EXPEDIUM 179732535Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM SI POLYAXIAL EXTEND TAB 5.5 x 5 x 35mm · Class 2 · Implant
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034073682), received between May 20, 2022 and Dec 7, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Migration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Migration2
- Break1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 7, 2022Migration; Migration
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A SPINAL FUSION (TH8 TO L3) FOR THE THORACOLUMBAR VERTEBRAE ON (B)(6) 2022. DURING INSERTION OF THE ROD, BACK-OUT OF THE SIX SCREWS IN QUESTION OCCURRED AT L1 TO L3. IT SEEMED THAT BONE QUALITY WAS DECLINING BECAUSE OF THE PATIENT AGE. SCREWS WERE REINSERTED BY INCREASING THE SCREW SIZE OR CHANGING THE TRAJECTORY OF THE INSERTION HOLE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES DELAY. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM SI POLYAXIAL EXTEND TAB 5.5 X 5 X 35MM. THIS IS REPORT 6 OF 6 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15931474 ↗MalfunctionDec 7, 2022Migration; Migration
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A SPINAL FUSION (TH8 TO L3) FOR THE THORACOLUMBAR VERTEBRAE ON (B)(6) 2022. DURING INSERTION OF THE ROD, BACK-OUT OF THE SIX SCREWS IN QUESTION OCCURRED AT L1 TO L3. IT SEEMED THAT BONE QUALITY WAS DECLINING BECAUSE OF THE PATIENT AGE. SCREWS WERE REINSERTED BY INCREASING THE SCREW SIZE OR CHANGING THE TRAJECTORY OF THE INSERTION HOLE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES DELAY. THE PATIENT OUTCOME WAS REPORTED TO BE STABLE. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM SI POLYAXIAL EXTEND TAB 5.5 X 5 X 35MM. THIS IS REPORT 5 OF 6 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15931473 ↗MalfunctionMay 20, 2022Break; Material Twisted/Bent; Break; Material Twisted/Bent
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PSF (TH11) FOR TREATING IDIOPATHIC SCOLIOSIS ON (B)(6) 2022, THE SURGEON TRIED TO DEPLOY THE SETSCREW ONTO THE SI SCREW, THE TABS ON THE SI SCREW BECAME BENT AND CAUSED CROSS-THREADING. THE SURGEON BROKE THE TABS AND SEPARATED THEM FROM THE SI SCREW. HOWEVER, THIS TAB BREAKING RESULTED IN DAMAGE TO THE SI SCREW¿S THREADS. THE SURGEON DECIDED TO REPLACE THE SI SCREW. THE PROCEDURE WAS COMPLETED WITH A FIFTEEN MINUTE DELAY. THIS REPORT IS FOR A SI POLYAXL EXT TAB 5 X 35MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14453613 ↗
Other Orthosis, spondylolisthesis spinal fixation devices
- CD HORIZON® Spinal System 7540020Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation257 reports
- MANTIS Redux 48289999Stryker Corporation · Orthosis, spondylolisthesis spinal fixation76 reports
- EXPEDIUM TOP NOTCH 179771555Depuy Spine, LLC · Orthosis, spondylolisthesis spinal fixation41 reports
- EXPEDIUM 179712650Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation31 reports
- EXPEDIUM VERSE 199721530Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4 5 48130000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation29 reports
- XIA 4.5 System - XIA CT 48850000Stryker Corporation · Orthosis, spondylolisthesis spinal fixation27 reports
- CD HORIZON® Spinal System 7576545Medtronic Sofamor Danek, Inc. · Orthosis, spondylolisthesis spinal fixation26 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.