BRITE TIP 401655MRecalls & reported problems
Cordis US Corp. · SI BRITE TIP 6F 55CM STR · Class 2
All models of BRITE TIP (20)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705032000277), received between Dec 8, 2021 and Nov 19, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Difficult to Remove1
- Material Frayed1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 19, 2024Material Frayed; Material Split, Cut or Torn; Material Frayed; Material Split, Cut or Torn
AS REPORTED, THE DOCTOR FELT RESISTANCE WHEN THE 6F 55CM BRITE TIP SHEATH INTRODUCER WAS INSERTED. THE SHEATH WAS REMOVED AND THE TIP GOT FRAYED. IT WAS REPLACED WITH A NEW BRITE TIP SHEATH (DIFFERENT LOT#) AND THE PROCEDURE WAS CONTINUED. THERE WAS NO REPORTED INJURY TO THE PATIENT. ADDITIONAL INFORMATION AND DEVICE WERE REQUESTED; HOWEVER, THE INFORMATION COULD NOT BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20730450 ↗InjuryOct 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE 6F 55CM BRITE TIP SHEATH INTRODUCER DID NOT SLIDE AT THE END OF THE INTERVENTION WHEN REMOVING THE INTRODUCER. THE PHYSICIAN MANAGED TO MAKE IT SLIDE WITH DELICATE AND SLOW MOVEMENTS. THE SHEATH BEGAN TO STRETCH AND FINALLY BECAME BROKEN, LEAVING THE DISTAL PART IN THE PATIENT. FORTUNATELY, USING CHRISTOPHE FORCEPS, THE PHYSICIAN WAS ABLE TO CATCH AND MANAGED TO RECOVER THE MISSING PART OF THE INTRODUCER. THE DEVICE WAS USED IN A CROSS OVER ¿AEL¿ PROCEDURE. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17978249 ↗MalfunctionDec 8, 2021Difficult to Remove; Difficult to Remove
DURING USE OF THE 6F BRITE TIP CATHETER SHEATH INTRODUCER (CSI), THE PHYSICIAN COULD NOT REMOVE THE SHEATH FROM PATIENT BECAUSE THE SHEATH KINKED. THERE WAS NO PATIENT INJURY. THE DEVICE IS AVAILABLE FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12960561 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page