Skip to content

VIPER 186788045Recalls & reported problems

Depuy Spine, LLC · VIPER 2 SYSTEM ROD, PRE-LORDOSED 45mm · Class 2 · Implant

Adverse event reports
10
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10705034131740), received between Nov 13, 2019 and Jan 8, 2024 — classified as 8 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K071860, K111136); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem7
  • Break2
  • Insufficient Device Problem Information1

Patient problems

  • Post Operative Wound Infection3
  • Osteolysis2
  • Unspecified Infection2
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 8, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SWITZERLAND AS FOLLOWS: THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: IMPLANT FAILURE. COMPLICATION TIME/INTERVENTION. REPEAT SURGERY. THIS IS FOR DEPUY SYNTHES SPINE VIPER AND VIPER /OTHER PRODUCTS. THIS IS REPORT 3 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18465394 ↗
  • InjuryJan 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: WOUND INFECTION DEEP, WOUND INFECTION SUPERFICIAL, COMPLICATION TIME/INTERVENTION, POSTOPERATIVE (BEFORE DISCHARGE). THIS IS FOR DEPUY SYNTHES SPINE VIPER /OTHER AND EXPEDIUM POLYAXIAL PRODUCTS (VIPER2 LORDOTIC ROD-45MM (B)(4).

    Patient problems: Post Operative Wound Infection

    MAUDE report 18454431 ↗
  • InjuryOct 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2023, AS SCHEDULED AND IT WAS COMPLETED WITH NO PROBLEMS. THE FOLLOWING PROCEDURES WERE DONE IN THE REVISION SURGERY. L3/4: XLIF WAS PERFORMED. L4/5: SET SCREWS AND RODS INSERTED IN THE INITIAL SURGERY WERE REMOVED. L2 AND L3: DIAMETER 6.0 - LENGTH 50 MM VIPER PRIME SCREWS WERE INSERTED. S: DIAMETER 6.0 - LENGTH 45 MM VIPER PRIME SCREW WAS INSERTED. L2/S: DIAMETER 5.5 MM VIPER ROD WAS INSERTED. SET SCREWS WERE INSERTED AND THE FINAL TIGHTENING WAS DONE.

    Patient problems: Osteolysis

    MAUDE report 17923237 ↗
  • InjuryOct 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2023, AS SCHEDULED, AND IT WAS COMPLETED WITH NO PROBLEMS. THE FOLLOWING PROCEDURES WERE DONE IN THE REVISION SURGERY. L3/4: XLIF WAS PERFORMED. L4/5: SET SCREWS AND RODS INSERTED IN THE INITIAL SURGERY WERE REMOVED. L2 AND L3: DIAMETER 6.0 - LENGTH 50 MM VIPER PRIME SCREWS WERE INSERTED. S: DIAMETER 6.0 - LENGTH 45 MM VIPER PRIME SCREW WAS INSERTED. L2/S: DIAMETER 5.5 MM VIPER ROD WAS INSERTED. SET SCREWS WERE INSERTED AND THE FINAL TIGHTENING WAS DONE.

    Patient problems: Osteolysis

    MAUDE report 17923225 ↗
  • InjuryMar 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A PLIF (L4/L5) FOR SPINAL STENOSIS ON (B)(6) 2023. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. AFTER SURGERY, THE INFECTION OCCURRED. ON (B)(6) 2023, DEBRIDEMENT WILL BE DONE WITH THE INCISION OPENED UNDER GENERAL ANESTHESIA DUE TO INFECTION. DURING THIS TIME, SOME IMPLANTED PRODUCTS MAY BE REMOVED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) VIPER 2 SYSTEM ROD, PRE-LORDOSED 45MM. THIS IS REPORT 5 OF 6 FOR COMPLAINT (B)(4).

    Patient problems: Unspecified Infection

    MAUDE report 16598338 ↗
  • InjuryMar 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A PLIF (L4/L5) FOR SPINAL STENOSIS ON (B)(6) 2023. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. AFTER SURGERY, THE INFECTION OCCURRED. ON (B)(6) 2023, DEBRIDEMENT WILL BE DONE WITH THE INCISION OPENED UNDER GENERAL ANESTHESIA DUE TO INFECTION. DURING THIS TIME, SOME IMPLANTED PRODUCTS MAY BE REMOVED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) VIPER 2 SYSTEM ROD, PRE-LORDOSED 45MM. THIS IS REPORT 6 OF 6 FOR COMPLAINT (B)(4) .

    Patient problems: Unspecified Infection

    MAUDE report 16598244 ↗
  • MalfunctionDec 23, 2022Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, A VIPER 2 50MM ROD WAS INSERTED INTO THE VIPER PRIME SCREWS, THE SET SCREWS WERE INSERTED AND TORQUED OFF, THE VIPER PRIME ROD HOLDER WAS REMOVED ACCORDING TO THE TECHNIQUE GUIDE. WHEN TAKING A FINAL INTRA-OP X-RAY, THE SURGEON DISCOVERED THAT THE VIPER 2 50MM ROD HAD BROKEN, AND A FRAGMENT OF THAT ROD WAS RETAINED IN THE PATIENT. UNDER FLUOROSCOPIC GUIDANCE, THE SURGEON CAREFULLY REMOVED THE FRAGMENT OF THE ROD, REMOVED ALL SET SCREWS, AND REMOVED THE ROD. A SECOND ROD WAS INSERTED USING THE SAME TECHNIQUE AS ABOVE. UPON A FINAL INTRA-OP X-RAY, THE SURGEON AGAIN NOTICED THAT THE VIPER ROD HAD BROKEN. BOTH TIMES, THIS HAPPENED AT THE CONNECTION POINT BETWEEN THE ROD AND THE ROD INSERTER INSTRUMENT. THE SECOND FAILURE RESULTED IN A SIMILAR BROKEN ROD FRAGMENT. THIS WAS REMOVED UNDER FLUOROSCOPIC GUIDANCE. THE REMAINING VIPER ROD REMAINED IN THE PATIENT AS IT WAS SECURED WITHIN THE CONSTRUCT, AND THE SURGEON DIDN¿T WANT TO PUT THE PATIENT THROUGH MORE TIME IN THE OR. TO THE BEST OF THE REPORTER¿S KNOWLEDGE, THERE WERE NO FRAGMENTS LEFT IN THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE VIPER 2 SYSTEM ROD, PRE-LORDOSE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16042127 ↗
  • InjuryDec 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT PRESENTED WITH ISSUES RELATED TO A PREVIOUS FUSION DONE AT (B)(6) HOSPITAL IN (B)(6) ON (B)(6) 2020 AND ALL THE HARDWARE WAS SUCCESSFULLY REMOVED. THE PROCEDURE WAS AN L4-5 RIGHT SIDED TLIF WITH VIPER PRIME AND CONDUIT PLIF CULTURES WERE TAKEN INTRA-OP FOR A SUSPECTED FUNGAL INFECTION. PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN. THIS COMPLAINT INVOLVES ELEVEN (11) DEVICES. THIS REPORT IS FOR (1) VIPER2 LORDOTIC ROD-45MM. THIS REPORT IS 5 OF 10 FOR (B)(4).

    Patient problems: Post Operative Wound Infection

    MAUDE report 11049317 ↗
  • InjuryDec 21, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT PRESENTED WITH ISSUES RELATED TO A PREVIOUS FUSION DONE AT (B)(6) ON (B)(6) 2020 AND ALL THE HARDWARE WAS SUCCESSFULLY REMOVED. THE PROCEDURE WAS AN L4-5 RIGHT SIDED TLIF WITH VIPER PRIME AND CONDUIT PLIF CULTURES WERE TAKEN INTRA-OP FOR A SUSPECTED FUNGAL INFECTION. PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN. THIS COMPLAINT INVOLVES ELEVEN (11) DEVICES. THIS REPORT IS FOR (1) VIPER2 LORDOTIC ROD-45MM. THIS REPORT IS 4 OF 10 FOR (B)(4).

    Patient problems: Post Operative Wound Infection

    MAUDE report 11049315 ↗
  • MalfunctionNov 13, 2019Break; Break

    ON (B)(6) 2017, AN INITIAL TLIF (TRANSFORAMINAL LUMBAR INTERBODY FUSION) PROCEDURE WAS PERFORMED TO STABILIZE L4-5 AT A DIFFERENT HOSPITAL. ON (B)(6) 2019, THE SURGEON PERFORMED A TLIF REVISION PROCEDURE AND A PPS (PERCUTANEOUS PEDICLE SCREW) EXTENSION PROCEDURE TO TREAT SPINAL STENOSIS, STABILIZING L5-S1. THE POLY DRIVER (279734000) BROKE OFF AGAINST THE HIGHLY OSSIFIED BONE. THE SURGEON REMOVED THE FRAGMENT AND A SCREW AND COMPLETED THE PROCEDURE WITH REPLACEMENTS LESS THAN A 30-MINUTE DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES ONE (1) DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9314687 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071860, K111136. One submission often covers many models, so check the product and lot details in each record.

  • Z-1674-2015 · Apr 21, 2015 · Terminated
    EXPEDIUM Spine System/Synapse System; MODEL: EXPEDIUM Offset Conical Lock Nut 8mm, CATALOG NO. : 1754-91-150, UDI NUMBER (if applicable): GTIN: 10705034032467

Other Orthosis, spondylolisthesis spinal fixation devices

All Orthosis, spondylolisthesis spinal fixation devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.