ETHIBOND EXCEL X423HRecalls & reported problems
Ethicon, LLC · Green Braided Polyester, Nonabsorbable Surgical Suture · Class 2 · Implant
All models of ETHIBOND EXCEL (133)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705031057258), received between Apr 27, 2023 and Apr 27, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
10 other recalls cover devices cleared under the same FDA submission (K946173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 27, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC HERNIA REPAIR PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE LAPAROSCOPIC HERNIA REPAIR, THE NEEDLE BROKE OFF INTRA ABDOMINAL AS THE SURGEON WAS TRYING TO TAKE IT OUT. HALF OF THE NEEDLE HAD TO BE FISHED OUT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16822745 ↗
FDA clearance and classification
Cleared through 510(k) K946173 on Jan 9, 1995 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Stainless Steel, Gut, Poly (Glycolide/L-Lactide) Polyamide, Polypropylene, Poly (Ethylene Terephthalate) And Silk”. The FDA review took 21 days. FDA record ↗
The FDA regulates product code GAQ (suture, nonabsorbable, steel, monofilament and multifilament, sterile) under 21 CFR 878.4495 as a Class II device. 21 CFR 878.4495 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K946173. One submission often covers many models, so check the product and lot details in each record.
- Z-0284-2027 · Sep 23, 2026 · Open, ClassifiedETHILON Nylon Suture, Product Code 697G
- Z-0153-2027 · Sep 8, 2026 · Open, ClassifiedPERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H
- Z-1516-2025 · Mar 7, 2025 · Open, ClassifiedETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
- Z-1097-2025 · Dec 20, 2024 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
- Z-2486-2020 · Jun 10, 2020 · Open, ClassifiedPERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
- Z-1841-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
- Z-1842-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
- Z-0747-2019 · Dec 18, 2018 · TerminatedPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
- Z-1839-2015 · May 8, 2015 · TerminatedEthicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
- Z-0492-2015 · Oct 17, 2014 · TerminatedEthicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Common questions about the ETHIBOND EXCEL X423H
Has the ETHIBOND EXCEL X423H been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705031057258), based on the FDA recall file dated Nov 27, 2026. However, 10 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ETHIBOND EXCEL X423H?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 27, 2023. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ETHIBOND EXCEL X423H?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ETHIBOND EXCEL X423H FDA approved?
It was cleared through 510(k) K946173 on Jan 9, 1995. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the ETHIBOND EXCEL X423H MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the ETHIBOND EXCEL X423H, and is it still sold?
The FDA's device database lists Ethicon, LLC as the company responsible for it (brand ETHIBOND EXCEL, model X423H). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE devices
- ETHICON M660GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE100 reports
- ETHICON M650GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE67 reports
- ETHIBOND EXCEL X425HEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall12 reports
- Steel 88862224-49Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE1 recall3 reports
- ETHICON M655GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE46 reports
- ETHIBOND EXCEL MB66GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE40 reports
- Steel 88862424-89Covidien LP · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE39 reports
- ETHICON M649GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE35 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page