on Jul 14, 2006; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the PDS PDP513G MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the PDS PDP513G, and is it still sold?
The FDA's device database lists Ethicon, LLC as the company responsible for it (brand PDS, model PDP513G). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE611 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE478 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE442 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE442 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone400 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE334 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death230 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page