PDS PDP771DRecalls & reported problems

Ethicon, LLC · Antibacterial Violet Monofilament (Polydioxanone), Absorbable Surgical Suture · Class 2 · Implant

All models of PDS (224)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon PDS absorbable surgical suture is a violet thread surgeons use to stitch tissue closed. Most of the 28 reports describe the needle or suture breaking mid-operation, sometimes several needles in a row, forcing surgeons to swap in another suture and lengthen the surgery. Three reports mention patient harm such as a leaking stitch, fistula or fluid buildup. No deaths and no recalls name this device, and reports rose from 3 in the previous 12 months to 15 in the most recent.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 10705031218772), received between Sep 12, 2022 and Sep 24, 2026 — classified as 3 injuries, 27 malfunctions.

The most frequently reported specific device problem is “Break” (20 reports).

9 other recalls cover devices cleared under the same FDA submission (K061037, K212380); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 12 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break20
  • Material Separation4
  • Adverse Event Without Identified Device or Use Problem3
  • Delivered as Unsterile Product2
  • Component Misassembled1
  • Dull, Blunt1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27
  • Failure to Anastomose3
  • Fistula3
  • Pleural Effusion3

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Delivered as Unsterile Product

    It was reported a patient underwent an unknown surgery on (b)(6) 2026 and suture was used. It was found that the needle had been protruded from the package before opening the package. Another product was used to complete the case. There were no adverse consequences to the patient or user. Further details are not provided. Additional information has been requested.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50887854 ↗
  • MalfunctionSep 24, 2026Break; Material Separation

    It was reported that a patient underwent a Laparoscopic partial hepatectomy and cholecystectomy under general anesthesia procedure on (b)(6) 2026, and suture was used. At 15:12, the surgery was performed. The doctor finished suturing the tissues in the abdominal cavity and used a needle holder to clamp the tail end suture. When removing the suture, the nurse and doctor found that there was only suture on the needle holder, no needle, and the tail end of the suture was broken. The doctor used laparoscopy to examine the abdominal cavity but was unsuccessful. The radiology technician used a C-arm to locate the needle in the abdominal cavity, and the surgeon used laparoscopy to completely remove the needle. Additional information was requested.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882340 ↗
  • MalfunctionJan 6, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. THE INSTRUMENT WAS USED FOR SURGICAL SUTURING AND WAS OPERATED ACCORDING TO THE INSTRUCTIONS. DURING USE, THE NEEDLE STEM SUDDENLY BROKE, CAUSING THE SURGERY TO COME TO A HALT. IMMEDIATELY, THE REMAINING NEEDLE AT THE BROKEN PART WAS FOUND, AND THE DOCTOR REPLACED THE SUTURE TO COMPLETE THE SURGERY. THE HANDLING OF THIS MALFUNCTION PROLONGS THE SURGICAL TIME, INCREASES THE RISK OF ANESTHESIA EXPOSURE FOR PATIENTS, AND MAY LEAD TO THE RISK OF INTRAOPERATIVE COMPLICATIONS, RESULTING IN SERIOUS CONSEQUENCES. RESULT: THE TREATMENT WAS TIMELY AND HAD NO IMPACT ON THE PATIENT EXCEPT FOR PROLONGING THE SURGICAL TIME. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23979551 ↗
  • MalfunctionDec 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING LOW ANTERIOR RESECTION FROM THE MORNING, WHEN THE NEEDLE WAS APPLIED TO STOMA, THE SWAGE PART WAS BROKEN 2 NEEDLES IN A ROW. THE APPLIED PART WAS STOMA, AND THE NEEDLE WAS HELD WITH THE NEEDLE-HOLDER, SO NO PIECES FELL INTO THE PATIENT. IT WAS SAID BY THE DOCTOR THAT THE NEEDLE SWAGE PART BROKE 2 NEEDLES IN A ROW, SO THIS BATCH WAS AVOIDED USING. IT WAS A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707699 ↗
  • MalfunctionDec 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING LOW ANTERIOR RESECTION FROM THE MORNING, WHEN THE NEEDLE WAS APPLIED TO STOMA, THE SWAGE PART WAS BROKEN 2 NEEDLES IN A ROW. THE APPLIED PART WAS STOMA, AND THE NEEDLE WAS HELD WITH THE NEEDLE-HOLDER, SO NO PIECES FELL INTO THE PATIENT. IT WAS SAID BY THE DOCTOR THAT THE NEEDLE SWAGE PART BROKE 2 NEEDLES IN A ROW, SO THIS BATCH WAS AVOIDED USING. IT WAS A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707677 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23647858 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23647845 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23647842 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23647840 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23647835 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23647833 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645671 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645670 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645669 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645663 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645660 ↗
  • MalfunctionNov 25, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23645479 ↗
  • MalfunctionApr 29, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST LUMP EXCISION SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED A MISSING NEEDLE TIP. IT WAS REPORTED THAT THE NEEDLE TIP WAS FOUND MISSING DURING OPERATION. THE DOCTOR TRIED TO FIND THE NEEDLE TIP IN PATIENT BUT FAILED. IT'S UNCERTAIN WHEN NEEDLE TIP WAS MISSING, PRE-OP OR INTRA-OP. THE PATIENT'S CONDITION IS STILL UNDER OBSERVATION. IT'S UNKNOWN HOW LONG THE SURGERY WAS PROLONGED. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21927372 ↗
  • MalfunctionDec 11, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE "ON THE 15TH" AROUND 12:00 PM AND SUTURE WAS USED. WHEN THE DOCTOR VISITED THE OPERATING ROOM, IT WAS FOUND THAT THE NEEDLE WAS BROKEN AT THE MIDDLE PART DURING INTERRUPTING SUTURE. THE NURSE SAID THAT AFTER APPLYING ONE STITCH, THE NEEDLE WAS BENT, SO THE NURSE STRETCHED OUT OF THE BEND AND WHEN IT WAS GRASPED AGAIN, THE NEEDLE WAS BROKEN. FORTUNATELY, THE BROKEN NEEDLE WAS FOUND IMMEDIATELY. IT WAS A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20906066 ↗
  • MalfunctionDec 4, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN OSTOMY SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. POST-OP, IT WAS REPORTED BY A SALES REP THAT, AFTER AN OSTOMY, THE PDS USED WAS NOT ABSORPTION EVEN MORE THAN 6 MONTHS, PROBABLY BECAUSE IT WAS USED OUTSIDE THE BODY, SO THE SUTURE WAS REMOVED. IT WAS A CONTROL RELEASE NEEDLE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO SAMPLE WILL BE RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20845016 ↗
  • MalfunctionJul 16, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ESOPHAGECTOMY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE PRIMARY PACKAGE OF THE SUTURE WAS FOUND DAMAGED WITH HOLE PRIOR TO USE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19751439 ↗
  • MalfunctionJun 4, 2024Material Separation; Material Separation

    IT WAS REPORTED FROM THE ANALYSIS OF THE RETURNED PRODUCT THAT NEEDLE BREAKAGE WAS OBSERVED. IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN SURGERY, AFTER OPENING THE PACKAGE AND BEFORE USE, WHEN THE SURGEON HELD THE PRODUCT WITH A NEEDLE HOLDER, THE SUTURE BROKE AT SWAGE PART. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19458705 ↗
  • MalfunctionJun 4, 2024Break; Dull, Blunt; Break; Dull, Blunt

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN PROCEDURE AND SUTURE WAS USED. DURING SURGERY, WHEN USING, THE TIP OF THE PRODUCT WAS CHECKED BECAUSE THE DOCTOR FELT THAT THERE WAS BAD PENETRATION, SO THE TIP WAS BROKEN. THE DOCTOR CONSIDERED THAT IT WAS PROBABLY BROKEN FROM THE BEGINNING. IT WAS A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19458067 ↗
  • MalfunctionApr 23, 2024Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN SURGERY ON 15-APRIL-2024 AND SUTURE WAS USED. DURING THE SURGERY, WHEN OPENING THE PACKAGE, THE TIP WAS MISSING. IT WAS A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19163463 ↗
  • MalfunctionJun 28, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE WAS BROKEN AT THE SWAGE PART DURING USE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17217184 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K061037, K212380. One submission often covers many models, so check the product and lot details in each record.

  • Z-2147-2024 · May 10, 2024 · Open, Classified
    PDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in oph
  • Z-1571-2023 · Mar 31, 2023 · Open, Classified
    PDS Plus Antibacterial (polydioxanone) Suture - PDS PLUS CLR 27IN(70CM) USP3-0(M2) S/A PC-25 PRIME. Intended for use in general soft tissue approximation, including use in pediatri
  • Z-1135-2019 · Mar 12, 2019 · Terminated
    VICRYL PLUS Sutures: VICRYL PLUS SUTURE 18"(45CM) 3-0 VIO VICRYL PLUS SUTURE 18"(45CM) 0 UND VICRYL PLUS SUTURE 18"(45CM) 0 VIO VICRYL PLUS SUTURE 18"(45CM) 0 VIO) VICRYL PLUS SU
  • Z-2533-2018 · Jun 8, 2018 · Terminated
    PDS Plus Antibacterial Suture-PDS PLUS Undyed Suture 45cm (18") 5-0 Single Armed P-3 Prime Needle Product Code: PDP493G (GTIN 10705031203945)
  • Z-2532-2018 · Jun 8, 2018 · Terminated
    PDS Plus Antibacterial (Polydioxanone) Suture-PDS PLUS Undyed Suture 45cm (18") 4-0 Single Armed PS-4 Prime Needle Product Code: PDP507G (GTIN 10705031203983)
  • Z-1338-2018 · Jan 10, 2018 · Terminated
    PDS Plus Antibacterial (Polydioxanone) Suture Size 0 PDS Plus Violet 36 Single Armed CT-1 Needle
  • Z-1340-2018 · Jan 10, 2018 · Terminated
    PDS Plus Antibacterial (Polydioxanone) Suture Size 0 PDS Plus Violet 36 Single Armed CTB-1 Needle
  • Z-1339-2018 · Jan 10, 2018 · Terminated
    PDS Plus Antibacterial (Polydioxanone) Suture Size 1 PDS PLUS Violet 27 Single Armed OS-4 Needle
  • Z-1314-2009 · Feb 2, 2009 · Terminated
    PDS Plus Antibacterial (polydioxanone) Suture Sutures are intended for general soft tissue approximation and/or ligation.

Other Suture, surgical, absorbable, polydioxanone devices

All Suture, surgical, absorbable, polydioxanone devices →