PDS Z464HRecalls & reported problems

Ethicon, LLC · Violet Monofilament (Polydioxanone), Synthetic Absorbable Surgical Suture · Class 2 · Implant

Quick summary

This Ethicon PDS absorbable surgical suture has 70 adverse event reports, all describing malfunctions with no deaths and no injuries. Reports mostly describe the suture or needle breaking during surgery, including needle fragments falling into the surgical field, with surgeons typically switching to another suture to finish. One report mentions a foreign body left in a patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
70
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 10705031462519), received between Oct 29, 2018 and Mar 13, 2026 — classified as 70 malfunctions.

The most frequently reported specific device problem is “Break” (48 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break48
  • Material Separation17
  • Component Misassembled4
  • Dull, Blunt1

Patient problems

  • No Clinical Signs, Symptoms or Conditions52
  • No Consequences Or Impact To Patient18
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WERE USED. DURING SURGERY, THE SUTURE WAS BROKEN DURING USE. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24590781 ↗
  • MalfunctionFeb 24, 2026Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2026 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, THERE WAS ONE PACKAGE WHICH HAD A PRODUCT THE SUTURE AND THE NEEDLE HAD BEEN DETACHED. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24432424 ↗
  • MalfunctionJan 15, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SWAGE PART OF TWO NEEDLES WERE BROKEN. NO NEEDLE PIECES FELL INTO THE SURGICAL FIELD. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24080244 ↗
  • MalfunctionJan 15, 2026Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE SWAGE PART OF TWO NEEDLES WERE BROKEN. NO NEEDLE PIECES FELL INTO THE SURGICAL FIELD. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24080229 ↗
  • MalfunctionOct 15, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE SUTURE WAS BROKEN WHEN TYING. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WAS NO EFFECT ON THE PATIENT. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23300471 ↗
  • MalfunctionOct 1, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE WAS BROKEN AT THE SWAGE PART DURING SUTURING. IT APPEARS TO HAVE BEEN A SCARRED AREA. WHEN THE NEEDLE WAS BROKEN, THE NEEDLE FRAGMENTS FELL INTO THE SURGICAL FIELD, BUT WERE QUICKLY RETRIEVED WITHOUT BEING LOST IN SIGHT. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23191252 ↗
  • MalfunctionAug 20, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, Z464H WAS BROKEN SEVERAL TIMES DURING SURGERY. THE BROKEN PART WAS IN THE MIDDLE OF THE SUTURE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED DUE TO THE PATIENT WITH AN INFECTIOUS DISEASE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22852341 ↗
  • MalfunctionAug 20, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, Z464H WAS BROKEN SEVERAL TIMES DURING SURGERY. THE BROKEN PART WAS IN THE MIDDLE OF THE SUTURE. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO SAMPLE WILL BE RETURNED DUE TO THE PATIENT WITH AN INFECTIOUS DISEASE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22852340 ↗
  • MalfunctionJul 30, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING AN UNKNOWN SEMI-OPEN SURGERY, WHEN THE NEEDLE PIERCED THE TISSUE DURING DERMAL SUTURE AND IT WAS CHECKING WHERE IT HAD EMERGED, IT WAS CONFIRMED THAT THE TIP OF THE NEEDLE HAD BROKEN. ALTHOUGH THERE HAVE BEEN NO PROBLEMS WITH THE PATIENT CURRENTLY, NO BROKEN PIECES OF THE NEEDLE WERE FOUND ON THE SPOT, AND THEY WERE NOT FOUND TO REMAIN INSIDE THE BODY EVEN BY AN X-RAY. IT WILL BE SENT ONE PRODUCT WITH A BROKEN NEEDLE AND ONE OF THE SAME STANDARDS AS A REFERENCE PRODUCT, FOR A TOTAL OF TWO. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22663313 ↗
  • MalfunctionJul 11, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PLASTIC SURGERY ON (B)(6) 2025 AND SUTURE WAS USED. DURING AN UNKNOWN PLASTIC SURGERY, THE SURGEON FELT THAT THE NEEDLE COULD NOT BE PASSED THROUGH THE TISSUE SMOOTHLY DURING USE AND THE SURGEON CHECKED THE TIP VISUALLY, IT WAS CONFIRMED THAT THE TIP OF THE NEEDLE WAS CHIPPED. IT WAS NOT A CONTROL RELEASE NEEDLE. ADDITIONAL SUTURE WAS PERFORMED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22478494 ↗
  • MalfunctionJun 30, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND BARBED SUTURE WAS USED. DURING SURGERY, THE NEEDLE AT THE PROXIMAL END WAS BROKEN. IT WAS NOT A CONTROL RELEASE NEEDLE. NO PIECES WERE LEFT INSIDE THE PATIENT. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22361646 ↗
  • MalfunctionMay 8, 2025Break; Break

    IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN RESPIRATORY SURGERY, IN A UROLOGIC CASE, THE NEEDLE WAS BROKEN AT THE SWAGE PART. IN A DA VINCI CASE, THE NEEDLE TIP DID NOT FALL OFF. NO PIECE FELL INTO THE PATIENT. IT WAS NOT A CONTROL RELEASE NEEDLE. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. WHEN WE SEND THE SAMPLE TO YOU, WE WILL LET YOU KNOW ITS RETURN DATE AND TRACKING NUMBER. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. AFFILIATE REFERENCE NUMBER: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21990785 ↗
  • MalfunctionJan 21, 2025Component Misassembled; Component Misassembled

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, THE PRODUCT WAS A SINGLE-ARMED SUTURE OF 4-0, BUT THERE WAS INCLUDED A DOUBLE-ARMED SUTURE IN THE PACKAGE. THE PRODUCT WAS NOT USED FOR THE PATIENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21198881 ↗
  • MalfunctionJan 15, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HEPATECTOMY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING A HEPATECTOMY, DURING INTERRUPTING SUTURE, THE SUTURE WAS BROKEN WHEN THE INSTRUMENT TIE WAS PERFORMED IN DERMAL SUTURE. THE PRODUCT WAS USED ON THE DERMIS. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21160983 ↗
  • MalfunctionJan 15, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A HEPATECTOMY PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING A HEPATECTOMY, DURING INTERRUPTING SUTURE, THE SUTURE WAS BROKEN WHEN THE INSTRUMENT TIE WAS PERFORMED IN DERMAL SUTURE. THE PRODUCT WAS USED ON THE DERMIS. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21160955 ↗
  • MalfunctionNov 4, 2024Dull, Blunt; Dull, Blunt

    IT WAS REPORTED FROM THE ANALYSIS OF THE RETURNED PRODUCT THAT NEEDLE BREAKAGE WAS OBSERVED. IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. IT WAS REPORTED WHEN THE PACKAGE WAS OPENED THE NEEDLE TIP WAS NOT SHARP AND COULD NOT BE USED AS A NEEDLE. NO HARM WAS DONE BECAUSE IT WAS BEFORE USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20599597 ↗
  • MalfunctionOct 28, 2024Material Separation; Material Separation

    IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN SURGERY ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE WAS DETACHED FROM THE NEEDLE WHEN THE NEEDLE WAS GRASPED WITH THE NEEDLE-HOLDER AND TAKEN OUT. IT WAS NOT A CONTROL RELEASE NEEDLE. FURTHER DETAILS ARE NOT PROVIDED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20550737 ↗
  • MalfunctionSep 13, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED BY A SALES REP THAT, DURING AN UNKNOWN SURGERY, THE NEEDLE WAS BROKEN WHEN THE NEEDLE WAS GRASPED TO USE THE DEVICE. IT WAS NOT A CONTROL RELEASE NEEDLE. FURTHER DETAILS ARE NOT PROVIDED FROM THE HP. ONE SAMPLE WILL BE RETURNED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20212914 ↗
  • MalfunctionSep 3, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SURGERY ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING THE PROCEDURE, WHEN TRIED TO USE THE PRODUCT ON THE DERMIS, THE NEEDLE WAS BROKEN AT THE SWAGE PART. THERE WAS NO EFFECT TO THE PATIENT. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20128774 ↗
  • MalfunctionAug 28, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. DURING THE PROCEDURE, SUTURE BREAKAGE OCCURRED. THE PRODUCT WAS USED ON THE DERMIS, AND THE OPERATION TIME WAS NO EXTENSION. IT WAS NOT A CONTROL RELEASE NEEDLE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20085350 ↗
  • MalfunctionMay 7, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. WHEN TAKING IT OUT FROM THE CASE, THE SUTURE AND NEEDLE WERE COME OFF. IT WAS NOT A CONTROL RELEASE NEEDLE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19258871 ↗
  • MalfunctionFeb 26, 2024Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT A LAPAROSCOPIC CHOLECYSTECTOMY ON AN UNKNOWN DATE, AND SUTURE WAS USED ON THE DERMIS. DURING THE SURGERY, THE SUTURE BROKE DURING USE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18784251 ↗
  • MalfunctionFeb 26, 2024Break; Break

    IT WAS REPORTED THAT PATIENT UNDERWENT A LAPAROSCOPIC CHOLECYSTECTOMY ON AN UNKNOWN DATE, AND SUTURE WAS USED. DURING THE SURGERY, THE NEEDLE WAS BROKEN AT THE TIME OF WOUND CLOSURE. THE BROKEN NEEDLE WAS BURIED UNDER SUBCUTANEOUS. THE NEEDLE PIECE WAS VISUALLY CONFIRMED; IT WAS RETRIEVED DIRECTLY. NO X-RAYS WERE TAKEN TO LOCATE THE BROKEN PIECE. THERE WAS NO ADDITIONAL TISSUE DAMAGE AS A RESULT OF SEARCHING FOR THE NEEDLE PIECE. NO ADDITIONAL INCISION WAS MADE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18784196 ↗
  • MalfunctionJan 9, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN ORTHOPEDIC SURGERY, THE SUTURE BROKE DURING USE AT THE TIME OF WOULD CLOSURE. AFTER THAT, THE NEEDLE WAS BROKEN AT THE BODY PART WHEN THE NEEDLE WAS HELD THE SECOND TIME. THESE BOTH THINGS WERE HAPPENED IN THE SAME NEEDLE-SUTURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18475524 ↗
  • MalfunctionJan 9, 2024Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ORTHOPEDIC PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. DURING AN UNKNOWN ORTHOPEDIC SURGERY, THE SUTURE BROKE DURING USE AT THE TIME OF WOULD CLOSURE. AFTER THAT, THE NEEDLE WAS BROKEN AT THE BODY PART WHEN THE NEEDLE WAS HELD THE SECOND TIME. THESE BOTH THINGS WERE HAPPENED IN THE SAME NEEDLE-SUTURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18475523 ↗

Other Suture, surgical, absorbable, polydioxanone devices

All Suture, surgical, absorbable, polydioxanone devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.