RAIN Sheath Transradial 506510NRecalls & reported problems
Cordis US Corp. · 5F10CM NW RADIAL · Class 2
All models of RAIN Sheath Transradial (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705032082150), received between May 30, 2023 and Jul 6, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Difficult to Remove1
- Material Frayed1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 6, 2024Material Frayed; Material Split, Cut or Torn; Material Frayed; Material Split, Cut or Torn
AS REPORTED, THE 5F 10CM TRANS-RADIAL RAIN SHEATH KEPT KINKING ON THE END AND IN THE CENTER. THERE WERE NO REPORTS OF PATIENT INJURY. THE DEVICE WAS OPENED IN A STERILE FIELD AND WAS USED IN PATIENT. THE PRODUCT WAS STORED AS PER LABELING. THE DEVICE WILL BE RETURNED FOR EVALUATION. ADDENDUM: PER PE FINDINGS, THE CANNULA OF THE CATHETER SHEATH INTRODUCER (CSI) PRESENTS AN ACCORDIONED CONDITION LOCATED APPROXIMATELY ON 8 CM FROM THE DISTAL TIP. ALSO, THE CANNULA PRESENTS A FRAYED CONDITION ON THE DISTAL TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19685995 ↗InjuryNov 15, 2023Fracture; Fracture
AS REPORTED, THE NITINOL WIRE IN THE 5F 10CM RAIN SHEATH TRANSRADIAL KIT BECAME KNOTTED UPON ADVANCEMENT THROUGH THE NEEDLE. IN ATTEMPTING TO REMOVE THE WIRE, THE DISTAL ASPECT OF THE WIRE BROKE OFF AND REMAINED IN THE PATIENT. THE INTENDED PROCEDURE WAS REPORTED TO BE A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA). THE TARGET LESION WAS THE RIGHT SUPERFICIAL FEMORAL ARTERY. LESION CALCIFICATION WAS MODERATE WITH NO VESSEL TORTUOSITY OR STENOSIS BUT A CHRONIC TOTAL OCCLUSION (CTO). THERE WERE NO ANOMALIES NOTED WHEN THE PRODUCT WAS REMOVED FROM THE PACKAGE. THE PRODUCT WAS STORED, HANDLED, INSPECTED, AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY EXPERIENCED IN PREPPING THE DEVICE. RESISTANCE WAS EXPERIENCED ADVANCING THE WIRE. NO UNUSUAL FORCE WAS NECESSARY DURING USE OF THE DEVICE. THE SEPARATED SEGMENT WAS LEFT IN THE PATIENT, BUT THE WIRE IS EXTRA VASCULAR, AND THE PATIENT IS FOLLOWING UP WITH GENERAL SURGERY. THE PATIENT WAS NOT HOSPITALIZED OR REQUIRED EXTENDED HOSPITALIZATION BECAUSE OF THIS EVENT. A PROCEDURAL CD IS NOT AVAILABLE. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18144742 ↗MalfunctionAug 4, 2023Fracture; Fracture
AS REPORTED, THE GUIDEWIRE OF A 5F 10CM RAIN TRANS-RADIAL CATHETER SHEATH INTRODUCER (CSI) WAS SPLIT INTO TWO PARTS DURING PREPARATION OF THE DEVICE. AS A RESULT, THE SAME 5F 10CM RAIN CSI WAS USED WITH AN UNKNOWN GUIDEWIRE TO CONTINUE THE PROCEDURE. THERE WAS NO REPORTED INJURY TO THE PATIENT. THERE WERE NO DIFFICULTIES DURING THE INSERTION OR WITHDRAWAL OF THE CSI. THE DEVICE WAS NOT RETURNED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17463278 ↗MalfunctionMay 30, 2023Fracture; Difficult to Remove; Fracture; Difficult to Remove
AS REPORTED, A 5F 10 CM TRANS RADIAL RAIN SHEATH ACCORDIONED AS IT WAS ATTEMPTED TO BE INSERTED IN A LEFT PEDAL ARTERY WITH GUIDED ULTRASOUND (US). THERE WAS DIFFICULTY INSERTING AND WITHDRAWING DUE TO THE ACCORDIONED CONDITION OF DEVICE. WHEN ATTEMPTING TO REMOVE THE SHEATH OVER THE WIRE. THE CSI WIRE BECAME ENTRAPPED DURING THE PROCEDURE AND ABNORMAL FORCE WAS REQUIRED WHEN WITHDRAWING THE DEVICE. THE WIRE AND SHEATH WERE REMOVED. THE DISTAL TIP OF THE WIRE BROKE OFF THE REST OF THE WIRE. THE WIRE WAS SUCCESSFULLY RETRIEVED AND REMOVED BY THE PHYSICIAN. THERE WERE NO REPORTS OF PATIENT INJURY. THE DEVICE WAS USED IN AN INTERVENTIONAL ENDOVASCULAR PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA). THE DEVICE WAS PREPPED, FLUSHED, AND WIPED WITH HEPARINIZED SALINE PRIOR TO INSERTION TO ACTIVATE THE HYDROPHILIC COATING. THE SHEATHS ARE STORED IN THE PRODUCT BOX IN A TEMP/HUMIDITY-CONTROLLED PROCEDURE ROOM. THE SHEATH WAS INSPECTED PRIOR TO USE, WITH NO DEFECTS NOTED. SEVERE CALCIFICATION WAS NOTED BY THE PHYSICIAN VIA ULTRASOUND. THE LENGTH OF STAY WAS NOT INCREASED DUE TO THE PRODUCT MALFUNCTION. ANOTHER 5F RAIN SHEATH WAS USED FOR THE PROCEDURE. THE VESSEL DILATOR WAS SNAPPED INTO PLACE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17026416 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page