Skip to content

BRITE TIP 401823MRecalls & reported problems

Cordis US Corp. · SI BRITE TIP 8F 23CM STR · Class 2

All models of BRITE TIP (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705032000383), received between Jul 22, 2021 and Apr 29, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 29, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, AN 8F 23 CM BRITE TIP CATHETER SHEATH INTRODUCER (CSI) BROKE OFF INSIDE THE PATIENT. THE PHYSICIAN WAS FIGURE 8 SUTURING THE ACCESS SITE TO REMOVE THE SHEATH WHEN THIS OCCURRED. MANUAL REMOVAL WAS ATTEMPTED WITH NO SUCCESS. THE PATIENT WAS THE CONVERTED FROM MONITORED ANESTHESIA CARE TO GENERAL ANESTHESIA FOR VASCULAR SURGERY TO MAKE AN INCISION AND EXTRACT THE SHEATH. THE SHEATH TIP WAS REMOVED SUCCESSFULLY FROM THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED BUT NOT PROVIDED. THE DEVICE WAS NOT RETURNED FOR EVALUATION AS PREVIOUSLY EXPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19201521 ↗
  • InjurySep 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, A PIECE OF AN 8F BRITE TIP 23CM STR SHEATH BROKE OFF DURING REMOVAL AND REMAINED IN THE PATIENT. THE DEVICE WAS FLUSHED PRIOR TO USE. THE VESSEL DILATOR WAS SNAPPED INTO PLACE AT THE HUB. THERE WERE NO DIFFICULTIES ENCOUNTERED DURING THE INSERTION PROCESS. THERE WERE NO DAMAGES NOTED TO THE DEVICE PRIOR TO USING IT. THE PATIENT WAS STABLE AND DISCHARGED. THE VESSEL AT THE INSERTION SITE WAS NOT NOTED AS A DIFFICULT SITE ACCORDING TO THE PHYSICIAN OPERATION (OP) NOTE. THE DISTAL PORTION, BLACK END THAT IS RADIOPAQUE BROKE OFF INTO THE PATIENT. THERE WAS NO ATTEMPT TO REMOVE THE PIECE THAT BROKE OFF. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: Foreign Body In Patient

    MAUDE report 17844087 ↗
  • MalfunctionJul 22, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    WHEN THE CUSTOMER OPENED THE 8F 23CM STRAIGHT (STR) BRITE TIP CATHETER SHEATH INTRODUCER (CSI), IT WAS NOTED THAT THE BRITE TIP/DISTAL TIP WAS FRACTURED-SEPARATED. THE DEVICE WAS NEVER USED IN THE PATIENT. THE DEVICE WAS STORED PER LABELLING. THE PROCEDURE WAS COMPLETED WITH ANOTHER 8X23 SHEATH. THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12208713 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates