PDS Z452GRecalls & reported problems
Ethicon, LLC · Violet Monofilament (Polydioxanone), Absorbable Synthetic Surgical Suture · Class 2 · Implant
All models of PDS (224)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10705031218703), received between Nov 15, 2018 and Jan 14, 2025 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Material Separation7
- Component Misassembled1
- Device Markings/Labelling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information5
- No Consequences Or Impact To Patient4
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 14, 2025Material Separation; Material Separation
IT WAS REPORTED AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THEY RECENTLY PURCHASED SUTURE. THE PRODUCT SEEMS TO BE FAULTY AS WHEN THE DOCTOR WAS PERFORMING SURGERY THE NEEDLE COMES OFF OF THE SUTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21153911 ↗MalfunctionJan 14, 2025Material Separation; Material Separation
IT WAS REPORTED AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THEY RECENTLY PURCHASED SUTURE. THE PRODUCT SEEMS TO BE FAULTY AS WHEN THE DOCTOR WAS PERFORMING SURGERY THE NEEDLE COMES OFF OF THE SUTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21153892 ↗MalfunctionJan 14, 2025Material Separation; Material Separation
IT WAS REPORTED AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THEY RECENTLY PURCHASED SUTURE. THE PRODUCT SEEMS TO BE FAULTY AS WHEN THE DOCTOR WAS PERFORMING SURGERY THE NEEDLE COMES OFF OF THE SUTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21153885 ↗MalfunctionMar 12, 2024Material Separation; Material Separation
IT WAS REPORTED AN ANIMAL UNDERWENT AN UNKNOWN VETERINARY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR BY THE CUSTOMER THAT 3 NEEDLES COME LOOSE FROM SUTURE DURING SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18885399 ↗MalfunctionMar 12, 2024Material Separation; Material Separation
IT WAS REPORTED AN ANIMAL UNDERWENT AN UNKNOWN VETERINARY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR BY THE CUSTOMER THAT 3 NEEDLES COME LOOSE FROM SUTURE DURING SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18885398 ↗MalfunctionMar 12, 2024Material Separation; Material Separation
IT WAS REPORTED AN ANIMAL UNDERWENT AN UNKNOWN VETERINARY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED TO THE DISTRIBUTOR BY THE CUSTOMER THAT 3 NEEDLES COME LOOSE FROM SUTURE DURING SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18885394 ↗MalfunctionMar 22, 2023Component Misassembled; Component Misassembled
IT WAS REPORTED THAT A BOX OF 12 SUTURES HAS 10 CORRECT PACKS OF SUTURE AND 2 INCORRECT PACKS INSIDE OF THE BOX. (ORDERED Z452G BOX 12. BOX HAS 10 PACKS OF Z452G AND 2 PACKS Z824G). THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16593119 ↗MalfunctionSep 13, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A FELINE UNDERWENT A SPAY PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE DETACHED FROM THE SUTURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12465524 ↗MalfunctionJul 1, 2021Break; Break
IT WAS REPORTED THAT AN ANIMAL UNDERWENT A CANINE OVARIOHYSTERECTOMY PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. POST-OPERATIVELY, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 12105130 ↗MalfunctionJul 1, 2021Break; Break
IT WAS REPORTED THAT AN ANIMAL UNDERWENT A CANINE OVARIOHYSTERECTOMY PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. POST-OPERATIVELY, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 12105129 ↗MalfunctionJul 1, 2021Break; Break
IT WAS REPORTED THAT AN ANIMAL UNDERWENT A CANINE OVARIOHYSTERECTOMY PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. POST-OPERATIVELY, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 12105128 ↗MalfunctionJul 1, 2021Break; Break
IT WAS REPORTED THAT AN ANIMAL UNDERWENT A CANINE OVARIOHYSTERECTOMY PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. POST-OPERATIVELY, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 12105127 ↗MalfunctionJul 1, 2021Break; Break
IT WAS REPORTED THAT AN ANIMAL UNDERWENT A CANINE OVARIOHYSTERECTOMY PROCEDURE ON (B)(6) 2021 AND SUTURE WAS USED. POST-OPERATIVELY, THE SUTURE BROKE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 12105126 ↗MalfunctionNov 10, 2020Break; Break
IT WAS REPORTED THAT AN ANIMAL UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2020; AND A SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE BROKE WHILE TRYING TO TIE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10816784 ↗MalfunctionMar 25, 2020Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2020 AND THE SUTURE WAS USED. DURING SURGERY, THE SUTURE KEPT BREAKING. UPON INSPECTING THE SUTURE LATER ON, SMALL WEAK SPOTS WERE NOTICED ALONG THE WHOLE LENGTH OF THE SUTURE. A LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9876354 ↗MalfunctionNov 28, 2018Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED A PATIENT UNDERWENT AND UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. SUTURE PACKAGES PRODUCT CODE INSIDE OF BOX DO NOT MATCH PRODUCT CODE ON BOX. SUTURE PACKET LABELED AS Z332H (PDS II SUTURE) AND OUTSIDE OF BOX IS PRODUCT CODE Z452G (PDS II SUTURE). CUSTOMER DID OPEN ONE PACKAGE BEFORE NOTICING THE ERROR. THERE WAS NO PATIENT CONSEQUENCE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8112194 ↗MalfunctionNov 15, 2018Break; Break
IT WAS REPORTED THAT AN ANIMAL UNDERWENT A CAT SPAYING PROCEDURE ON (B)(6) 2018 AND A SUTURE WAS USED. AFTER SURGERY, THE ANIMAL WAS X-RAYED AND FOUND THAT THE SUTURE HAD BROKEN IN THE MIDDLE. THE KNOTS WERE INTACT BUT THE MIDDLE OF THE SUTURE LINE WAS BROKEN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8073936 ↗
Other SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE devices
- STRATAFIX SXPP1A405Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE617 reports
- STRATAFIX SXPP1A400Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE485 reports
- STRATAFIX SXPP1A401Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE446 reports
- STRATAFIX SXPP1A301Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE443 reports
- STRATAFIX SXPP2B405Ethicon, LLC · Suture, surgical, absorbable, polydioxanone406 reports
- STRATAFIX SXPP1A406Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE339 reports
- STRATAFIX SXPP1A404Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death319 reports
- STRATAFIX SXPP1B420Ethicon, LLC · SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE1 death229 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page