PDS PDP509GRecalls & reported problems
Ethicon, LLC · Antibacterial Clear Monofilament (Polydioxanone), Absorbable Surgical Suture · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10705031207240), received between Jun 7, 2019 and May 4, 2026 — classified as 2 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break11
- Material Separation3
- Adverse Event Without Identified Device or Use Problem1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Consequences Or Impact To Patient4
- Device Embedded In Tissue or Plaque1
- Foreign Body Reaction1
- Impaired Healing1
- Insufficient Information1
- Pocket Erosion1
- Wound Dehiscence1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 4, 2026Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE NEEDLE WAS BROKEN DURING USE AND USE WAS STOPPED. NO PIECES WERE LEFT INSIDE THE PATIENT. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25073286 ↗MalfunctionDec 16, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE NEEDLE COULD NOT BE PASSED THROUGH FROM THE FIRST STITCH. IT COULD BE CONSIDERED THAT THE NEEDLE TIP WAS BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23821096 ↗MalfunctionNov 13, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE WAS BROKEN DURING USE. IT IS UNKNOWN WHAT SURGERY PERFORMED. AFTER RECEIVING THE PRODUCT, WHEN THE SALES REP CHECKED IT, HE SPECULATES THAT THE SUTURE WAS DETACHED RATHER THAN THE NEEDLE BREAKAGE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23542755 ↗InjuryFeb 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED FOR FACIAL SUTURING. THE SUTURE WAS USED FOR BURIED SUTURE, AND THE USED SITE WAS REINFORCED WITH TAPING. POST-OP, AFTER APPROXIMATELY 3 MONTHS, THE SUTURE WAS PARTLY PROTRUDED AS IF A REACTION OF FOREIGN MATTER, SO WHEN THE SUTURE WAS REMOVED WITH THE PATIENT'S AGREEMENT, THE WOUND WAS OPENED. THE OPEN WOUND WAS SUTURED AND TREATED WITH ANOTHER NYLON SUTURE, BUT IT WAS A SERIOUS EXPERIENCE FOR BOTH THE PATIENT AND THE DOCTOR. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THE PATIENT HAS BEEN DISCHARGED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Foreign Body Reaction; Impaired Healing; Pocket Erosion; Wound Dehiscence
MAUDE report 21358643 ↗InjuryFeb 6, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC CHOLECYSTECTOMY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN SUTURING THE DERMIS ON WOUND CLOSURE, THE NEEDLE BECAME DETACHED FROM THE SUTURE ON THE WAY AND THE NEEDLE WAS LOST IN SIGHT. THE SURGEON THOUGHT THAT THE NEEDLE FELL OUT OF THE SURGICAL FIELD INSTEAD OF INSIDE THE BODY, SO THE SURGEON PROCEEDED TO SUTURE THE WOUND WITH A NEW NEEDLE, WHILE LOOKING OUT FOR THE FALLEN NEEDLE. AFTER COMPLETING THE SUTURE, THERE WAS ALSO NO NEEDLE IN SIGHT, SO AN X-RAY WAS PERFORMED. A SHADOW THAT LOOKED AS IF THE NEEDLE WAS INSIDE THE BODY WAS SEEN, AND ADDITIONAL FLUOROSCOPY WAS PERFORMED. AFTER FLUOROSCOPY, THE NEEDLE WAS CONFIRMED TO BE INSIDE THE BODY, SO WHEN THE WOUND WAS OPENED AGAIN, THE NEEDLE WAS FOUND TO HAVE FALLEN INTO THE SUBCUTANEOUS AREA AND IT WAS REMOVED. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THE EFFECT ON THE PATIENT WAS THE RISK OF A SMALL AMOUNT OF EXPOSURE DUE TO ADDITIONAL FLUOROSCOPY, AND AN INCREASE OF THE ENTIRE BODY LOAD DUE TO THE EXTENSION OF THE OPERATION TIME BY ABOUT 1 HOUR AND HALF. ADDITIONAL INFORMATION AS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 21325217 ↗MalfunctionApr 3, 2024Break; Break
IT WAS REPORTED THAT PATIENT UNDERWENT A LAPAROSCOPIC CHOLECYSTECTOMY ON (B)(6) 2024, AND SUTURE WAS USED ON THE DERMIS. DURING SURGERY, WHEN PULLING THE SUTURE DURING LIGATION, IT WAS BROKEN. IT WAS USED WITH THE USUAL WAY. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19034531 ↗MalfunctionJun 22, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17181243 ↗MalfunctionMay 9, 2023Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED FROM THE ANALYSIS OF THE RETURNED PRODUCT THAT NEEDLE BREAKAGE OCCURRED, A FRACTURE WAS OBSERVED NEAR THE TIP AND THROUGH THE BODY OF THE NEEDLE. IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. IT WAS FOUND THAT THE NEEDLE TIP HAD BEEN BENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16898444 ↗MalfunctionJan 11, 2023Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE NEEDLE WAS BROKEN DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16135656 ↗MalfunctionNov 9, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. WHEN THE SURGEON TRIED TO USE THE PRODUCT, IT WAS FOUND THAT THE NEEDLE WAS ALREADY DETACHED FROM THE SUTURE FROM THE BEGINNING. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. A FRACTURE WAS OBSERVED AT THE SUTURE ATTACHMENT OF THE NEEDLE DURING DEVICE ANALYSIS. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15759912 ↗MalfunctionAug 17, 2022Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND THE SUTURE WAS USED. DURING SURGERY, THE SUTURE WAS DETACHED FROM THE NEEDLE EASILY DURING CONTINUOUS SUTURING. IT WAS NOT A CONTROL RELEASE NEEDLE. THE PRODUCT WAS DISCARDED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15244396 ↗MalfunctionApr 25, 2022Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ORTHOPEDIC PROCEDURE ON (B)(6) 2022 AND SUTURE WAS USED. WHEN PUTTING THE FIRST STITCH WITH THE USUAL WAY, THE NEEDLE TIP WAS BROKEN. THE BROKEN FRAGMENT WAS CONFIRMED VISUALLY AND IT WAS RETRIEVED. THERE WAS NO POSSIBILITY OF INTERNAL REMNANT. THE OPERATION TIME WAS NOT EXTENDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14193953 ↗MalfunctionFeb 13, 2020Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN OB-GYN PROCEDURE ON AN UNKNOWN DATE AND THE SUTURE WAS USED. IT WAS REPORTED THAT DURING SURGERY, THE SURGEON HELD THE NEEDLE WITH A NEEDLE-HOLDER, THE NEEDLE WAS BROKEN EASILY AT THE SWAGE PART THOUGH NO EXTRA FORCE WAS APPLIED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9708982 ↗MalfunctionOct 31, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A BREAST CONSERVATION SURGERY ON (B)(6) 2019 AND SUTURE WAS USED. THE SURGEON FELT THAT THE NEEDLE BECAME DULL WHILE HE WAS DOING INTERRUPTED SUTURING BY INSTRUMENT TYING. WHEN HE CHECKED THE NEEDLE TIP, IT WAS FOUND THAT THE NEEDLE TIP HAD BEEN BROKEN. THE BROKEN NEEDLE TIP COULD NOT BE FOUND ON THE SURGICAL FIELD NOR THE INSTRUMENT TABLE. IT COULD NOT BE FOUND ON X-RAY IMAGES, NEITHER. SO, THE SURGERY WAS COMPLETED WITHOUT FINDING THE BROKEN NEEDLE TIP. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PATIENT HAS BEEN DISCHARGED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Device Embedded In Tissue or Plaque; No Consequences Or Impact To Patient
MAUDE report 9261150 ↗MalfunctionSep 19, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN ORTHOPEDIC SURGERY ON UNKNOWN DATE AND THE SUTURE WAS USED. DURING THE SURGERY, THE NEEDLE TIP WAS BROKEN DURING USE THOUGH IT WAS USED IN USUAL WAY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9086167 ↗MalfunctionJun 7, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A CARDIOVASCULAR PROCEDURE ON AN UNKNOWN DATE IN 2019 AND SUTURE WAS USED. THE SUTURE BROKE DURING USE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8680692 ↗
Other Suture, surgical, absorbable, polydioxanone devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE617 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE485 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE446 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE443 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone406 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE339 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death229 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page