PDS Z442HRecalls & reported problems
Ethicon Inc. · Clear Monofilament (Polydioxanone), Absorbable Surgical Suture · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705031060975), received between Oct 31, 2019 and Aug 30, 2024 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Break” (12 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Aug 10, 2023Open, Classified
Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.
Z-2617-2023 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Break12
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Local Reaction1
- Not Applicable1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO SALES REP BY SURGEON THAT EVERY SUTURE WE¿VE USED FROM THIS BOX HAS BROKEN WHILE TYING THE KNOT. DEVICE AVAILABILITY UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20107115 ↗MalfunctionAug 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO SALES REP BY SURGEON THAT EVERY SUTURE WE¿VE USED FROM THIS BOX HAS BROKEN WHILE TYING THE KNOT. DEVICE AVAILABILITY UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20107085 ↗MalfunctionAug 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO SALES REP BY SURGEON THAT EVERY SUTURE WE¿VE USED FROM THIS BOX HAS BROKEN WHILE TYING THE KNOT. DEVICE AVAILABILITY UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20107077 ↗MalfunctionAug 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO SALES REP BY SURGEON THAT EVERY SUTURE WE¿VE USED FROM THIS BOX HAS BROKEN WHILE TYING THE KNOT. DEVICE AVAILABILITY UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20107051 ↗MalfunctionAug 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO SALES REP BY SURGEON THAT EVERY SUTURE WE¿VE USED FROM THIS BOX HAS BROKEN WHILE TYING THE KNOT. DEVICE AVAILABILITY UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20107039 ↗MalfunctionAug 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO SALES REP BY SURGEON THAT EVERY SUTURE WE¿VE USED FROM THIS BOX HAS BROKEN WHILE TYING THE KNOT. DEVICE AVAILABILITY UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20106998 ↗MalfunctionAug 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO SALES REP BY SURGEON THAT EVERY SUTURE WE¿VE USED FROM THIS BOX HAS BROKEN WHILE TYING THE KNOT. DEVICE AVAILABILITY UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20106985 ↗MalfunctionAug 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO SALES REP BY SURGEON THAT EVERY SUTURE WE¿VE USED FROM THIS BOX HAS BROKEN WHILE TYING THE KNOT. DEVICE AVAILABILITY UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20106974 ↗MalfunctionAug 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO SALES REP BY SURGEON THAT EVERY SUTURE WE¿VE USED FROM THIS BOX HAS BROKEN WHILE TYING THE KNOT. DEVICE AVAILABILITY UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20106971 ↗MalfunctionMay 6, 2024Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE IN 2024 AND SUTURE WAS USED. THE NEEDLE BROKE DURING SURGERY. NO PATIENT HARM WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19247804 ↗MalfunctionMay 17, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN THE SURGEON SEAMED IN THE MIDDLE, THE SUTURE BROKE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11838196 ↗InjuryOct 31, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT OTOPLASTY ON (B)(6) 2019 AND SUTURE WAS USED FOR PLICATION AND ATTACHMENT OF CONCHA. IN (B)(6) 2019, TWO EARS GOT PEELED OFF. THE PATIENT WAS OPERATED AGAIN ON (B)(6) 2019 AND NO SUTURE WAS FOUND EXCEPT A 3.0 ABSORBABLE SUTURE ON THE RIGHT CONCHA. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Local Reaction; Not Applicable
MAUDE report 9261574 ↗
Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE617 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE485 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE446 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE443 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone406 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE339 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death229 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.