MOUNTAINEER 188318328Recalls & reported problems
Medos International Sàrl · MOUNTAINEER OCT SPINAL SYSTEM MINI POLYAXIAL BONE SCREW 3.5 x 3.5 x 28mm · Class 2 · Implant
All models of MOUNTAINEER (51)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034157252), received between May 15, 2018 and Oct 25, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
Patient problems
- No Consequences Or Impact To Patient2
- Not Applicable1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 25, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THE INNER SCREW OF THE SCREW BACKED OUT - COULD NOT INSERTER THE SCREW. PATIENT CONSEQUENCE? : NO. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? : YES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8006344 ↗MalfunctionMay 15, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
ON (B)(6) 2018, THE SURGERY FOR THE CERVICAL OPLL WAS PERFORMED BY USING THE MOUNTAINEER SYSTEM. THE FIXED AREA WAS C2 ¿ 5. THE REPORTED FAVORED ANGLE SCREW WAS USED WHEN THE SURGEON WAS INSERTING THE C2 LAMINA SCREW. WHEN THE SURGEON INSTALLED THE SCREW TO THE DRIVER, THERE WAS NO PROBLEM. WHEN THE SURGEON REMOVED THE SCREW FROM THE SCREW DRIVER, SINCE THE SURGEON PUSHED THE BUTTON ON THE DRIVER INNER PORTION, THE SCREW COULD NOT BE DETACHED FROM THE SCREW DRIVER. THE SURGEON INSTALLED THE SCREW TO THE SCREW DRIVER AGAIN. HOWEVER, THE SURGEON PUSHED THE BUTTON ON THE DRIVER INNER PORTION THREE TIMES AND REPEATED THE SCREW INSTALLATION, THE INNER PART OF THE SCREW HEAD WAS COME UP; THUS, THE SURGEON COULD NOT PLACE THE CORE DRIVER OR THE ROD ANYMORE. THE SURGERY WAS COMPLETED WITH A 30-MINUTE-DELAY, AND THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient; Not Applicable
MAUDE report 7515767 ↗
Other Orthosis, cervical pedicle screw spinal fixation devices
- INFINITY™ Occipitocervical Upper Thoracic System 3604026Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation2 recalls1 report
- CONNECTOR 79-2003Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 360PT4022Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- CONNECTOR 79-2002Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604532Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604024Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- OASYS 48551000Stryker Corporation · Orthosis, cervical pedicle screw spinal fixation45 reports
- YUKON OCT Spinal System 7601-05030VB Spine LLC · Orthosis, cervical pedicle screw spinal fixation15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page