Skip to content

RAIN Sheath Transradial 506616SRecalls & reported problems

Cordis US Corp. · 6F16cm SW Radial · Class 2

All models of RAIN Sheath Transradial (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705032079006), received between Dec 21, 2021 and Dec 21, 2021 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Necrosis1
  • Vascular Dissection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 21, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    APPROXIMATELY 9 DAYS AFTER AN ELECTIVE ANGIOGRAPHIC PROCEDURE INVOLVING A 6F 16CM RAIN SHEATH, A PATIENT´S FINGERS WERE FADING INTO NECROSIS. THE PATIENT WAS TREATED BY A VASCULAR SURGEON WHO FOUND THAT A TRAUMA/WOUND IN THE VASCULAR WALL, A DISSECTION, HAD CAUSED THE NEAR NECROSIS. TREATMENT WAS A BYPASS SURGERY FROM THE BRACHIAL ARTERY TO THE RADIAL ARTERY. THE PATIENT WAS STILL SUFFERING WITH THE FINGERTIP NECROSIS. 2 FINGER AMPUTATION IS AN OPTION. THE RAIN SHEATH WAS USED ACCORDING TO THE INSTRUCTIONS FOR USE. THE USER WAS TRAINED. THE PRODUCT APPEARED NORMAL WHEN TAKEN FROM ITS PACKAGING. THERE WAS NO DIFFICULTY PREPPING THE COMPONENTS. AN ULTRASOUND WAS NOT USED FOR OBTAINING ACCESS AS IT IS NOT COMMONLY USED. THERE WAS NO DIFFICULTY OBTAINING ACCESS. THERE WERE NO MEDICATIONS WERE GIVEN INTRA-ARTERIALLY. THE PATIENT¿S ADMITTING DIAGNOSIS WAS STROKE WITH SOME CHAMBER INSUFFICIENCY. THERE WAS NO DIFFICULTY REMOVING THE RAIN SHEATH AFTER THE PROCEDURE. THE ANGIOGRAPHIC CATHETER AND GUIDEWIRE WERE NON-CORDIS. THERE WAS NO DIFFICULTY REMOVING ANY OF THE OTHER DEVICES. THE RAIN SHEATH WAS REMOVED RIGHT AFTER THE ANGIOGRAPHY. THE PATIENT WAS DISCHARGED AND RETURNED WITH NEAR NECRO

    Patient problems: Necrosis; Vascular Dissection

    MAUDE report 13052162 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page