Skip to content

EXPEDIUM 179788650Recalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM UNIPLANAR SCREW 5.5 x 6.0 x 50mm · Class 2 · Implant

All models of EXPEDIUM (291)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034081410), received between Aug 16, 2019 and Sep 22, 2020 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Slipped” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Slipped1
  • Loose or Intermittent Connection1

Patient problems

  • No Code Available2
  • Failure of Implant1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 22, 2020Device Slipped; Device Slipped

    IT WAS REPORTED THAT ON UNKNOWN DATE, THERE WAS A POST-OPERATIVE DISCONNECTION WITH A FAILURE AT SCREW, ROD INTERFACE AT DISTAL END OF SCOLIOSIS CORRECTING SEGMENTAL PEDICAL SCREW CONSTRUCT. THE SURGEON SUSPECTS THAT A CERTAIN AMOUNT OF MEDIAL OR LATERAL FORCE FROM THE ROD ON THE TULIP HEAD SPLAYS HEAD ENOUGH TO ALLOW SLIPPAGE. THERE IS NO APPARENT MALFUNCTION WITH THE ROD. THE PATIENT IS CURRENTLY CHARACTERIZED AS AN ASYMPTOMATIC HARDWARE FAILURE. NO FURTHER INFORMATION PROVIDED. CONCOMITANT DEVICE REPORTED: ROD (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 2); SINGLE-INNER SETSCREW (PART # 179702000, LOT # UNKNOWN, QUANTITY 1); SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 4). THIS REPORT IS FOR ONE (1) EXP TI UNI SCREW 6MM X 50MM. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: Failure of Implant; No Code Available

    MAUDE report 10567796 ↗
  • InjuryAug 16, 2019Loose or Intermittent Connection; Loose or Intermittent Connection

    REVISION FUSION: (B)(6), (B)(6) 2019. PRIMARY IMPLANT DATE: (B)(6) 2018. REVISION FOR DISLODGED RODS. SET SCREWS WERE NO LONGER IN SCREW HEADS RIGHT SIDE ROD NO LONGER STABLE CONCOMITANT DEVICE REPORTED: UNKNOWN ROD (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY ONE). THIS COMPLAINT INVOLVES 11 (ELEVEN) DEVICES.

    Patient problems: No Code Available

    MAUDE report 8902463 ↗

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page