Skip to content

PDS ZB740Recalls & reported problems

Ethicon, LLC · Violet Monofilament (Polydioxanone), Absorbable Synthetic Surgical Suture · Class 2 · Implant

Adverse event reports
18
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 10705031062108), received between May 8, 2018 and Jan 31, 2025 — classified as 4 injuries, 14 malfunctions.

The most frequently reported specific device problem is “Break” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break9
  • Material Separation6
  • Device Contaminated During Manufacture or Shipping2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient6
  • No Code Available4
  • Wound Dehiscence4
  • Impaired Healing1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 31, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. THE NEEDLE AND SUTURE HAD COME OFF DURING USE IN FASCIA SUTURING OF THE WOUND CLOSURE. IT WAS A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21287573 ↗
  • MalfunctionAug 14, 2024Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. WHEN OPENED THE PACKAGE, THERE WAS A BROWN TRASH IN IT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19987311 ↗
  • MalfunctionMar 26, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A COLECTOMY PROCEDURE ON (B)(6) 2024, AND SUTURE WAS USED. DURING INTERRUPTED SUTURING, THE SUTURE WAS DETACHED FROM THE NEEDLE ALTHOUGH NO EXTRA FORCE WAS APPLIED ON THE PART OF CR. IT OCCURRED IN ONE OF THE PRODUCT IN EACH TWO PACKAGES. IT WAS A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18977952 ↗
  • MalfunctionMar 26, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A COLECTOMY PROCEDURE ON (B)(6) 2024, AND SUTURE WAS USED. DURING INTERRUPTED SUTURING, THE SUTURE WAS DETACHED FROM THE NEEDLE ALTHOUGH NO EXTRA FORCE WAS APPLIED ON THE PART OF CR. IT OCCURRED IN ONE OF THE PRODUCT IN EACH TWO PACKAGES. IT WAS A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18977948 ↗
  • MalfunctionAug 11, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS FOUND THAT THE SUTURE WAS DETACHED FROM THE NEEDLE. IT WAS A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17511780 ↗
  • MalfunctionMay 26, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. AFTER OPENING THE PACKAGE AND BEFORE USE, IT WAS CONFIRMED THAT A FOREIGN SUBSTANCE OF BIOLOGICAL ORIGIN, SUCH AS HUMAN HAIR OR INSECT, WAS FOUND IN THE STERILE PACKAGE. THE OPERATION TIME WAS EXTENDED WITHIN 30 MINUTES. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17009629 ↗
  • MalfunctionJul 28, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN TAKING OUT FROM THE PACKAGE, IT WAS FOUND THAT THE NEEDLE TIP HAD BEEN MISSING. THE NEEDLE TIP WAS SEARCHED, BUT IT COULD NOT BE FOUND, SO IT WAS COMMENTED THAT IT MAY HAVE BEEN IN THIS STATE FROM THE BEGINNING. THE NEEDLE TIP DID NOT FALL INTO THE PATIENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12240083 ↗
  • MalfunctionJan 6, 2021Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2020 AND THE SUTURE WAS USED. DURING SURGERY, AFTER OPENING THE PACKAGE, THE NEEDLE WAS BROKEN IMMEDIATELY WHEN HOLDING THE NEEDLE. THE SUTURE BROKE BEFORE USE ON THE PATIENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11128214 ↗
  • MalfunctionJul 1, 2020Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2020 AND SUTURE WAS USED. DURING WOUND CLOSURE BY INTERRUPTED SUTURING, AFTER PASSING THE NEEDLE THROUGH THE ABDOMINAL FASCIA, WHEN TRYING TO TIE A KNOT, THE SUTURE WAS BROKEN EASILY. ANOTHER PACKAGE WAS USED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10217489 ↗
  • MalfunctionJul 1, 2020Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2020 AND SUTURE WAS USED. DURING WOUND CLOSURE BY INTERRUPTED SUTURING, AFTER PASSING THE NEEDLE THROUGH THE ABDOMINAL FASCIA, WHEN TRYING TO TIE A KNOT, THE SUTURE WAS BROKEN EASILY. ANOTHER PACKAGE WAS USED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10217484 ↗
  • MalfunctionJul 1, 2020Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2020 AND SUTURE WAS USED. DURING WOUND CLOSURE BY INTERRUPTED SUTURING, AFTER PASSING THE NEEDLE THROUGH THE ABDOMINAL FASCIA, WHEN TRYING TO TIE A KNOT, THE SUTURE WAS BROKEN EASILY. ANOTHER PACKAGE WAS USED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10217483 ↗
  • MalfunctionMay 14, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2020 AND THE SUTURE WAS USED. IT WAS REPORTED THAT WHEN TRYING TO USE THE PRODUCTS, IT WAS FOUND THAT SOME SUTURES HAD BEEN DETACHED FROM THE NEEDLES. THEY WERE CONTROL RELEASE NEEDLES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10060031 ↗
  • InjuryFeb 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN OB-GYN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE LONGITUDINAL INCISION WAS MADE NOT EXACTLY ON THE MIDLINE (LINEA ALBA) BUT WAS SLIGHTLY APART FROM THE MIDLINE. THE BOTH ENDS OF THE LONGITUDINAL INCISION WERE SUTURED BY INTERRUPTED SUTURING, AND THE MIDDLE OF THE INCISION WAS SUTURED BY CONTINUOUS SUTURING. ABOUT ONE WEEK AFTER THE PROCEDURE, WOUND DEHISCENCE OCCURRED IN THE WARD ON THE FASCIA. IT WAS REPORTED THAT CONTRIBUTING FACTORS WERE THE PATIENT WAS OBESE WITH BMI 30. IT WAS ALSO REPORTED THAT CONTRIBUTING FACTORS WERE TENSION WAS APPLIED TO A PART OF THE SUTURE AND DELAYED HEALING AT THE INCISION SITE. NO SUTURE BREAKAGE OCCURRED AND THE FASCIA WAS BROKEN AND TORN. DURING THE REOPERATION, IT WAS CONFIRMED THAT THE SUTURE REMAINED ON ONE SIDE OF THE INCISION. THE WOUND OPENING WAS NOT EXACTLY ON THE MIDLINE (LINEA ALBA) BUT WAS SLIGHTLY APART FROM THE MIDLINE. THE WOUND ON THE PERITONEUM AND THE WOUND ON THE FASCIA WERE SUTURED INDIVIDUALLY BY INTERRUPTED SUTURING. FOR REINFORCEMENT OF THE CLOSURE THE MIDDLE PART, EXCEPT FOR BOTH ENDS OF THE WOUND ON THE FASCIA, WAS SUTURED BY CONTINUOUS SU

    Patient problems: Impaired Healing; No Code Available; Wound Dehiscence

    MAUDE report 9752985 ↗
  • MalfunctionSep 6, 2019Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2019 AND SUTURE WAS USED. WHEN THE SURGEON WAS HOLDING THE NEEDLE BY A NEEDLE-HOLDER, THE NEEDLE WAS BROKEN AT THE BODY PART. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8972656 ↗
  • MalfunctionFeb 22, 2019Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT LAPAROSCOPY-ASSISTED DISTAL GASTRECTOMY ON (B)(6) 2019 AND SUTURE WAS USED. WHEN CLOSING WOUND OF SMALL INCISION SITE, ONE OF THE NEEDLES WAS DETACHED FROM THE SUTURE EASILY DURING PASSING THE NEEDLE THROUGH THE FASCIA BY INTERRUPTED SUTURING. IT WAS A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8363844 ↗
  • InjuryMay 8, 2018Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT OPEN TOTAL GASTRECTOMY PROCEDURE ON (B)(6) 2018 AND SUTURE WAS USED TO CLOSE THE WOUND ON FASCIA WITH INTERRUPTED TECHNIQUE. POST OPERATIVELY, THE PATIENT COUGHED IN THE WARD. A SECOND PROCEDURE WAS PERFORMED ON (B)(6) 2018 AND SUTURE WAS FOUND BROKEN, THE TISSUE WAS NOT CUT WHEN WOUND DEHISCED. THE SURGEON OPINED THAT STRONG FORCE WAS APPLIED TO THE SUTURE DURING BUCKING SINCE THE PATIENT WAS OBESE. THE PATIENT IS REPORTED AS STABLE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Code Available; Wound Dehiscence

    MAUDE report 7495956 ↗
  • InjuryMay 8, 2018Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT OPEN TOTAL GASTRECTOMY PROCEDURE ON (B)(6) 2018 AND SUTURE WAS USED TO CLOSE THE WOUND ON FASCIA WITH INTERRUPTED TECHNIQUE. POST OPERATIVELY, THE PATIENT COUGHED IN THE WARD. A SECOND PROCEDURE WAS PERFORMED ON (B)(6) 2018 AND SUTURE WAS FOUND BROKEN, THE TISSUE WAS NOT CUT WHEN WOUND DEHISCED. THE SURGEON OPINED THAT STRONG FORCE WAS APPLIED TO THE SUTURE DURING BUCKING SINCE THE PATIENT WAS OBESE. THE PATIENT IS REPORTED AS STABLE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Code Available; Wound Dehiscence

    MAUDE report 7495910 ↗
  • InjuryMay 8, 2018Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT OPEN TOTAL GASTRECTOMY PROCEDURE ON 4/16/2018 AND SUTURE WAS USED TO CLOSE THE WOUND ON FASCIA WITH INTERRUPTED TECHNIQUE. POST OPERATIVELY, THE PATIENT COUGHED IN THE WARD. A SECOND PROCEDURE WAS PERFORMED ON (B)(6) 2018 AND SUTURE WAS FOUND BROKEN, THE TISSUE WAS NOT CUT WHEN WOUND DEHISCED. THE SURGEON OPINED THAT STRONG FORCE WAS APPLIED TO THE SUTURE DURING BUCKING SINCE THE PATIENT WAS OBESE. THE PATIENT IS REPORTED AS STABLE. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Code Available; Wound Dehiscence

    MAUDE report 7495845 ↗

Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices

All SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices →