PDS II (polydioxanone) Suture, Product Numbers Z844G-50… recallZ-2617-2023 · Ethicon, LLC · initiated Aug 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ethicon, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT: MEDICAL DEVICE RECALL (REMOVAL) notification letter dated 8/10/23 was sent to customers. ACTION REQUIRED 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). If you have product subject to this recall, please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email to Sedgwick at 877-550-0702 or [email protected] within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 4. Customers are required to return unused PDS" Sutures subject to this recall that are in inventory immediately. To receive credit reimbursement, customers must return product subject to this recall no later than November 30, 2023. Any non-affected product and any product returned after the date specified will not receive credit reimbursement. 5. To return product subject to this recall, photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for the shipping charges only if the authorized label is used. Extra shipping labels may be obtained by calling Sedgwick at 877-479-8088. Your account number and mailing address have been pre-populated on the BRF. 6. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Sedgwick. 7. If product subject to this recall is contained in a custom kit, please contact you
Recall details
- Recall number
- Z-2617-2023
- Status
- Open, Classified
- Initiated by the company
- Aug 10, 2023
- Posted by the FDA
- Sep 30, 2024
- Recalling firm
- Ethicon, LLC
- Quantity
- 868,363 total units
- Product code
- NEW
- Distribution
- US Nationwide. Global Distribution.
Product description
PDS II (polydioxanone) Suture, Product Numbers Z844G-50, D10065-05, D10085-05, D6776-05, D6776-05, D9297-05, D9297-05, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z421H-50, Z421H-50, Z421H-50, Z422H-50, Z422H-50, Z422H-50, Z422H-50, Z422H-50, Z422H-50, Z422H-50, Z423H-50, Z423H-50, Z423H-50, Z442H-50, Z442H-50, Z442H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z489G-50, Z489G-50, Z489G-50, Z489G-50, Z493G-50, Z493G-50, Z493G-50, Z493G-50, Z493G-50, Z494G-50, Z495G-05, Z495G-05, Z495G-05, Z496G-50, Z496G-50, Z496G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z503G-50, Z504G-50, Z507G-50, Z507G-50, Z683G-05, Z833G-05, Z833G-05, Z844G-50
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
UDI-DI: 10705031026742, 10705031031234, 10705031032620, 10705031033894, 10705031134386, 10705031134492, 10705031118102, 10705031073487, 10705031149410, 10705031149427, 10705031075443, 10705031149441, 10705031149472, 10705031086906, 10705031130296, 10705031130302, 10705031082960, 10705031151819, 10705031130333, 10705031130357, 10705031130357, 10705031130418, 10705031130418, 10705031083028, 10705031130425, 10705031130562, 10705031151895, 10705031151895, 10705031151901, 10705031151918, 10705031130708, 10705031130722, 10705031131057, 10705031131057, 10705031461536, 10705031060340, 10705031113602, 10705031113619, 10705031060852, 10705031114050, 10705031060876, 10705031060883, 10705031114067, 10705031060890, 10705031060906, 10705031114074, 10705031114081, 10705031060913, 10705031114111, 10705031060968, 10705031060975, 10705031060982, 10705031114135, 10705031060999, 10705031061002, 10705031114142, 10705031114258, 10705031061132, 10705031061149, 10705031114265, 10705031114272, 10705031061156, 10705031061163, 10705031114296, 10705031061170, 10705031061187, 10705031061194, 10705031114319, 10705031061217, 10705031061224, 10705031114333, 10705031061231, 10705031467217, 10705031061248, 10705031114340, 10705031061255, 10705031061262, 10705031061279, 10705031061293, 10705031061309, 10705031061323, 10705031061347, 10705031061385, 10705031061460, 10705031061507, 10705031061538, 10705031061545, 10705031114654, 10705031061828, 10705031061835, 10705031061842, 10705031467354; Lot Numbers: RGMBEK, QMMCKS, QLMKHK, REMMRB, RKMASZ, PBM649, SJMLRE, PG6842, QJMCQC, QLMBXL, QKMKRK, QLMDJA, QLMJQH, QMMAKJ, QMMLBP, QPMCMQ, RCMJDB, RCMPXB, RDMERH, RGMDMM, RGMDBJ, SCMJET, REMJAJ, MKZ221, QPMDDQ, QPMDHD, RBMAJE, RGMAQU, RHMBMR, RLMPMM, QGMHTJ, QKMMPQ, RCMKHS, MMK957, QMMCPS, RDMKAJ, REMKDP, QHMMPR, RKMAAS, SGMCTP, REMBUE, RKMMTD, SCMAXS, QHMEDK, REMEQC, RJMPQP, QKMHTR, QLMKPU, QMMHUU, RAMLZJ, RDMACU, RDMECL, RGMLCS, SCMELT, QBMBQA, QBMKCR, RPMBJS, SBMJZM, QLMDDC, QLMPAR, QPMBDT, RGMBBK, RHMJHU, QHMMEP, QLMDDQ, QMMBQJ, RCMRUD, RGMALA, RCMBDB, TBMDPK, PGZ426, QKMHTM, QKMHRP, QLMHBX, QPMBMD, QPMDBX, RCMJEK, RDMDAS, RKMDJA, RLMHPQ, QPMCAD, RGMDBP, QCMCZZ, RJMLEL, TAMQSS, MPZ365, SCMBJZ, RBMEDE
Devices named in this recall
28 devices in the FDA device registry whose identifier appears in the recall's code information.
- PDS Z494GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall159 reports
- PDS Z493GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall56 reports
- PDS Z496GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall47 reports
- PDS Z507GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall17 reports
- PDS Z497GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall16 reports
- PDS Z423HEthicon Inc. · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall12 reports
- PDS Z442HEthicon Inc. · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall12 reports
- PDS Z503GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall12 reports
- PDS Z422HEthicon Inc. · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall11 reports
- PDS Z259HEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall8 reports
- PDS Z489GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall7 reports
- PDS Z509GEthicon, LLC · Suture, surgical, absorbable, polydioxanone1 recall7 reports
- PDS Z495GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall6 reports
- PDS D8790Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall4 reports
- PDS Z443HEthicon Inc. · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall3 reports
- PDS Z416HEthicon Inc. · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall1 report
- PDS Z516GEthicon, LLC · Suture, surgical, absorbable, polydioxanone1 recall1 report
- PDS Z844GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall1 report
- PDS D6776Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
- PDS D9297Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
- PDS D9632Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
- PDS Z421HEthicon Inc. · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
- PDS Z504GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
- PDS Z663HEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
- PDS Z683GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
- PDS Z833GEthicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
- PDS D10065Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
- PDS D10085Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92843).
- Z-2616-2023 · Open, ClassifiedPDS Plus Antibacterial (polydioxanone) Suture, Product Numbers PDP072H40, PDP259H50, PDP416H50, PDP441H50, PDP442H50, PDP489G50, PDP489H40, PDP497G51, PDP497H16, PDP497H40, PDP498H…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page