PDS Z990GRecalls & reported problems
Ethicon, LLC · Violet Monofilament (Polydioxanone), Absorbable Synthetic Surgical Suture · Class 2 · Implant
All models of PDS (224)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon PDS Z990G is an absorbable surgical suture (violet polydioxanone monofilament) used to close tissue. The 32 reports mostly describe the suture or its needle breaking during surgery — in one case a needle tip broke off and an X-ray was used to find the fragment — plus a few mentions of the material separating or twisting. No deaths and no injuries were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 10705031061989), received between Aug 7, 2018 and May 29, 2025 — classified as 32 malfunctions.
The most frequently reported specific device problem is “Break” (30 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break30
- Material Separation2
- Material Twisted/Bent2
Patient problems
- No Clinical Signs, Symptoms or Conditions25
- No Consequences Or Impact To Patient7
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 29, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED WHILE CLOSING THE FACIA OF A C-SECTION THAT WHEN THE SUTURE WAS HANDED BACK TO THE TECH IT WAS NOTICED THAT THE TIP OF THE NEEDLE HAD BROKEN OFF. X-RAY WAS USED TO FIND AND RECOVER THE FRAGMENT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22105656 ↗MalfunctionApr 29, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT DURING A LABOR AND DELIVERY CASE, A PDS LOOPED SUTURE BROKE AT THE NEEDLE SEVERAL TIMES. SO MUCH SO THE SURGEON HAD THE SUTURE REMOVED FROM THE SHELF AND FINISHED WITH THE ZERO VICRYL. NO PATIENT HARM WAS REPORTED. DEVICE IS AVIALALBE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21926150 ↗MalfunctionApr 29, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY THE SALES REP THAT DURING A LABOR AND DELIVERY CASE, A PDS LOOPED SUTURE BROKE AT THE NEEDLE SEVERAL TIMES. SO MUCH SO THE SURGEON HAD THE SUTURE REMOVED FROM THE SHELF AND FINISHED WITH THE ZERO VICRYL. NO PATIENT HARM WAS REPORTED. DEVICE IS AVIALALBE FOR RETURN. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21926148 ↗MalfunctionFeb 19, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED TO THAT THE SURGEON HAD AN ISSUE WITH SUTURE. THE SCRUB TECHS WENT TO OPEN THE SUTURE AND THE NEEDLE WAS BROKEN. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21414035 ↗MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355472 ↗MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355447 ↗MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355427 ↗MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6)2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355398 ↗MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355355 ↗MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355340 ↗MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355326 ↗MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355307 ↗MalfunctionMay 20, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355303 ↗MalfunctionApr 9, 2024Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19071631 ↗MalfunctionApr 9, 2024Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BENT OVER AND BROKE OFF THE SUTURE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19071627 ↗MalfunctionMar 5, 2024Break; Break
IT WAS REPORTED THAT PATIENT UNDERWENT A C-SECTION ON AN UNKNOWN DATE IN 2024, AND SUTURE WAS USED. WHILE COMPLETING THE C-SECTION, THE SUTURE POPPED. THE SURGEON GRABBED ANOTHER SUTURE TO COMPLETE THE CASE WITHOUT PATIENT HARM. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18841419 ↗MalfunctionSep 15, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, TWO OF THE SUTURES BROKE WHILE USING THEM DURING A C-SECTION. NO PATIENT HARM. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17756785 ↗MalfunctionSep 15, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, TWO OF THE SUTURES BROKE WHILE USING THEM DURING A C-SECTION. NO PATIENT HARM. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17756784 ↗MalfunctionJan 9, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE FASCIAL CLOSURE THE NEEDLE BROKE. THE FRAGMENT WAS RETRIEVED, AND A NEW SUTURE WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16126446 ↗MalfunctionJul 1, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A C-SECTION ON (B)(6) 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE. A LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14888196 ↗MalfunctionApr 7, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IN 2 INSTANCES THE NEEDLE BROKE WHILE BEING USE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14043634 ↗MalfunctionApr 7, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, IN 2 INSTANCES THE NEEDLE BROKE WHILE BEING "USE". THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14043631 ↗MalfunctionApr 7, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THAT THEY HAD TWO INSTANCES THAT THE NEEDLE BROKE WHILE BEING USE IN SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14043128 ↗MalfunctionApr 7, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THAT THEY HAD TWO INSTANCES THAT THE NEEDLE BROKE WHILE BEING USE IN SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14043123 ↗MalfunctionApr 7, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN 2022 AND SUTURE WAS USED. DURING THE PROCEDURE, THAT THEY HAD TWO INSTANCES THAT THE NEEDLE BROKE WHILE BEING USE IN SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14043115 ↗
Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE617 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE485 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE446 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE443 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone406 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE339 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death229 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗